# K173280: Neonatal ConchaSmart Breathing Circuits

- Pathway: 510(k) (Traditional)
- Applicant: Teleflexmedical, Inc.
- Decided: Aug 2, 2018 after 290 days
- Product code: BZE (Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer)
- Device class: 2
- Regulation: 21 CFR 868.5270
- Recalls of this device: 1 (1 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 6 generations back to K913368 (1992). 10 ancestors in all.
No later device cites it as a predicate yet.
1 of 11 devices in its family (counting itself) has a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K103767: Dual Heated Infant Vreathing Cirouits, Fisher & Paykel Healthcare, Ltd. (2012) Summary page 4.

## Recalls of this device
- May 10, 2019 · Class I · Z-1912-2019: Reported complaints of cracks being observed, prior to use, on swivel wye adaptors of the Neonatal ConchaSmart Breathing Circuit.

## Indications for use
The Neonatal ConchaSmart Breathing Circuits are intended for neonatal and infant patients in professional healthcare environments as a conduit for respiratory gas between a patient and a ventilator or Infant Flow Generator (for single limb configuration) and include heated wire(s) for use with the Hudson RCI Neptune Heated Humidifier. The heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.