510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

GYN-Pump PH304

K173489510(k) TraditionalCleared Mar 28, 2018135-day reviewClass II21 CFR 884.1700Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

W.O.M. World of Medicine GmbH · HIG Insufflator, Hysteroscopic

Cleared on Mar 28, 2018 as substantially equivalent to K112642 (W.O.M. World of Medicine AG, 2012) and 1 other predicate after a 135-day review. Its family tree reaches back 7 generations to 1996. It has been recalled once.

Decided
Mar 28, 2018
Review
135 days
Class
Class II
Product code
HIG
Recalls
1
Predicates
2

Its longest line of descent runs back 7 generations to K950035 (1996). 29 ancestors in all, the oldest K903419 (1990). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2017: 1 other ancestor2008: 1 other ancestor2003: 3 other ancestors2001: 1 other ancestor1999: 1 other ancestor1995: 3 other ancestors2007: 1 other ancestor2006: 1 other ancestor2000: 1 other ancestor1998: 1 other ancestor1996: 3 other ancestors1994: 2 other ancestors1991: 2 other ancestors1990: 1 other ancestorK950035 · Surgiflator 20 · W.O.M. GmbH · 1996K955791 · Surgiflator-20 Pim · W.O.M. World of Medicine GmbH · 1996K981859 · Surgiflator-20 Ecu · W.O.M. World of Medicine GmbH · 1998K990443 · Arthro-Surgimat-2000 Ecu · W.O.M. World of Medicine GmbH · 1999K030402 · Arthroscopy Pump, Model A107 · W.O.M. World of Medicine AG · 2003K062380 · Flosimple Arthroscopy Pump A120 · W.O.M. World of Medicine AG · 2006 · recalledK112642 · Aquilex Fluid Control System H112 · W.O.M. World of Medicine AG · 20121990199520002005201020152020
  1. Root · 7 generations backK950035 Surgiflator 20W.O.M. GmbH · 1996
  2. 6 generations backK955791 Surgiflator-20 PimW.O.M. World of Medicine GmbH · 1996
  3. 5 generations backK981859 Surgiflator-20 EcuW.O.M. World of Medicine GmbH · 1998
  4. 4 generations backK990443 Arthro-Surgimat-2000 EcuW.O.M. World of Medicine GmbH · 1999
  5. 3 generations backK030402 Arthroscopy Pump, Model A107W.O.M. World of Medicine AG · 2003
  6. 2 generations backK062380 Flosimple Arthroscopy Pump A120W.O.M. World of Medicine AG · 2006 · Recalled
  7. 1 generation backK112642 Aquilex Fluid Control System H112W.O.M. World of Medicine AG · 2012
  8. This deviceK173489 GYN-Pump PH304W.O.M. World of Medicine GmbH · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K950035Surgiflator 20W.O.M. GmbH1996—
−6K955791Surgiflator-20 PimW.O.M. World of Medicine GmbH1996—
−5K981859Surgiflator-20 EcuW.O.M. World of Medicine GmbH1998—
−4K990443Arthro-Surgimat-2000 EcuW.O.M. World of Medicine GmbH1999—
−3K030402Arthroscopy Pump, Model A107W.O.M. World of Medicine AG2003—
−2K062380Flosimple Arthroscopy Pump A120W.O.M. World of Medicine AG20061
−1K112642Aquilex Fluid Control System H112W.O.M. World of Medicine AG2012—
0 (this device)K173489GYN-Pump PH304W.O.M. World of Medicine GmbH2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199120
199420
199530
199630
199810
199910
200010
200110
200330
200610
200710
200810
201710
FIG. 1 — Lineage of K173489, 1996–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HIG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HIG device

  1. #1 K261706 Endomat Select (UP220)score 14.1 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112642Aquilex Fluid Control System H112
Summary page 5
…rket Notification 3. Primary Predicate Device Trade Name: Aquilex Fluid Control System H112 510(k) Number: K112642 Classification Name: Hysteroscopic Insufflator Regulation Number: 21 C.F.R. § 884.1…
W.O.M. World of Medicine AG2012—
*K163320LAP-Pump PP110
Summary page 5
…subject to a design related recall. 4. Secondary Predicate Device Trade Name: LAP-Pump PP110 510(k) Number: K163320 Classification Name: Gynecologic laparoscope and accessories Regulation Number: 21…
W.O.M. World of Medicine GmbH2017—

Also cleared against K112642: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 18, 2021Class IIZ-2561-2021Software designWhen the display of inflow volume to the uterus reaches 32450 ml, the fluid management system control unit display will freeze because the internal software calculation threshold is reached. The outflow measurement will continue, and so…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 3 companies; a new generation was cleared every 3.7 years on median.
  • 3 of the 29 ancestors have a recall of any class.
  • 4 other devices also cite its predicate K112642.

How does its category compare?

All HIG devices

HIG Insufflator, Hysteroscopic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · HIG

1in 2024

2010: 12011: 12012: 12013: 12014: 12017: 12018: 42020: 12021: 12024: 12026 (partial year): 2
2007high 4 (2018) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HIG

175in 2018

2018: 175 days median
2007high 175 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HIG

1in 2007

2007: 12026 (partial year): 1
2007high 1 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HIG

45in 2025

2008: 22009: 32010: 12011: 42012: 42013: 52014: 92015: 102016: 102017: 112018: 102019: 662020: 312021: 1302022: 662023: 332024: 452025: 45
2007high 130 (2021) · low 0 (2007)2025

MAUDE reports naming product code HIG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20080—02
20090—03
20101—01
20111—04
20121—04
20131—05
20141—09
20150—010
20160—010
20171—011
20184175010
20190—066
20201—031
20211—0130
20220—066
20230—033
20241—045
20250—045
20262—1—

What is it indicated for?

The GYN-Pump PH304 is a combined suction and irrigation pump for use in hysteroscopic and laparoscopic interventions. During diagnostic and operative hysteroscopy, the GYN-Pump PH304 is intended to provide liquid distension of the uterus and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus. During diagnostic and therapeutic laparoscopic procedures, the GYN-Pump PH304 is intended to irrigate fluid into and remove fluid from the abdominal cavity.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K173489: GYN-Pump PH304.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173489

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026