510(k) · Cleared
ClearedPF Nitrile Exam Glove with pH coating, Black-Gray Color, Tested for Use with Chemo Drugs
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Shen Wei USA, Inc. · LZC Medical Glove, Specialty
Cleared on Jul 8, 2018 after a 209-day review. FDA’s file for it is a 510(k) statement, which names no predicate.
- Decided
- Jul 8, 2018
- Review
- 209 days
- Class
- Class I
- Product code
- LZC
- Recalls
- None
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LZC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LZC device
- #1 K201390 Medline Powder-Free Light Blue Nitrile Exam…score 19.2 · 5 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
How does its category compare?
All LZC devicesLZC Medical Glove, Specialty: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.
FDA decisions per year · LZC
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LZC
85in 2022
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LZC
1in 2021
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LZC
10in 2025
MAUDE reports naming product code LZC, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 0 | 0 |
| 2008 | 3 | 166 | 1 | 0 |
| 2009 | 5 | 253 | 0 | 1 |
| 2010 | 5 | 68 | 0 | 0 |
| 2011 | 0 | — | 4 | 1 |
| 2012 | 4 | 149 | 0 | 1 |
| 2013 | 0 | — | 0 | 1 |
| 2014 | 0 | — | 0 | 3 |
| 2015 | 1 | — | 0 | 0 |
| 2016 | 3 | 258 | 0 | 5 |
| 2017 | 6 | 121 | 0 | 7 |
| 2018 | 6 | 105 | 0 | 6 |
| 2019 | 7 | 200 | 0 | 8 |
| 2020 | 3 | 167 | 0 | 1 |
| 2021 | 3 | 94 | 1 | 2 |
| 2022 | 7 | 85 | 0 | 2 |
| 2023 | 0 | — | 0 | 1 |
| 2024 | 0 | — | 0 | 16 |
| 2025 | 1 | — | 0 | 10 |
| 2026 | 1 | — | 1 | — |
What is it indicated for?
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. This device is single use only. Chemotherapy Drug and Concentration Breakthrough Detection…
Read the full statement (142 words)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. This device is single use only.
Chemotherapy Drug and Concentration Breakthrough Detection Time in Minutes
1. Carmustine (BCNU), 3.3mg/ml 27.1
2. Cisplatin, 1.0mg/ml >240
3. Cyclophosphamide (Cytoxan), 20mg/ml >240
4. Dacarbazine (DTIC) 10g mg/ml >240
5. Doxorubicin Hydrochloride, 2.0 mg/ml >240
6. Etoposide (Toposar), 20mg/ml >240
7. Fluorouracil 50mg/ml >240
8. Paclitaxel (Taxol), 6.0mg/ml >240
9. Thiotepa, 10mg/ml 274
*Please note that Carmustine (BCNU) has low permeation time of 27.1 minutes and Thiotepa has permeation time of 27.4 minutes.
WARNING: Do not use with Carmustine and Thiotepa.
Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K173766 · printed Nov 6, 2026
Cite this page
510(k) Report. “K173766: PF Nitrile Exam Glove with pH coating, Black-Gray Color, Tested for Use with Chemo Drugs.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173766
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026