510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

GammaPod

K180571510(k) SpecialCleared Apr 4, 201830-day reviewClass II21 CFR 892.5750Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Xcision Medical Systems, LLC · IWB System, Radiation Therapy, Radionuclide

Cleared on Apr 4, 2018 as substantially equivalent to K172706 (Xcision Medical Systems, LLC, 2017) after a 30-day review. Its family tree reaches back 3 generations to 1997. It has been recalled 2 times.

Decided
Apr 4, 2018
Review
30 days
Class
Class II
Product code
IWB
Recalls
2
Predicates
1

Its longest line of descent runs back 3 generations to K970647 (1997). No later device cites it as a predicate yet.

  1. Root · 3 generations backK970647 Our Rotating Gamma SystemOur Scientific Intl., Inc. · 1997
  2. 2 generations backK060314 Rotating Gamma System. Gamma Art-6000American Radiosurgery, Inc. · 2006
  3. 1 generation backK172706 GammaPod - Model AXcision Medical Systems, LLC · 2017 · Recalled
  4. This deviceK180571 GammaPodXcision Medical Systems, LLC · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K970647Our Rotating Gamma SystemOur Scientific Intl., Inc.1997—
−2K060314Rotating Gamma System. Gamma Art-6000American Radiosurgery, Inc.2006—
−1K172706GammaPod - Model AXcision Medical Systems, LLC20171
0 (this device)K180571GammaPodXcision Medical Systems, LLC2018—
FIG. 1 — Lineage of K180571, 1997–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IWB device

  1. #1 K250391 Leksell Gamma Knife® (Elekta Esprit); Leksell…score 11.2 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K172706GammaPod - Model A
Summary page 4
…IWB Device Classification: 21 CFR 892.5750, Class II III. PREDICATE DEVICE IDENTIFICATION Predicate Device: GammaPod (K172706) IV. DEVICE DESCRIPTION The GammaPod™ is a teletherapy device that uses rotating, multi-source Cobalt-60 gamm…
Xcision Medical Systems, LLC20171

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 30, 2023Class IIZ-0750-2024Software designWrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected
Jul 3, 2019Class IIZ-2081-2019Nonconforming Material/ComponentOne bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 8.9 years on median.
  • 1 of the 3 ancestors has a recall of any class.

How does its category compare?

All IWB devices

IWB System, Radiation Therapy, Radionuclide: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · IWB

1in 2025

2007: 12009: 12011: 12012: 12014: 32015: 32016: 12017: 12018: 32019: 12020: 12021: 22022: 12024: 12025: 1
2007high 3 (2014) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IWB

57in 2018

2014: 170 days median2015: 79 days median2018: 57 days median
2007high 170 (2014) · low 57 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IWB

1in 2023

2007: 32008: 32009: 12010: 12011: 22014: 12023: 1
2007high 3 (2007) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IWB

3in 2025

2007: 82008: 152009: 82010: 92011: 52012: 32013: 22014: 82015: 52016: 72017: 42018: 22019: 32020: 22023: 12025: 3
2007high 15 (2008) · low 0 (2021)2025

MAUDE reports naming product code IWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—38
20080—315
20091—18
20100—19
20111—25
20121—03
20130—02
2014317018
201537905
20161—07
20171—04
201835702
20191—03
20201—02
20212—00
20221—00
20230—11
20241—00
20251—03
20260—0—

What is it indicated for?

GammaPod™ is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K180571: GammaPod.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K180571

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026