510(k) Report

510(k) · Cleared

Cleared

Blue Non-Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Low Dermatitis and Fentanyl Permeation Resistance Claim

K183149510(k) TraditionalCleared Jul 4, 2019233-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Central Medicare Sdn Bhd · LZC Medical Glove, Specialty

Cleared on Jul 4, 2019 after a 233-day review. FDA’s file for it is a 510(k) statement, which names no predicate.

Decided
Jul 4, 2019
Review
233 days
Class
Class I
Product code
LZC
Recalls
None
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K183149Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZC device

  1. #1 K201390 Medline Powder-Free Light Blue Nitrile Exam…score 19.2 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All LZC devices

LZC Medical Glove, Specialty: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · LZC

1in 2025

2007: 12008: 32009: 52010: 52012: 42015: 12016: 32017: 62018: 62019: 72020: 32021: 32022: 72025: 12026 (partial year): 1
2007high 7 (2019) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZC

85in 2022

2008: 166 days median2009: 253 days median2010: 68 days median2012: 149 days median2016: 258 days median2017: 121 days median2018: 105 days median2019: 200 days median2020: 167 days median2021: 94 days median2022: 85 days median
2007high 258 (2016) · low 68 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZC

1in 2021

2008: 12011: 42021: 12026 (partial year): 1
2007high 4 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZC

10in 2025

2009: 12011: 12012: 12013: 12014: 32016: 52017: 72018: 62019: 82020: 12021: 22022: 22023: 12024: 162025: 10
2007high 16 (2024) · low 0 (2007)2025

MAUDE reports naming product code LZC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
2008316610
2009525301
201056800
20110—41
2012414901
20130—01
20140—03
20151—00
2016325805
2017612107
2018610506
2019720008
2020316701
202139412
202278502
20230—01
20240—016
20251—010
20261—1—

What is it indicated for?

Blue Non Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs, with Low Dermatitis and Fentanyl Permeation Resistance Claim is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Low Dermatitis Potential - this product demonstrated reduced potential for sensitizing users to chemical additives. Tested for Use with Chemotherapy Drugs -…

Read the full statement (276 words)

Blue Non Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs, with Low Dermatitis and Fentanyl Permeation Resistance Claim is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Low Dermatitis Potential - this product demonstrated reduced potential for sensitizing users to chemical additives.

Tested for Use with Chemotherapy Drugs - Gloves have been tested for use with chemotherapy drugs using ASTM D6978-05 and will be labeled with a statement of compliance and a summary of the testing results.

Chemotherapy Drug Permeation

The following chemicals have been tested with these gloves:

Tested Chemotherapy Drug Concentrate(mg/ml) Minimum Breakthrough Detection Time (minutes)

Bleomycin Sulfate 15 >240

Busulfan 6 >240

Carboplatin 10 >240

*Carmustine 3.3 12.4

Cisplatin 1 >240

Chloroquine 50 >240

Cyclophosphamide 20 >240

Cyclosporin 100 >240

Cytarabine HCI 100 >240

Dacarbazine 10 >240

Daunorubicin HCI 5 >240

Docetaxel 10 >240

Doxorubicin Hydrochloride 2 >240

Epirubicin (Ellence) 2 >240

Etoposide 20 >240

Fludarabine 25 >240

Fluorouracil 50 >240

Gemcitabine 38 >240

Idarubicin 1 >240

Ifosfamide 50 >240

Irinotecan 20 >240

Mechlorethamine HCI 1 >240

Melphalan 5 >240

Methotrexate 25 >240

Mitomycin 0.5 >240

Mitoxantrone 2 >240

Oxaliplatin 5 >240

Paclitaxel 6 >240

Paraplatin 10 >240

Retrovir 10 >240

Rituximab 10 >240

*Thiotepa 10 24.4

Topotecane 1 >240

Trisonex 1 >240

Velcade 1 >240

Vincristine Sulfate 1 >240

* Please note that the following drugs have extremely low permeation times:

Carmustine: 12.4 minutes and Thiotepa: 24.4 minutes

Fentanyl Permeation Resistance Claim - Under the testing conditions of ASTM D6978-05, Fentanyl Citrate Injection (100mcg/2mL) was found to have no breakthrough detected up to 240 minutes.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K183149: Blue Non-Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Low Dermatitis and Fentanyl Permeation Resistance Claim.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K183149

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026