510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Single Use Electrosurgical Snare SD-400

K172734510(k) TraditionalCleared Dec 7, 201787-day reviewClass II21 CFR 876.4300Recalls 41

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Medical Systems Corp. · FDI Snare, Flexible

Cleared on Dec 7, 2017 as substantially equivalent to K133987 (Boston Scientific Corporation, 2014) after an 87-day review. Its family tree reaches back 4 generations to 1995. 3 later devices descend from it. It has been recalled 41 times.

Decided
Dec 7, 2017
Review
87 days
Class
Class II
Product code
FDI
Recalls
41
Predicates
1

Its longest line of descent runs back 4 generations to K950496 (1995). 6 ancestors in all, the oldest K941750 (1994). 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1994: 1 other ancestor2018: 1 descendant cleared2022: 1 descendant cleared2019: 1 descendant clearedK950496 · Snare, Flexible, Electrosurgical Cutting & Coagulation Device & Accessories · Boston Scientific Corp · 1995K992477 · Rotosnare · Inscope, LLC · 1999×2: 2 generations cleared close together, K131700 (2013), K133987 (2014) 1990200020102020
  1. Root · 4 generations backK950496 Snare, Flexible, Electrosurgical Cutting & Coagulation Device &…Boston Scientific Corp · 1995
  2. 3 generations backK992477 RotosnareInscope, LLC · 1999
  3. 2 generations backK131700 Sensation Short Throw, Captiflex, Captivator, Captivator II…Boston Scientific Corp · 2013 · Recalled
  4. 1 generation backK133987 Captivator II,Single-Use Polypectomy SnaresBoston Scientific Corporation · 2014 · Recalled
  5. This deviceK172734 Single Use Electrosurgical Snare SD-400Olympus Medical Systems Corp. · 2017
  6. Descendant · cites it directly · 1 of its ownK173673 AcuSnare Polypectomy SnareWilson-Cook Medical, Inc. · 2018
  7. Descendant · cites it directly · 0 of its ownK220089 Disposable Polyp SnareJiangsu Vedkang Medical Science and Technology Co., Ltd. · 2022
  8. Descendant · 2 generations on · 0 of its ownK191048 AcuSnare Polypectomy Snare - DuckbillWilson-Cook Medical, Inc. · 2019

Most influential descendants

Showing 3 of 3 · See all
  • K173673AcuSnare Polypectomy SnareWilson-Cook Medical, Inc. · 2018 · cites it directly1descendant
  • K220089Disposable Polyp SnareJiangsu Vedkang Medical Science and Technology Co., Ltd. · 2022 · cites it directly0descendants
  • K191048AcuSnare Polypectomy Snare - DuckbillWilson-Cook Medical, Inc. · 2019 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K950496Snare, Flexible, Electrosurgical Cutting & Coagulation Device & AccessoriesBoston Scientific Corp1995—
−3K992477RotosnareInscope, LLC1999—
−2K131700Sensation Short Throw, Captiflex, Captivator, Captivator II, Profile…Boston Scientific Corp20138
−1K133987Captivator II,Single-Use Polypectomy SnaresBoston Scientific Corporation20141
0 (this device)K172734Single Use Electrosurgical Snare SD-400Olympus Medical Systems Corp.2017—
+1K173673AcuSnare Polypectomy SnareWilson-Cook Medical, Inc.2018—
+1K220089Disposable Polyp SnareJiangsu Vedkang Medical Science and Technology Co., Ltd.2022—
+2K191048AcuSnare Polypectomy Snare - DuckbillWilson-Cook Medical, Inc.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
201910
202210
FIG. 1 — Lineage of K172734, 1995–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FDI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FDI device

  1. #1 K251692 Advanced Tissue Resection Devicescore 10.9 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133987Captivator II,Single-Use Polypectomy Snares
Summary page 5
…cate Device Device name Applicant 510(k)# CAPTIVATOR II,SINGLE-USE Boston Scientific Corporation K133987 POLYPECTOMY SNARES IV. Device Description The subject device is designed to be used with an Olympus endoscope for the removal an…
Boston Scientific Corporation20141

Also cleared against K133987: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K173673AcuSnare Polypectomy SnareWilson-Cook Medical, Inc.20181
K220089Disposable Polyp SnareJiangsu Vedkang Medical Science and Technology Co., Ltd.20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 6, 2025Class IIZ-0483-2026PackagingSterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.
Oct 6, 2025Class IIZ-0482-2026PackagingSterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.
Jan 4, 2021Class IIZ-0991-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0993-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0994-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0996-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0997-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1000-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1002-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1004-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1005-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1006-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1010-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1017-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1023-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1025-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0992-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0995-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1001-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1007-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1008-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1009-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1012-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1014-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1018-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1026-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1027-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1028-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1029-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1003-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1011-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1015-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1016-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1019-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1020-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1022-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0998-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0999-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1013-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1021-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-1024-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 4.4 years on median.
  • 4 of the 6 ancestors have a recall of any class.
  • Of its descendants, K173673 (Wilson-Cook Medical, Inc., 2018) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K133987.

How does its category compare?

All FDI devices

FDI Snare, Flexible: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · FDI

1in 2023

2008: 22009: 12011: 12012: 12013: 32014: 52015: 32016: 12017: 12018: 42019: 32020: 22021: 12022: 42023: 12026 (partial year): 1
2007high 5 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FDI

247in 2022

2013: 87 days median2014: 47 days median2015: 62 days median2018: 214 days median2019: 100 days median2022: 247 days median
2007high 247 (2022) · low 47 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FDI

2in 2025

2010: 12015: 12025: 2
2007high 2 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FDI

274in 2025

2007: 482008: 2012009: 2652010: 1742011: 1772012: 1622013: 1912014: 3452015: 1952016: 2232017: 1542018: 1812019: 1352020: 4872021: 2112022: 2462023: 2092024: 2792025: 274
2007high 487 (2020) · low 48 (2007)2025

MAUDE reports naming product code FDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—048
20082—0201
20091—0265
20100—1174
20111—0177
20121—0162
20133870191
20145470345
20153621195
20161—0223
20171—0154
201842140181
201931000135
20202—0487
20211—0211
202242470246
20231—0209
20240—0279
20250—2274
20261—0—

What is it indicated for?

These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172734: Single Use Electrosurgical Snare SD-400.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172734

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026