510(k) Report

510(k) · Cleared

Cleared

PLANET Onco Dose

K182966510(k) TraditionalCleared Mar 6, 2019131-day reviewClass II21 CFR 892.2050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dosisoft · LLZ System, Image Processing, Radiological

Cleared on Mar 6, 2019 as substantially equivalent to K173636 (Varian Medical Systems, Inc., 2018) after a 131-day review. Its family tree reaches back 13 generations to 1996. 1 later device descends from it.

Decided
Mar 6, 2019
Review
131 days
Class
Class II
Product code
LLZ
Recalls
None
Predicates
1

Its longest line of descent runs back 13 generations to K954874 (1996). 72 ancestors in all, the oldest K871859 (1987). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2011: 1 other ancestor2007: 2 other ancestors2004: 6 other ancestors2001: 7 other ancestors2006: 1 other ancestor2005: 2 other ancestors2003: 3 other ancestors2000: 5 other ancestors1997: 3 other ancestors2002: 3 other ancestors1998: 3 other ancestors1995: 4 other ancestors1993: 3 other ancestors1987: 1 other ancestor1999: 1 other ancestor1994: 2 other ancestors1992: 2 other ancestors1996: 3 other ancestors1991: 4 other ancestors1990: 1 other ancestor1989: 1 other ancestor2025: 1 descendant clearedK954874 · Spectef Protocol · GE Medical Systems · 1996K980914 · Emory Cardiac Tool Box · Emory University · 1998K992450 · Emory Cardiac Toolbox* 2.0 Executing on Nuclear Medicine Computers · Emory University · 1999K014033 · Emory Cardiac Toolbox 2.1 Executing on Nuclear Medicine Computers · Syntermed, Inc. · 2002K040141 · Emory Cardiac Toolbox Version 2.6 · Syntermed, Inc. · 2004K052379 · Mim 3.5 (Circa) · Mimvista Corp. · 2005×2: 2 generations cleared close together, K060816 (2006), K062163 (2006) K103785 · Mobile Mim · Mim Software, Inc. · 2011K112930 · Mobile Mim · Mim Software, Inc. · 2011K151913 · MIM-Thin Client (mobile) · Mim Software, Inc. · 2016×2: 2 generations cleared close together, K172218 (2017), K173636 (2018) 1990200020102020
  1. Root · 13 generations backK954874 Spectef ProtocolGE Medical Systems · 1996
  2. 9 generations between
  3. 3 generations backK151913 MIM-Thin Client (mobile)Mim Software, Inc. · 2016
  4. 2 generations backK172218 MIM – Y90 DosimetryMim Software, Inc. · 2017
  5. 1 generation backK173636 VelocityVarian Medical Systems, Inc. · 2018
  6. This deviceK182966 PLANET Onco DoseDosisoft · 2019
  7. Descendant · cites it directly · 0 of its ownK241765 PLANET Onco Dose (3.2)Dosisoft SA · 2025
Show all 13 generations
  1. Root · 13 generations backK954874 Spectef ProtocolGE Medical Systems · 1996
  2. 12 generations backK980914 Emory Cardiac Tool BoxEmory University · 1998
  3. 11 generations backK992450 Emory Cardiac Toolbox* 2.0 Executing on Nuclear Medicine ComputersEmory University · 1999
  4. 10 generations backK014033 Emory Cardiac Toolbox 2.1 Executing on Nuclear Medicine ComputersSyntermed, Inc. · 2002
  5. 9 generations backK040141 Emory Cardiac Toolbox Version 2.6Syntermed, Inc. · 2004
  6. 8 generations backK052379 Mim 3.5 (Circa)Mimvista Corp. · 2005
  7. 7 generations backK060816 Mim 4.0 (NeuroMimvista Corp. · 2006 · Recalled
  8. 6 generations backK062163 MimviewerMimvista Corp. · 2006
  9. 5 generations backK103785 Mobile MimMim Software, Inc. · 2011
  10. 4 generations backK112930 Mobile MimMim Software, Inc. · 2011
  11. 3 generations backK151913 MIM-Thin Client (mobile)Mim Software, Inc. · 2016
  12. 2 generations backK172218 MIM – Y90 DosimetryMim Software, Inc. · 2017
  13. 1 generation backK173636 VelocityVarian Medical Systems, Inc. · 2018
  14. This deviceK182966 PLANET Onco DoseDosisoft · 2019

Most influential descendants

Showing 1 of 1 · See all
  • K241765PLANET Onco Dose (3.2)Dosisoft SA · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−13K954874Spectef ProtocolGE Medical Systems1996—
−12K980914Emory Cardiac Tool BoxEmory University1998—
−11K992450Emory Cardiac Toolbox* 2.0 Executing on Nuclear Medicine ComputersEmory University1999—
−10K014033Emory Cardiac Toolbox 2.1 Executing on Nuclear Medicine ComputersSyntermed, Inc.2002—
−9K040141Emory Cardiac Toolbox Version 2.6Syntermed, Inc.2004—
−8K052379Mim 3.5 (Circa)Mimvista Corp.2005—
−7K060816Mim 4.0 (NeuroMimvista Corp.20061
−6K062163MimviewerMimvista Corp.2006—
−5K103785Mobile MimMim Software, Inc.2011—
−4K112930Mobile MimMim Software, Inc.2011—
−3K151913MIM-Thin Client (mobile)Mim Software, Inc.2016—
−2K172218MIM – Y90 DosimetryMim Software, Inc.2017—
−1K173636VelocityVarian Medical Systems, Inc.2018—
0 (this device)K182966PLANET Onco DoseDosisoft2019—
+1K241765PLANET Onco Dose (3.2)Dosisoft SA2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
198910
199010
199140
199220
199330
199420
199540
199630
199730
199830
199910
200050
200170
200230
200330
200460
200520
200610
200720
200810
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K182966, 1996–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K173636Velocity
Summary page 5
…System, Image Processing, Radiological (21 CFR 892.2050, Class II, Product Code LLZ) 4. Predicate Device: K173636 – Velocity – Varian Medical Systems, Inc. Executive Summary 5. Product Description: PLANET Onco Dose is a software platform ded…
Varian Medical Systems, Inc.2018—

Also cleared against K173636: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K241765PLANET Onco Dose (3.2)Dosisoft SA20250

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 7 companies; a new generation was cleared every 1.6 years on median.
  • 18 of the 72 ancestors have a recall of any class.
  • Of its descendants, K241765 (Dosisoft SA, 2025) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K173636.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

PLANET Onco Dose is a standalone software intended to be used with PET or SPECT hybrid imaging systems in order to manage, process, display and analyze nuclear medicine and other digital medical image series, to assist in medical diagnosis, to assist in treatment analysis and in therapy response assessment, to assist in the contouring of region of interest for radiotherapy. PLANET Onco Dose is dedicated to be used by qualified medical professionals in…

Read the full statement (428 words)

PLANET Onco Dose is a standalone software intended to be used with PET or SPECT hybrid imaging systems in order to manage, process, display and analyze nuclear medicine and other digital medical image series, to assist in medical diagnosis, to assist in treatment analysis and in therapy response assessment, to assist in the contouring of region of interest for radiotherapy.

PLANET Onco Dose is dedicated to be used by qualified medical professionals in Molecular Imaging, and/or Medical Oncology.

The medical modalities of these medical imaging systems include CT, MRI, SPECT, PET, XA, RT Struct and RT Dose as supported by ACR/NEMA DICOM 3 standard format.

PLANET Onco Dose provides the user with the means to segment structures in medical image volumes by providing dedicated delineation, contouring and propagation tools for both tumors and normal tissues (i.e. Regions of Interest (ROI)).

PLANET Onco Dose provides tools to display, co-register (including deformable registration), compute Standardized Uptake Value (SUV) and import / export results (contours and dosimetries) to / from Treatment Planning Systems (TPS) and PACS devices for assessment and response of patients undergoing a course of oncology treatment.

PLANET Onco Dose provides the user with the means to assist in the assessment quantification and verification of radiation doses received by tissues as a result of administering a radionuclide (e.g. Permanent Yttrium-90 microsphere implants).

PLANET Onco Dose provides tools for post-treatment absorbed dose calculation and evaluation on PET and SPECT images. The following functions are available to allow dose calculations for patients after they have received a treatment using permanent Yttrium-90 (Y90) microspheres:

- 3D liver-lung shunt assessment;

- Local Deposition Model;

- Voxel S Value approach based on the schema in MIRD Pamphlet 17 [1];

- Dosimetry based on 90Y-microspheres-PET (or SPECT Bremsstrahlung) series;

- Compatible with PET images acquired with another radioisotope instead of Y90 when Y90 acquisitions are not supported by the scanner (correction of branching ratio and decay parameters).

For Y90 microspheres, PLANET Onco Dose cannot be used to prescribe the radionuclide activity to be administered to the patient for the therapy. The user has to provide the radionuclide parameters (e.g. activity) in order for PLANET Onco Dose to estimate the radiation doses that the tissues received as a result of the administration.

PLANET Onco Dose should only be used for the retrospective determination of dose and not for the case where there is a need for retreatment using Y90 microspheres.

[1] Bolch WE et al. MIRD pamphlet no. 17: the dosimetry of nonuniform activity distributions-Radionuclide S values at the voxel level. J Nucl Med, 1999, 40:11S-36S

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K182966: PLANET Onco Dose.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K182966

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026