# K173477: ChemoCLAVE Cytotoxic Medication Preparation and Delivery System

- Pathway: 510(k) (Traditional)
- Applicant: Icu Medical, Inc.
- Decided: Aug 31, 2018 after 291 days
- Product code: ONB (Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System)
- Device class: 2
- Regulation: 21 CFR 880.5440
- Recalls of this device: 9 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 4 generations back to K980580 (1998). 18 ancestors in all.
No later device cites it as a predicate yet.
2 of 19 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K081361: Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXX, Icu Medical, Inc. (2008) Summary page 4.

## Recalls of this device
- May 12, 2026 · Class II · Z-3159-2026: Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber 
- May 12, 2026 · Class II · Z-3161-2026: Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber 
- Mar 16, 2021 · Class II · Z-1578-2021: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Mar 16, 2021 · Class II · Z-1579-2021: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Mar 16, 2021 · Class II · Z-1580-2021: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Mar 16, 2021 · Class II · Z-1576-2021: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Mar 16, 2021 · Class II · Z-1577-2021: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Aug 31, 2020 · Class II · Z-0704-2021: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous me
- Jul 6, 2020 · Class II · Z-0248-2021: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

## Indications for use
The ChemoClave is a needle-free Closed System Transfer Device (CSTD) that mechanically prohibits the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system during drug preparation and administration, thereby minimizing exposure of individuals, healthcare personnel and the environment to hazardous drugs.

Source page: https://510kreport.com/device/K173477