Recalled · Class II · 2020Vitality® Spinal Fixation System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Zimmer Biomet Spine, Inc. · NKB Thoracolumbosacral Pedicle Screw System
Cleared on Apr 8, 2019 after a 109-day review. The FDA summary names no predicate by number. 9 later devices descend from it. It has been recalled 23 times.
- Decided
- Apr 8, 2019
- Review
- 109 days
- Class
- Class II
- Product code
- NKB
- Recalls
- 23
- Predicates
- 0
9 later devices descend from it, 1 citing it directly.
- This deviceK183550 Vitality® Spinal Fixation SystemZimmer Biomet Spine, Inc. · 2019
- Descendant · cites it directly · 8 of its ownK203507 Vitality® Spinal Fixation SystemZimmer Biomet Spine, Inc. · 2021
- Descendant · 2 generations on · 5 of its ownK230116 Vital™ Spinal Fixation SystemZimvie Spine (Zimmer Biomet Spine, Inc.) · 2023
- Descendant · 3 generations on · 4 of its ownK240539 Vital™ Spinal Fixation SystemZimvie Spine (Zimmer Biomet Spine, Inc.) · 2024
- Descendant · 4 generations on · 2 of its ownK253721 KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O · 2025
- Descendant · 5 generations on · 1 of its ownK260786 KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O · 2026
Showing 5 of 9 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 9 · See all- K203507Vitality® Spinal Fixation SystemZimmer Biomet Spine, Inc. · 2021 · cites it directly8descendants
- K230116Vital™ Spinal Fixation SystemZimvie Spine (Zimmer Biomet Spine, Inc.) · 2023 · 2 generations on5descendants
- K240539Vital™ Spinal Fixation SystemZimvie Spine (Zimmer Biomet Spine, Inc.) · 2024 · 3 generations on4descendants
- K253721KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O · 2025 · 4 generations on2descendants
- K260786KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O · 2026 · 5 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K183550 | Vitality® Spinal Fixation System | Zimmer Biomet Spine, Inc. | 2019 | — |
| +1 | K203507 | Vitality® Spinal Fixation System | Zimmer Biomet Spine, Inc. | 2021 | — |
| +2 | K230116 | Vital™ Spinal Fixation System | Zimvie Spine (Zimmer Biomet Spine, Inc.) | 2023 | — |
| +3 | K240539 | Vital™ Spinal Fixation System | Zimvie Spine (Zimmer Biomet Spine, Inc.) | 2024 | — |
| +4 | K253721 | KHEIRON® Spinal Fixation System, including patient specific K-ROD | S.M.A.I.O | 2025 | — |
| +5 | K260786 | KHEIRON® Spinal Fixation System, including patient specific K-ROD | S.M.A.I.O | 2026 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2021 | 1 | 0 |
| 2022 | 1 | 0 |
| 2023 | 1 | 0 |
| 2024 | 2 | 0 |
| 2025 | 2 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in NKB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NKB device
- #1 K223494 CD Horizon Spinal Systemscore 40.4 · 30 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
The FDA summary names no predicate by number.
What descends from it? (9)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K203507 | Vitality® Spinal Fixation System | 2021 | 8 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 5, 2020 | Class II | Z-2212-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2231-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2225-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2229-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2219-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2221-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2218-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2230-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2226-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2211-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2224-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2217-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2216-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2214-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2232-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2223-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2222-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2233-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2220-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2227-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2234-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2210-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
| May 5, 2020 | Class II | Z-2228-2020 | Under Investigation by firm | There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively. |
The 5 latest of 23 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K203507 (Zimmer Biomet Spine, Inc., 2021) has gone furthest, with 8 descendants of its own.
How does its category compare?
All NKB devicesNKB Thoracolumbosacral Pedicle Screw System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.
FDA decisions per year · NKB
24in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · NKB
61in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NKB
1in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · NKB
1,171in 2025
MAUDE reports naming product code NKB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 38 | 87 | 4 | 34 |
| 2008 | 36 | 59 | 5 | 82 |
| 2009 | 44 | 71 | 33 | 171 |
| 2010 | 37 | 111 | 4 | 462 |
| 2011 | 48 | 84 | 3 | 577 |
| 2012 | 60 | 79 | 13 | 1,747 |
| 2013 | 35 | 73 | 0 | 1,962 |
| 2014 | 46 | 90 | 3 (1) | 1,671 |
| 2015 | 55 | 90 | 3 | 1,446 |
| 2016 | 50 | 84 | 1 | 1,585 |
| 2017 | 63 | 69 | 1 | 1,618 |
| 2018 | 56 | 61 | 3 | 1,684 |
| 2019 | 39 | 84 | 1 | 1,915 |
| 2020 | 48 | 68 | 1 | 2,076 |
| 2021 | 32 | 61 | 0 | 1,745 |
| 2022 | 34 | 88 | 5 | 1,339 |
| 2023 | 44 | 72 | 2 | 919 |
| 2024 | 39 | 57 | 3 | 1,614 |
| 2025 | 24 | 61 | 1 | 1,171 |
| 2026 | 30 | 91 | 3 | — |
https://510kreport.com/device/K183550 · printed Nov 5, 2026
Cite this page
510(k) Report. “K183550: Vitality® Spinal Fixation System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K183550
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026