510(k) Report

510(k) · Cleared

Cleared

Yumizen C1200, Yumizen C1200 Glucose HK, Sodium Electrode, Potassium Electrode, Chloride Electrode

K183375510(k) TraditionalCleared Feb 12, 201968-day reviewClass I21 CFR 862.2160

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Horiba, Ltd. · JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Cleared on Feb 12, 2019 as substantially equivalent to K161837 (Beckman Coulter, Inc., 2016) and 1 other predicate after a 68-day review. Its family tree reaches back 5 generations to 1997.

Decided
Feb 12, 2019
Review
68 days
Class
Class I
Product code
JJE
Recalls
None
Predicates
2

Its longest line of descent runs back 5 generations to K971954 (1997). 17 ancestors in all, the oldest K883962 (1988). No later device cites it as a predicate yet.

  1. Root · 5 generations backK971954 Liquichek Urine Chemistry Control (Cat.NO. 397,398)Bio-Rad · 1997
  2. 4 generations backK020817 Liquichek Urine Chemistry ControlBio-Rad Laboratories, Inc. · 2002
  3. 3 generations backK070249 Abx Pentra 400, Calcium CP, Creatinine CP, Phosphorus CP, Amylase…Horiba Abx · 2007 · Recalled
  4. 2 generations backK070146 Abx Pentra Glucose HK CP, Urea CP, Uric Acid CPHoriba Abx · 2007
  5. 1 generation backK081276 Abx Pentra Glucose HK CP, Abx Pentra Uric Acid CP, Abx Pentra Urine…Horiba Abx · 2008 · Recalled
  6. This deviceK183375 Yumizen C1200, Yumizen C1200 Glucose HK, Sodium Electrode, Potassium…Horiba, Ltd. · 2019
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K971954Liquichek Urine Chemistry Control (Cat.NO. 397,398)Bio-Rad1997—
−4K020817Liquichek Urine Chemistry ControlBio-Rad Laboratories, Inc.2002—
−3K070249Abx Pentra 400, Calcium CP, Creatinine CP, Phosphorus CP, Amylase CP, Urine…Horiba Abx20071
−2K070146Abx Pentra Glucose HK CP, Urea CP, Uric Acid CPHoriba Abx2007—
−1K081276Abx Pentra Glucose HK CP, Abx Pentra Uric Acid CP, Abx Pentra Urine Control L/HHoriba Abx20081
0 (this device)K183375Yumizen C1200, Yumizen C1200 Glucose HK, Sodium Electrode, Potassium…Horiba, Ltd.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199010
199410
199510
199730
200020
200510
201110
201610
FIG. 1 — Lineage of K183375, 1997–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JJE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K161837ISE Reagent, Glucose, CRP Latex, DxC 700 AU Clinical Chemistry Analyzer
Summary page 7
…manufacturer Yumizen C1200 DxC 700 AU Clinical Chemistry Analyzer Beckman Coulter, Inc. k161837 Yumizen C1200 Glucose HK ABX PENTRA Glucose HK CP on Pentra C400…
Beckman Coulter, Inc.2016—
*K081276Abx Pentra Glucose HK CP, Abx Pentra Uric Acid CP, Abx Pentra Urine Control L/H
Summary page 7
…e HK ABX PENTRA Glucose HK CP on Pentra C400 HORIBA ABX SAS k081276 CLINICAL CHEMISTRY ANALYZER Sodium Electrode Potassium Electrode I…
Horiba Abx20081

Also cleared against K161837: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 4.8 years on median.
  • 4 of the 17 ancestors have a recall of any class.
  • 1 other device also cites its predicate K161837.

How does its category compare?

All JJE devices

JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · JJE

1in 2019

2008: 12011: 22012: 12013: 12014: 32019: 1
2007high 3 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JJE

224in 2014

2014: 224 days median
2007high 224 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JJE

6in 2025

2007: 202008: 202009: 142010: 322011: 252012: 162013: 212014: 142015: 152016: 92017: 182018: 172019: 72020: 72021: 62023: 122024: 62025: 62026 (partial year): 5
2007high 32 (2010) · low 0 (2022)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JJE

4,400in 2025

2007: 2,1812008: 2,1592009: 3,4402010: 3,2732011: 11,6062012: 3,7072013: 2,4892014: 2,8412015: 2,7932016: 2,5802017: 1,8692018: 4,5902019: 3,9482020: 3,9862021: 3,9202022: 3,9982023: 3,7832024: 4,2212025: 4,400
2007high 11,606 (2011) · low 1,869 (2017)2025

MAUDE reports naming product code JJE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—202,181
20081—202,159
20090—143,440
20100—323,273
20112—2511,606
20121—163,707
20131—212,489
20143224142,841
20150—152,793
20160—92,580
20170—181,869
20180—174,590
20191—73,948
20200—73,986
20210—63,920
20220—03,998
20230—123,783
20240—64,221
20250—64,400
20260—5—

What is it indicated for?

The Yumizen C1200 Glucose HK reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma, and urine using a glucose hexokinase method by colorimetry. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. The sodium electrode is an ion selective…

Read the full statement (343 words)

The Yumizen C1200 Glucose HK reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma, and urine using a glucose hexokinase method by colorimetry. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

The sodium electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The sodium electrode is used to quantify the concentrations of sodium ions in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison’s disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

The potassium electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The potassium electrode is used to quantify the concentrations of potassium ions in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases and conditions characterized by low or high blood potassium levels.

The chloride electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The chloride electrode is used to quantify the concentrations of chloride ions in serum, plasma, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

The Yumizen C1200 is an automatic chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. Applications include colorimetric and ion selective electrode. This analyzer is intended for professional use in a laboratory environment only. Tests performed using this analyzer are intended for in vitro diagnostic use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K183375: Yumizen C1200, Yumizen C1200 Glucose HK, Sodium Electrode, Potassium Electrode, Chloride Electrode.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K183375

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026