# K173607: SOMATOM Edge Plus

- Pathway: 510(k) (Traditional)
- Applicant: Siemens Medical Solutions USA, Inc.
- Decided: Mar 21, 2018 after 120 days
- Product code: JAK (System, X-Ray, Tomography, Computed)
- Device class: 2
- Regulation: 21 CFR 892.1750
- Recalls of this device: 3 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 8 generations back to K982349 (1998). 11 ancestors in all.
20 later devices descend from it; 1 cites it directly.
0 of 32 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K152036: SOMATOM Definition Edge, SOMATOM Definition AS/AS+, Siemens Medical Solutions USA, Inc. (2015) Summary page 5.

## Most influential descendants
- K190578: SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM Definition AS Open, SOMATOM Edge Plus, SOMATOM Definition AS/AS+, SOMATOM Confidence (2019), 19 descendants of its own
- K192061: SOMATOM Go Platform – SOMATOM go.Now, go.Up, go.All, go.Top, go.Sim, go.Open Pro, Scan & GO (2019), 16 descendants of its own
- K191891: SOMATOM X.cite (2019), 15 descendants of its own
- K200524: SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All, go. Sim, go. Open Pro, Scan&GO (2020), 14 descendants of its own
- K211373: SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X. Platform Scanners - SOMATOM X.cite and SOMATOM X.ceed, Scan&GO Software (2021), 12 descendants of its own

## Recalls of this device
- Nov 24, 2020 · Class II · Z-0771-2021: System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan
- Jul 3, 2019 · Class II · Z-2273-2019: Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.
- Feb 14, 2019 · Class II · Z-1246-2019: Potential risk of scans being aborted when using the optional Dual Spiral Dual Engery mode due to a software issue found in syngo CT VB10A for the Siemens Healthineers CT scanners with a single X-ray 

## Indications for use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011;365:395-409) and subsequent literature, for further information.

Source page: https://510kreport.com/device/K173607