510(k) Report

510(k) · Cleared

Cleared

BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use, BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx

K180878510(k) TraditionalCleared Dec 17, 2018258-day reviewClass II21 CFR 862.3610

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shanghai Venture Bio-Tech Co., Ltd. · DJC Thin Layer Chromatography, Methamphetamine

Cleared on Dec 17, 2018 as substantially equivalent to K173303 (Alfa Scientific Designs, Inc., 2018) after a 258-day review. Its family tree reaches back 7 generations to 2000.

Decided
Dec 17, 2018
Review
258 days
Class
Class II
Product code
DJC
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K001178 (2000). 17 ancestors in all, the oldest K971218 (1997). No later device cites it as a predicate yet.

  1. Root · 7 generations backK001178 Cedia Dau 6-Acetylmorphine AssayMicrogenics Corp. · 2000
  2. 6 generations backK040316 Cedia Buprenorphine AssayMicrogenics Corp. · 2004
  3. 5 generations backK042988 Rapid One Step Buprenorphine Test CardTianjin New Bay Bioresearch Co., Ltd. · 2005
  4. 4 generations backK060527 Instant-View Bup/Nbup Urine Test. Instant-Verdict Bup/Nbup Urine…Alfa Scientific Designs, Inc. · 2006
  5. 3 generations backK063545 Multi-Drug of Abuse Urine Test,Urine Cup Test, Amphetamine Urine…Alfa Scientific Designs, Inc. · 2007
  6. 2 generations backK152122 Instant-View Multi-Drug Urine Test Cup (Home Use), Instant-View…Alfa Scientific Designs, Inc. · 2016
  7. 1 generation backK173303 INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC…Alfa Scientific Designs, Inc. · 2018
  8. This deviceK180878 BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE…Shanghai Venture Bio-Tech Co., Ltd. · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K001178Cedia Dau 6-Acetylmorphine AssayMicrogenics Corp.2000—
−6K040316Cedia Buprenorphine AssayMicrogenics Corp.2004—
−5K042988Rapid One Step Buprenorphine Test CardTianjin New Bay Bioresearch Co., Ltd.2005—
−4K060527Instant-View Bup/Nbup Urine Test. Instant-Verdict Bup/Nbup Urine Test…Alfa Scientific Designs, Inc.2006—
−3K063545Multi-Drug of Abuse Urine Test,Urine Cup Test, Amphetamine Urine Test…Alfa Scientific Designs, Inc.2007—
−2K152122Instant-View Multi-Drug Urine Test Cup (Home Use), Instant-View Multi-Drug…Alfa Scientific Designs, Inc.2016—
−1K173303INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use)…Alfa Scientific Designs, Inc.2018—
0 (this device)K180878BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid…Shanghai Venture Bio-Tech Co., Ltd.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199720
200020
200110
200240
200410
FIG. 1 — Lineage of K180878, 2000–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DJC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DJC device

  1. #1 K253705 Healgen® Accurate Oral Fluid Drug Testscore 12.2 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K173303INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use), INSTANT-VIEW…
Summary page 6
…Cannabinoid Test System DNK, NGI 21 CFR, 862.3640 Toxicology Morphine test system 5.3 Predicate Device K173303 INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use) INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (P…
Alfa Scientific Designs, Inc.2018—

Also cleared against K173303: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 1.2 years on median.
  • None of the 17 ancestors has a recall.
  • 2 other devices also cite its predicate K173303.

How does its category compare?

All DJC devices

DJC Thin Layer Chromatography, Methamphetamine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · DJC

1in 2025

2008: 12013: 12016: 12018: 32025: 12026 (partial year): 1
2007high 3 (2018) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DJC

258in 2018

2018: 258 days median
2007high 258 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DJC

No recall names DJC since 2007.

Adverse event reports per year · DJC

2in 2018

2017: 12018: 2
2007high 2 (2018) · low 0 (2007)2025

MAUDE reports naming product code DJC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—00
20090—00
20100—00
20110—00
20120—00
20131—00
20140—00
20150—00
20161—00
20170—01
2018325802
20190—00
20200—00
20210—00
20220—00
20230—00
20240—00
20251—00
20261—0—

What is it indicated for?

BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, Oxazepam, Benzoylecgonine…

Read the full statement (392 words)

BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, Oxazepam, Benzoylecgonine, d-Methamphetamine, Morphine and Delta-9-THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

Test Calibrators Cut-off Level

Amphetamine (AMP) d-Amphetamine 1000ng/ml

Oxazepam Oxazepam 300ng/ml

Cocaine (COC) Benzoylecgonine 300ng/ml

Methamphetamine (MET) d-Methamphetamine 1000ng/ml

Morphine(MOP) Morphine 300ng/ml

Marijuana (THC) Delta-9-THC-COOH 50ng/ml

The cassette test may be configured as single drug test in any drug analytes listed in the table.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method.

Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over the counter use.

BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Morphine Test Cassette for Rx Use, BIO-VENTURE Rapid Marijuana Test Cassette for Rx Use is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, Oxazepam, Benzoylecgonine, d-Methamphetamine, Morphine and Delta-9-THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

Test Calibrators Cut-off Level

Amphetamine (AMP) d-Amphetamine 1000ng/ml

Oxazepam Oxazepam 300ng/ml

Cocaine (COC) Benzoylecgonine 300ng/ml

Methamphetamine (MET) d-Methamphetamine 1000ng/ml

Morphine(MOP) Morphine 300ng/ml

Marijuana (THC) Delta-9-THC-COOH 50ng/ml

The cassette test may be configured as single drug test in any drug analytes listed in the table.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method.

Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for prescription use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K180878: BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use, BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K180878

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026