510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Intellijoint HIP System (Cart), Intellijoint HIP System (Portable), Intellijoint HIP Instrument Tray

K172849510(k) SpecialCleared Nov 16, 201757-day reviewClass II21 CFR 882.4560Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Intellijoint Surgical, Inc. · OLO Orthopedic Stereotaxic Instrument

Cleared on Nov 16, 2017 as substantially equivalent to K151364 (Intellijoint Surgical, Inc., 2015) and 1 other predicate after a 57-day review. Its family tree reaches back 10 generations to 2001. 1 later device descends from it. It has been recalled once.

Decided
Nov 16, 2017
Review
57 days
Class
Class II
Product code
OLO
Recalls
1
Predicates
2

Its longest line of descent runs back 10 generations to K010612 (2001). 24 ancestors in all, the oldest K983831 (1999). 1 later device descends from it, 1 citing it directly.

  1. Root · 10 generations backK010612 Vectorvision KneeBrainlab AG · 2001
  2. 6 generations between
  3. 3 generations backK133759 Intellijoint HipIntellijoint Surgical, Inc. · 2014
  4. 2 generations backK151364 Intellijoint HIP (tm) SystemIntellijoint Surgical, Inc. · 2015
  5. 1 generation backK162364 Intellijoint HIP Generation 2A SystemIntellijoint Surgical, Inc. · 2017
  6. This deviceK172849 Intellijoint HIP System (Cart), Intellijoint HIP System (Portable)…Intellijoint Surgical, Inc. · 2017
  7. Descendant · cites it directly · 0 of its ownK203751 OMNIVision systemCorin USA · 2021
Show all 10 generations
  1. Root · 10 generations backK010612 Vectorvision KneeBrainlab AG · 2001
  2. 9 generations backK021306 Vectorvision CT-Free KneeBrainlab AG · 2002
  3. 8 generations backK040368 Vectorvision Hip 3.0Brainlab AG · 2004
  4. 7 generations backK052213 Navigation SW Hip 3.1 on CIBrainlab AG · 2005
  5. 6 generations backK060468 Vectorvision HipBrainlab AG · 2006
  6. 5 generations backK083483 Brainlab Hip Unlimited, CI Hip UnlimitedBrainlab AG · 2009
  7. 4 generations backK110021 Dash HipBrainlab AG · 2011
  8. 3 generations backK133759 Intellijoint HipIntellijoint Surgical, Inc. · 2014
  9. 2 generations backK151364 Intellijoint HIP (tm) SystemIntellijoint Surgical, Inc. · 2015
  10. 1 generation backK162364 Intellijoint HIP Generation 2A SystemIntellijoint Surgical, Inc. · 2017
  11. This deviceK172849 Intellijoint HIP System (Cart), Intellijoint HIP System (Portable)…Intellijoint Surgical, Inc. · 2017

Most influential descendants

Showing 1 of 1 · See all
  • K203751OMNIVision systemCorin USA · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K010612Vectorvision KneeBrainlab AG2001—
−9K021306Vectorvision CT-Free KneeBrainlab AG2002—
−8K040368Vectorvision Hip 3.0Brainlab AG2004—
−7K052213Navigation SW Hip 3.1 on CIBrainlab AG2005—
−6K060468Vectorvision HipBrainlab AG2006—
−5K083483Brainlab Hip Unlimited, CI Hip UnlimitedBrainlab AG2009—
−4K110021Dash HipBrainlab AG2011—
−3K133759Intellijoint HipIntellijoint Surgical, Inc.2014—
−2K151364Intellijoint HIP (tm) SystemIntellijoint Surgical, Inc.2015—
−1K162364Intellijoint HIP Generation 2A SystemIntellijoint Surgical, Inc.2017—
0 (this device)K172849Intellijoint HIP System (Cart), Intellijoint HIP System (Portable)…Intellijoint Surgical, Inc.2017—
+1K203751OMNIVision systemCorin USA2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199910
200020
200120
200230
200320
200410
200620
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
FIG. 1 — Lineage of K172849, 2001–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OLO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OLO device

  1. #1 K240662 Spine Guidance Software (version 5.0); Xia 3/…score 23.7 · 15 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K151364Intellijoint HIP (tm) System
Summary page 4
…ssisted Orthopedic Surgery System. 4. Predicate Device Information 510(k) No. Device Manufacturer K151364 Intellijoint HIP® Intellijoint Surgical Inc. Concurrence: 28-Dec-2015 Generatio…
Intellijoint Surgical, Inc.2015—
*K162364Intellijoint HIP Generation 2A System
Summary page 4
…K151364 Intellijoint HIP® Intellijoint Surgical Inc. Concurrence: 28-Dec-2015 Generation 2 System K162364 Intellijoint HIP® Intellijoint Surgical Inc. Concurrence: 2-Mar-2017 Generatio…
Intellijoint Surgical, Inc.2017—

Also cleared against K151364: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K203751OMNIVision systemCorin USA20210

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 28, 2019Class IIZ-2228-2019Nonconforming Material/ComponentApple has determined that, in a limited number of older generation 15-inch MacBook Pro Units, the battery may overheat and pose a fire safety risk. To date, Intellijoint has distributed, 15-inch MacBook Pros which are the Workstation…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 2 companies; a new generation was cleared every 1.4 years on median.
  • 4 of the 24 ancestors have a recall of any class.
  • Of its descendants, K203751 (Corin USA, 2021) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K151364.

How does its category compare?

All OLO devices

OLO Orthopedic Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · OLO

40in 2025

2009: 112010: 92011: 112012: 72013: 72014: 102015: 102016: 142017: 332018: 242019: 352020: 342021: 282022: 402023: 452024: 442025: 402026 (partial year): 42
2007high 45 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OLO

119in 2025

2009: 126 days median2010: 113 days median2011: 138 days median2012: 129 days median2013: 137 days median2014: 119 days median2015: 176 days median2016: 120 days median2017: 91 days median2018: 97 days median2019: 95 days median2020: 89 days median2021: 91 days median2022: 89 days median2023: 90 days median2024: 90 days median2025: 119 days median2026 (partial year): 109 days median
2007high 176 (2015) · low 89 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OLO

1in 2025

2010: 12011: 12012: 12014: 32015: 22016: 32017: 22018: 12019: 12020: 62022: 52023: 7, 1 Class I2024: 42025: 12026 (partial year): 6
2007high 7 (2023) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · OLO

3,306in 2025

2008: 12009: 42010: 42011: 102012: 602013: 362014: 2452015: 4662016: 7502017: 1,0762018: 1,2302019: 1,5852020: 2,7462021: 1,6962022: 1,5682023: 1,4932024: 2,3532025: 3,306
2007high 3,306 (2025) · low 0 (2007)2025

MAUDE reports naming product code OLO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—01
20091112604
2010911314
201111138110
20127129160
20137137036
2014101193245
2015101762466
2016141203750
2017339121,076
2018249711,230
2019359511,585
2020348962,746
2021289101,696
2022408951,568
202345907 (1)1,493
2024449042,353
20254011913,306
2026421096—

What is it indicated for?

The Intellijoint HIP Generation 2C System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The system is only compatible with straight acetabular cup impactors. The Intellijoint HIP Generation 2C System is indicated for patients undergoing orthopedic…

Read the full statement (115 words)

The Intellijoint HIP Generation 2C System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The system is only compatible with straight acetabular cup impactors.

The Intellijoint HIP Generation 2C System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective. The system aids the surgeon in performing intra-operative measurements including measurements of limb position, joint center-of-rotation, and implant component positioning.

Example orthopedic surgical procedures include, but are not limited to:

- Total Hip Arthroplasty

- Minimally Invasive Hip Arthroplasty

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172849: Intellijoint HIP System (Cart), Intellijoint HIP System (Portable), Intellijoint HIP Instrument Tray.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172849

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026