St. Jude Medical (Now Part of Abbott Medical) · DXO Transducer, Pressure, Catheter Tip

Cleared on Mar 28, 2018 as substantially equivalent to K161171 (St. Jude Medical, Inc., 2016) after a 27-day review. Its family tree reaches back 11 generations to 2001.

Decided
Mar 28, 2018
Review
27 days
Class
Class II
Product code
DXO
Recalls
None
Predicates
1

Its longest line of descent runs back 11 generations to K003122 (2001). 21 ancestors in all, the oldest K883651 (1988). No later device cites it as a predicate yet.

  1. Root · 11 generations backK003122 Smartflow, Model Sfip 2000Florence Medical, Ltd. · 2001
  2. 7 generations between
  3. 3 generations backK131452 Pressurewire; Certus, Certus 300, Aeris, Aeris 300St. Jude Medical Systems AB · 2013
  4. 2 generations backK140466 Pressurewire: Aeris Agile Tip, Aeris Agile Tip 300, Certus Agile…St. Jude Medical Systems AB · 2014
  5. 1 generation backK161171 PressureWire X GuidewireSt. Jude Medical, Inc. · 2016
  6. This deviceK180558 PressureWire XSt. Jude Medical (Now Part of Abbott Medical) · 2018
Show all 11 generations
  1. Root · 11 generations backK003122 Smartflow, Model Sfip 2000Florence Medical, Ltd. · 2001
  2. 10 generations backK012947 SmartflowFlorence Medical, Ltd. · 2001
  3. 9 generations backK020127 Smartflow Pulse Transmission CoefficientFlorence Medical, Ltd. · 2002
  4. 8 generations backK022188 Radianalyzer, Model 12710Radi Medical Systems AB · 2002
  5. 7 generations backK031662 Pressure Wire Sensor, Models 12004/12304Radi Medical Systems AB · 2003
  6. 6 generations backK062769 Pressurewire Certus, Models 12006 and 12306Radi Medical Systems AB · 2006
  7. 5 generations backK080813 Pressurewire Certus, Certus (300CM), Aeris, Aeris (300CM), ReceiverRadi Medical Systems AB · 2008
  8. 4 generations backK113584 Pressurewire Certus, Certus 300, Aeris, Aeris 300, and ReceiverSt Jude Medical · 2012
  9. 3 generations backK131452 Pressurewire; Certus, Certus 300, Aeris, Aeris 300St. Jude Medical Systems AB · 2013
  10. 2 generations backK140466 Pressurewire: Aeris Agile Tip, Aeris Agile Tip 300, Certus Agile…St. Jude Medical Systems AB · 2014
  11. 1 generation backK161171 PressureWire X GuidewireSt. Jude Medical, Inc. · 2016
  12. This deviceK180558 PressureWire XSt. Jude Medical (Now Part of Abbott Medical) · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K003122Smartflow, Model Sfip 2000Florence Medical, Ltd.2001—
−10K012947SmartflowFlorence Medical, Ltd.2001—
−9K020127Smartflow Pulse Transmission CoefficientFlorence Medical, Ltd.2002—
−8K022188Radianalyzer, Model 12710Radi Medical Systems AB2002—
−7K031662Pressure Wire Sensor, Models 12004/12304Radi Medical Systems AB2003—
−6K062769Pressurewire Certus, Models 12006 and 12306Radi Medical Systems AB2006—
−5K080813Pressurewire Certus, Certus (300CM), Aeris, Aeris (300CM), ReceiverRadi Medical Systems AB2008—
−4K113584Pressurewire Certus, Certus 300, Aeris, Aeris 300, and ReceiverSt Jude Medical2012—
−3K131452Pressurewire; Certus, Certus 300, Aeris, Aeris 300St. Jude Medical Systems AB2013—
−2K140466Pressurewire: Aeris Agile Tip, Aeris Agile Tip 300, Certus Agile Tip, Certus…St. Jude Medical Systems AB2014—
−1K161171PressureWire X GuidewireSt. Jude Medical, Inc.2016—
0 (this device)K180558PressureWire XSt. Jude Medical (Now Part of Abbott Medical)2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199810
200130
200210
200320
200410
200510
FIG. 1 — Lineage of K180558, 2001–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXO device

  1. #1 K241418 OptoMonitor 3score 14.8 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K161171PressureWire X Guidewire
Summary page 4
…870.2910 – Transmitters and Receivers, Physiological Signal, Radiofrequency Device Class: II Predicate Device: PressureWire™ X (K161171 ) The PressureWire™ X guidewire has an integrated sensor element at the tip to enable…
St. Jude Medical, Inc.2016—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 6 companies; a new generation was cleared every 1.5 years on median.
  • None of the 21 ancestors has a recall.

How does its category compare?

All DXO devices

DXO Transducer, Pressure, Catheter Tip: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · DXO

1in 2025

2007: 12008: 12010: 12011: 12012: 12013: 12014: 32015: 22016: 22017: 22018: 12019: 42020: 22022: 12024: 12025: 12026 (partial year): 1
2007high 4 (2019) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXO

120in 2019

2014: 176 days median2019: 120 days median
2007high 176 (2014) · low 120 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXO

1in 2015

2015: 1
2007high 1 (2015) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXO

183in 2025

2007: 32008: 22009: 82010: 742011: 1112012: 822013: 932014: 852015: 1142016: 1522017: 1862018: 1792019: 1532020: 1582021: 1162022: 1442023: 1942024: 1972025: 183
2007high 197 (2024) · low 2 (2008)2025

MAUDE reports naming product code DXO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—03
20081—02
20090—08
20101—074
20111—0111
20121—082
20131—093
20143176085
20152—1114
20162—0152
20172—0186
20181—0179
201941200153
20202—0158
20210—0116
20221—0144
20230—0194
20241—0197
20251—0183
20261—0—

What is it indicated for?

The PressureWire™ X guidewire is indicated to direct a catheter through a blood vessel and to measure physiological parameters in the heart and in the coronary and peripheral blood vessels. Physiological parameters include blood pressure. The PressureWire™ X guidewire can also measure blood temperature.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K180558: PressureWire X.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K180558

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026