510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Fenom Pro Nitric Oxide Test

K182874510(k) TraditionalCleared Feb 13, 2019124-day reviewClass II21 CFR 862.3080Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spirosure, Inc. · MXA System, Test, Breath Nitric Oxide

Cleared on Feb 13, 2019 as substantially equivalent to K072816 (Aerocrine AB, 2008) after a 124-day review. Its family tree reaches back 2 generations to 2003. 2 later devices descend from it. It has been recalled once.

Decided
Feb 13, 2019
Review
124 days
Class
Class II
Product code
MXA
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to DEN030001 (2003). 2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2023: 1 descendant cleared2025: 1 descendant clearedDEN030001 · Niox Article NO: 02-1000 · Aerocrine AB · 2003K072816 · Niox Mino · Aerocrine AB · 2008200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant K072816 · Niox Mino · Aerocrine AB · 2008DEN030001 · Niox Article NO: 02-1000 · Aerocrine AB · 20030
  1. Root · 2 generations backDEN030001 Niox Article NO: 02-1000Aerocrine AB · 2003
  2. 1 generation backK072816 Niox MinoAerocrine AB · 2008
  3. This deviceK182874 Fenom Pro Nitric Oxide TestSpirosure, Inc. · 2019
  4. Descendant · cites it directly · 1 of its ownK213611 Fenom ProCaire Diagnostics, Inc. · 2023
  5. Descendant · 2 generations on · 0 of its ownK251674 Fenom Flo™ FeNO Monitoring SystemMgc Diagnostics Corporation · 2025

Most influential descendants

Showing 2 of 2 · See all
  • K213611Fenom ProCaire Diagnostics, Inc. · 2023 · cites it directly1descendant
  • K251674Fenom Flo™ FeNO Monitoring SystemMgc Diagnostics Corporation · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2DEN030001Niox Article NO: 02-1000Aerocrine AB2003—
−1K072816Niox MinoAerocrine AB2008—
0 (this device)K182874Fenom Pro Nitric Oxide TestSpirosure, Inc.2019—
+1K213611Fenom ProCaire Diagnostics, Inc.2023—
+2K251674Fenom Flo™ FeNO Monitoring SystemMgc Diagnostics Corporation2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202510
FIG. 1 — Lineage of K182874, 2003–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MXA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MXA device

  1. #1 K251674 Fenom Flo™ FeNO Monitoring Systemscore 11.3 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072816Niox Mino
Summary page 4
…Classification: 21 CFR Part 862.3080 807.92 (a)(3): Identification of the legally marketed predicate devices NIOX MINO (Aerocrine AB, Morrisville, NC), K072816 807.92 (a)(4): Device Description Fenom Pro™ is a point-of-care breath analyzer that uses solid-state electrochemical technology…
Aerocrine AB2008—

Also cleared against K072816: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K213611Fenom ProCaire Diagnostics, Inc.20231

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 26, 2021Class IIZ-1284-2021Process controlErroneously high FeNO levels which may contribute to premature discontinuation of effective asthma treatment and persistent underlying disease. Two issues affecting the FeNO readings: 1) Drift in the calibration gases and 2) Software…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 2 companies; a new generation was cleared every 10.9 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K213611 (Caire Diagnostics, Inc., 2023) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K072816.

How does its category compare?

All MXA devices

MXA System, Test, Breath Nitric Oxide: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · MXA

2in 2025

2008: 22009: 12010: 12012: 12014: 12015: 12017: 12019: 12021: 12023: 12024: 12025: 2
2007high 2 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MXA

Too few MXA 510(k)s in any year since 2007 to give a median.

Recalls per year · MXA

No recall names MXA since 2007.

Adverse event reports per year · MXA

No MAUDE report names MXA since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—00
20091—00
20101—00
20110—00
20121—00
20130—00
20141—00
20151—00
20160—00
20171—00
20180—00
20191—00
20200—00
20211—00
20220—00
20231—00
20241—00
20252—00
20260—0—

What is it indicated for?

Fenom Pro™ Nitric Oxide Test is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and often decreases in response to anti-inflammatory treatment. Measurement of FeNO by Fenom Pro™ is a method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological…

Read the full statement (138 words)

Fenom Pro™ Nitric Oxide Test is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and often decreases in response to anti-inflammatory treatment. Measurement of FeNO by Fenom Pro™ is a method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FeNO levels. FeNO measurements are to be used as an adjunct to established clinical assessments. Fenom Pro™ is suitable for children, approximately 7-17 years, and adults 18 years and older.

Testing using the Fenom Pro™ should only be done in a point-of-care healthcare setting under professional supervision. Fenom Pro™ should not be used in critical care, emergency care or in anesthesiology.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K182874: Fenom Pro Nitric Oxide Test.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K182874

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026