# K182063: VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides, VITROS Chemistry Products TBIL Slides, VITROS XT 7600 Integrated System

- Pathway: 510(k) (Traditional)
- Applicant: Ortho-Clinical Diagnostics, Inc.
- Decided: Oct 30, 2018 after 90 days
- Product code: JFY (Enzymatic Method, Creatinine)
- Device class: 2
- Regulation: 21 CFR 862.1225
- Recalls of this device: 20 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K001310 (2000). 8 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 10 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K160495: VITROS® Automation Solutions, Ortho-Clinical Diagnostics, Inc. (2016) Summary page 5.
- K063591: Vitros Chemistry Products Crea Slides,Model# 6802584, Vitros Chemistry Products Calibrator Kit 1,Model#1882208, Ortho-Clinical Diagnostics, Inc. (2006) Summary page 5.
- K840880: Ektachem Clinical Chem. Slides, Eastman Kodak Company (1984) Summary page 5.
- K081543: Vitros Immunodiagnostic Products TSH Reagent Pack and Calibrators, Vitros 5600 Integrated System, Ortho-Clinical Diagnostics, Inc. (2008) Summary page 5.

## Most influential descendants
- K190326: VITROS XT Chemistry Products UREA-CREA Slides (2019), 0 descendants of its own

## Recalls of this device
- May 9, 2025 · Class II · Z-2039-2025: a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a she
- Dec 19, 2024 · Class II · Z-1006-2025: The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the 
- Jan 26, 2024 · Class II · Z-1232-2024: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurabl
- Jan 26, 2024 · Class II · Z-1227-2024: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient res
- Jan 26, 2024 · Class II · Z-1228-2024: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient res
- Jan 26, 2024 · Class II · Z-1233-2024: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurabl
- Jan 22, 2024 · Class II · Z-1222-2024: Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems which may lead to erron
- Oct 10, 2023 · Class II · Z-0546-2024: During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred
- Oct 10, 2023 · Class II · Z-0547-2024: During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred
- Aug 9, 2023 · Class II · Z-0008-2024: Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as inten
- Mar 10, 2023 · Class II · Z-1555-2023: MicroTip Pack Opener Assembly Potentially Not Removing or Replacing MicroTip Pack Caps on VITROS¿ 5600 and XT 7600 Integrated Systems, results may be delayed results if a pack becomes unusable
- Jul 19, 2022 · Class II · Z-1676-2022: Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imag
- Jul 19, 2022 · Class II · Z-1675-2022: Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imag
- Mar 26, 2021 · Class II · Z-1489-2021: Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpec
- Jul 20, 2020 · Class II · Z-2879-2020: When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results.
- Jul 20, 2020 · Class II · Z-2878-2020: When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results.
- Jul 13, 2020 · Class II · Z-2873-2020: Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides
- Mar 20, 2019 · Class II · Z-1364-2019: Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported
- Mar 11, 2019 · Class II · Z-1222-2019: One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error c
- Feb 20, 2019 · Class II · Z-1291-2019: Luminometer Malfunction May Cause Inability to Process MicroWell Assays on VITROS¿ Systems

## Indications for use
1. VITROS Chemistry Products CRBM Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Products CRBM Slides quantitatively measure carbamazepine (CRBM) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Measurements obtained are used in monitoring levels of carbamazepine to help ensure appropriate therapy. 2. VITROS Chemistry Products CREA Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Product CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. 3. VITROS Chemistry Products TBIL Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Products TBIL Slides quantitatively measure total bilirubin (TBIL) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Measurements of the levels of bilirubin are used in the diagnosis and treatment of liver, hemolytic hematological and metabolic disorders, including hepatitis and gall bladder block. 4. VITROS XT 7600 Integrated System: Rx Only. For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest.