510(k) Report

510(k) · Cleared

Cleared

Hemoclip

K172727510(k) TraditionalCleared Apr 16, 2018217-day reviewClass II21 CFR 876.4400

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hangzhou AGS MedTech Co., Ltd. · PKL Hemostatic Metal Clip For The Gi Tract

Cleared on Apr 16, 2018 as substantially equivalent to K122660 (Boston Scientific Corporation, 2012) and 1 other predicate after a 217-day review. Its family tree reaches back 7 generations to 1997. 8 later devices descend from it.

Decided
Apr 16, 2018
Review
217 days
Class
Class II
Product code
PKL
Recalls
None
Predicates
2

Its longest line of descent runs back 7 generations to K963160 (1997). 8 later devices descend from it, 6 citing it directly.

  1. Root · 7 generations backK963160 HX-5/6-1 Endoscopic Clipping DeviceOlympus America, Inc. · 1997
  2. 6 generations backK990687 HX-5/6-1 Endoscopic Clipping Device, Endoscopic ClipsOlympus Optical, Co. · 1999
  3. 5 generations backK013066 HX-5/6-1 Endoscopic Clipping DeviceOlympus America, Inc. · 2001 · Recalled
  4. 4 generations backK040148 Resolution Hemostasis Clipping DeviceBoston Scientific Corp · 2004 · Recalled
  5. 3 generations backK122660 Resolution Clip Model M00522600, M00522601, M00522602, M00522610…Boston Scientific Corporation · 2012
  6. 2 generations backK152001 Sterile Repositionable Hemostasis Clipping DeviceMicro-Tech (Nanjing) Co., Ltd. · 2015
  7. 1 generation backK161463 SureClip (TM) Repositionable Hemostasis ClipMicro-Tech (Nanjing) Co., Ltd. · 2017
  8. This deviceK172727 HemoclipHangzhou AGS MedTech Co., Ltd. · 2018
  9. Descendant · cites it directly · 2 of its ownK211787 HemoclipHangzhou AGS MedTech Co., Ltd. · 2022
  10. Descendant · 2 generations on · 0 of its ownK252271 HemoclipHangzhou AGS MedTech Co., Ltd. · 2025
  11. Descendant · cites it directly · 0 of its ownK241005 Haemostasis ClipsJiangxi Zhuoruihua Medical Instrument Co., Ltd. · 2025
  12. Descendant · 2 generations on · 0 of its ownK234124 Single Use HemoclipsPromisemed Hangzhou Meditech Co., Ltd. · 2024
  13. Descendant · cites it directly · 0 of its ownK232450 Disposable Hemostatic ClipsZhejiang Soudon Medical Technology Co., Ltd. · 2024

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2022: 22023: 12024: 32025: 2
2022high 3 (2024) · low 0 (2026)2026

Most influential descendants

Showing 5 of 8 · See all
  • K211787HemoclipHangzhou AGS MedTech Co., Ltd. · 2022 · cites it directly2descendants
  • K252271HemoclipHangzhou AGS MedTech Co., Ltd. · 2025 · 2 generations on0descendants
  • K241005Haemostasis ClipsJiangxi Zhuoruihua Medical Instrument Co., Ltd. · 2025 · cites it directly0descendants
  • K234124Single Use HemoclipsPromisemed Hangzhou Meditech Co., Ltd. · 2024 · 2 generations on0descendants
  • K232450Disposable Hemostatic ClipsZhejiang Soudon Medical Technology Co., Ltd. · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K963160HX-5/6-1 Endoscopic Clipping DeviceOlympus America, Inc.1997—
−6K990687HX-5/6-1 Endoscopic Clipping Device, Endoscopic ClipsOlympus Optical, Co.1999—
−5K013066HX-5/6-1 Endoscopic Clipping DeviceOlympus America, Inc.20011
−4K040148Resolution Hemostasis Clipping DeviceBoston Scientific Corp20042
−3K122660Resolution Clip Model M00522600, M00522601, M00522602, M00522610, M00522611…Boston Scientific Corporation2012—
−2K152001Sterile Repositionable Hemostasis Clipping DeviceMicro-Tech (Nanjing) Co., Ltd.2015—
−1K161463SureClip (TM) Repositionable Hemostasis ClipMicro-Tech (Nanjing) Co., Ltd.2017—
0 (this device)K172727HemoclipHangzhou AGS MedTech Co., Ltd.2018—
+1K211787HemoclipHangzhou AGS MedTech Co., Ltd.2022—
+2K252271HemoclipHangzhou AGS MedTech Co., Ltd.2025—
+1K241005Haemostasis ClipsJiangxi Zhuoruihua Medical Instrument Co., Ltd.2025—
+2K234124Single Use HemoclipsPromisemed Hangzhou Meditech Co., Ltd.2024—
+1K232450Disposable Hemostatic ClipsZhejiang Soudon Medical Technology Co., Ltd.2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202220
202310
202430
202520
FIG. 1 — Lineage of K172727, 1997–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PKL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PKL device

  1. #1 K202333 Lockado Repositionable Hemostasis Clipscore 17.5 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122660Resolution Clip Model M00522600, M00522601, M00522602, M00522610, M00522611, M00522612
Summary page 4
…Name: Hemorrhoidal ligator Product Code Description: Hemostatic Metal Clip For The Gi Tract Product Code: PKL 5.3 Predicate Device K122660 - ResolutionTM Hemostasis Clipping Device(Primary) K161463- SureClip(TM) Repositionable Hemostasis Clip(Secondary) 5.4 Devic…
Boston Scientific Corporation2012—
*K161463SureClip (TM) Repositionable Hemostasis Clip
Summary page 4
…n: Hemostatic Metal Clip For The Gi Tract Product Code: PKL 5.3 Predicate Device K122660 - ResolutionTM Hemostasis Clipping Device(Primary) K161463- SureClip(TM) Repositionable Hemostasis Clip(Secondary) 5.4 Device Description The Hemoclip is a sterile device consisting of a…
Micro-Tech (Nanjing) Co., Ltd.2017—

Also cleared against K122660: 2 other devices →

What descends from it? (8)

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211787HemoclipHangzhou AGS MedTech Co., Ltd.20222
K241005Haemostasis ClipsJiangxi Zhuoruihua Medical Instrument Co., Ltd.20250
K232450Disposable Hemostatic ClipsZhejiang Soudon Medical Technology Co., Ltd.20240
K231633Disposable Hemoclip (AF series)Alton (Shanghai) Medical Instruments Co., Ltd.20240
K230004Disposable HemoclipYangzhou Fartley Medical Instrument Technology Co., Ltd.20230
K213143HemoclipHangzhou AGS MedTech Co., Ltd.20220

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 6 companies; a new generation was cleared every 2.4 years on median.
  • 2 of the 7 ancestors have a recall of any class.
  • Of its descendants, K211787 (Hangzhou AGS MedTech Co., Ltd., 2022) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K122660.

How does its category compare?

All PKL devices

PKL Hemostatic Metal Clip For The Gi Tract: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · PKL

6in 2025

2010: 22012: 32013: 22014: 12015: 32016: 12017: 22018: 52019: 42020: 62021: 22022: 92023: 22024: 82025: 62026 (partial year): 7
2007high 9 (2022) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PKL

162in 2025

2012: 102 days median2015: 143 days median2018: 116 days median2019: 231 days median2020: 164 days median2022: 259 days median2024: 181 days median2025: 162 days median2026 (partial year): 231 days median
2007high 259 (2022) · low 102 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PKL

1in 2019

2019: 1
2007high 1 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PKL

1,170in 2025

2012: 12014: 42015: 342016: 1282017: 3982018: 3932019: 6992020: 5042021: 6682022: 5222023: 8632024: 1,2162025: 1,170
2007high 1,216 (2024) · low 0 (2007)2025

MAUDE reports naming product code PKL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20102—00
20110—00
2012310201
20132—00
20141—04
20153143034
20161—0128
20172—0398
201851160393
201942311699
202061640504
20212—0668
202292590522
20232—0863
2024818101,216
2025616201,170
202672310—

What is it indicated for?

The hemoclip is indicated for endoscopic clip placement within the gastrointestinal tract for the purpose of:

1. Endoscopic marking

2. Hemostasis for

Mucosal/sub-mucosal defects <3cm

Bleeding ulcers

Arteries<2mm

Polyps<1.5cm in diameter

Diverticula in the colon

Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection

3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel

4. As a supplementary method, closure for GI tract luminal perforations<20mm that can be treated conservatively.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K172727: Hemoclip.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172727

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026