510(k) Report

510(k) · Decided

Recalled · Class II · 2018

Elite HV Radial

K173287510(k) TraditionalDecided Dec 21, 201766-day reviewClass II21 CFR 870.1340Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Galt Medical Corp. · DYB Introducer, Catheter

Cleared on Dec 21, 2017 as substantially equivalent to K152173 (Terumo Medical Corporation, 2015) and 2 other predicates after a 66-day review. Its family tree reaches back 5 generations to 1995. 1 of the 16 devices in its family, counting itself, has a Class I recall. It has been recalled 2 times.

Decided
Dec 21, 2017
Review
66 days
Class
Class II
Product code
DYB
Recalls
2
Predicates
3

Its longest line of descent runs back 5 generations to K954234 (1995). 15 ancestors in all, the oldest K810675 (1981). No later device cites it as a predicate yet. 1 of the 16 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year cleared 2015: 1 other ancestor2005: 1 other ancestor2012: 1 other ancestor2000: 2 other ancestors1996: 1 other ancestor1995: 1 other ancestor1993: 1 other ancestor1991: 1 other ancestor1981: 1 other ancestor, 1 with a Class I recall1K954234 · Radifocus Introducer II Kit · Terumo Medical Corp. · 1995K033681 · Glidesheath · Terumo Corp. · 2003K062858 · Modification to Glidesheath · Terumo Medical Corp. · 2006K082644 · Glidesheath · Terumo Medical Corp. · 2008K152173 · Glidesheath · Terumo Medical Corporation · 20151980199020002010
Other ancestors, by generation 1 generation back: 2 other ancestors2 generations back: 3 other ancestors3 generations back: 1 other ancestor4 generations back: 4 other ancestors, 1 with a Class I recall1 K152173 · Glidesheath · Terumo Medical Corporation · 2015K082644 · Glidesheath · Terumo Medical Corp. · 2008K062858 · Modification to Glidesheath · Terumo Medical Corp. · 2006K033681 · Glidesheath · Terumo Corp. · 2003K954234 · Radifocus Introducer II Kit · Terumo Medical Corp. · 1995−50
  1. Root · 5 generations backK954234 Radifocus Introducer II KitTerumo Medical Corp. · 1995
  2. 4 generations backK033681 GlidesheathTerumo Corp. · 2003
  3. 3 generations backK062858 Modification to GlidesheathTerumo Medical Corp. · 2006
  4. 2 generations backK082644 GlidesheathTerumo Medical Corp. · 2008
  5. 1 generation backK152173 GlidesheathTerumo Medical Corporation · 2015
  6. This deviceK173287 Elite HV RadialGalt Medical Corp. · 2017

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K954234Radifocus Introducer II KitTerumo Medical Corp.1995—
−4K033681GlidesheathTerumo Corp.2003—
−3K062858Modification to GlidesheathTerumo Medical Corp.2006—
−2K082644GlidesheathTerumo Medical Corp.2008—
−1K152173GlidesheathTerumo Medical Corporation2015—
0 (this device)K173287Elite HV RadialGalt Medical Corp.2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198111
199110
199310
199510
199610
200020
200510
201210
201510
FIG. 1 — Lineage of K173287, 1995–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DYB device

  1. #1 K251051 AuST Steerable Sheathscore 18.0 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K152173Glidesheath
Summary page 4
…Catheter Introducer (DYB), 21 CFR 870.1340 Device Classification: Class II (Cardiovascular) Predicate Device: Glidesheath (K152173) Elite HV (K043525 Gateway Hemostasis Valve (152528) Manufacturer:…
Terumo Medical Corporation2015—
*K043525Catheter Introducer
Summary page 4
…21 CFR 870.1340 Device Classification: Class II (Cardiovascular) Predicate Device: Glidesheath (K152173) Elite HV (K043525 Gateway Hemostasis Valve (152528) Manufacturer: Galt Medical…
Galt Medical Corp.20053
*K152528Attachable Cath Lab Hemostasis Valve
Summary page 5
…cock and dilator), which are packaged together with a guide wire. The Elite HV product is configured with a silicone valve or a TPE valve cleared under 510(k) K152528. Some product configurations also include an entry needle, guide wire inserter and a flushing syringe. The…
Galt Medical Corp.2015—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 2, 2018Class IIZ-2562-2018OtherThe products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce…
May 2, 2018Class IIZ-2558-2018OtherThe products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K810675Arrow Radial Artery Catheterization SetArrow Intl., Inc. · 1981 · 4 generations back · ◆ in the 1981 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 2.9 years on median.
  • 5 of the 15 ancestors have a recall of any class; 1 has a Class I recall.

How does its category compare?

All DYB devices

DYB Introducer, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · DYB

16in 2025

2007: 122008: 172009: 142010: 182011: 172012: 222013: 262014: 302015: 172016: 192017: 242018: 192019: 182020: 92021: 122022: 132023: 152024: 182025: 162026 (partial year): 17
2007high 30 (2014) · low 9 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DYB

63in 2025

2007: 95 days median2008: 37 days median2009: 52 days median2010: 123 days median2011: 100 days median2012: 128 days median2013: 118 days median2014: 122 days median2015: 129 days median2016: 95 days median2017: 70 days median2018: 156 days median2019: 47 days median2020: 49 days median2021: 78 days median2022: 143 days median2023: 55 days median2024: 56 days median2025: 63 days median2026 (partial year): 59 days median
2007high 156 (2018) · low 37 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DYB

3in 2025

2007: 152008: 92009: 252010: 182011: 42012: 32013: 5, 1 Class I2014: 22015: 2, 1 Class I2016: 45, 3 Class I2017: 22, 1 Class I2018: 12019: 22020: 21, 21 Class I2021: 62022: 22023: 102024: 22025: 32026 (partial year): 23, 21 Class I
2007high 45 (2016) · low 1 (2018)2026

Class I (48 since 2007) Class II/III not classified

Adverse event reports per year · DYB

4,539in 2025

2007: 3162008: 4232009: 3992010: 4452011: 6282012: 8982013: 7432014: 7802015: 9132016: 1,2342017: 1,6722018: 1,9032019: 2,4502020: 2,4652021: 2,8352022: 4,0842023: 3,7752024: 3,9312025: 4,539
2007high 4,539 (2025) · low 316 (2007)2025

MAUDE reports naming product code DYB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007129515316
200817379423
2009145225399
20101812318445
2011171004628
2012221283898
2013261185 (1)743
2014301222780
2015171292 (1)913
2016199545 (3)1,234
2017247022 (1)1,672
20181915611,903
2019184722,450
202094921 (21)2,465
2021127862,835
20221314324,084
20231555103,775
2024185623,931
2025166334,539
2026175923 (21)—

What is it indicated for?

The Galt Medical Corp Elite HV Radial is indicated to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K173287: Elite HV Radial.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173287

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026