# K183592: ClearLLab 10C Panels (B, T, M1, M2), Navios Flow Cytometer, Navios EX Flow Cytometer

- Pathway: 510(k) (Traditional)
- Applicant: Beckman Coulter
- Decided: Mar 21, 2019 after 90 days
- Product code: PWD (Flow Cytometric Test System For Hematopoietic Neoplasms)
- Device class: 2
- Regulation: 21 CFR 864.7010
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 6 generations back to K922745 (1993). 11 ancestors in all.
1 later device descends from it; 1 cites it directly.
3 of 13 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- DEN160047: ClearLLab T1, ClearLLab T2, ClearLLab B1, ClearLLab B2, ClearLLab M, Beckman Coulter (2017) Summary page 8.
- K130373: Navios Flow Cytometer System, Beckman Coulter, Inc. (2013) Summary page 8.
- K162897: Navios EX Flow Cytometer, 6 Color/2 Laser, Navios EX Flow Cytometer, 8 Color/2 Laser, Navios EX Flow Cytometer, 10 Color/3 Laser, Beckman Coulter (2017) Summary page 8.

## Most influential descendants
- K232600: DxFLEX Flow Cytometer; ClearLLab 10C Panels (2023), 0 descendants of its own

## Indications for use
The ClearLLab 10C Panels are intended for in vitro diagnostic use for qualitative identification of cell populations by multiparameter immunophenotyping on the Navios and Navios EX flow cytometers. These reagents are used as an aid in the differential diagnosis of hematologically abnormal patients having, or suspected of having, the following hematopoietic neoplasms: chronic leukemia, acute leukemia, non-Hodgkin lymphoma, myeloma, myelodysplastic syndrome (MDS), and/or myeloproliferative neoplasms (MPN). The reagents can be used with peripheral whole blood (collected in K2EDTA, Acid Citrate Dextrose (ACD) or Heparin), bone marrow (collected in K2EDTA, ACD or Heparin) and lymph node specimens. Interpretation of the results should be confirmed by a pathologist or equivalent professional in conjunction with other clinical and laboratory findings. These reagents provide multiparameter, qualitative results for the surface antigens listed below: • ClearLLab 10C B Cell Tube: Kappa, Lambda, CD10, CD5, CD200, CD34, CD38, CD20, CD19, CD45 • ClearLLab 10C T Cell Tube: TCRȖį, CD4, CD2, CD56, CD5, CD34, CD3, CD8, CD7, CD45 • ClearLLab 10C M1 Cell Tube: CD16, CD7, CD10, CD13, CD64, CD34, CD14, HLA-DR, CD11b, CD45 • ClearLLab 10C M2 Cell Tube: CD15, CD123, CD117, CD13, CD33, CD34, CD38, HLA-DR, CD19, CD45 The Navios Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to ten fluorescent detection channels using three lasers (488 nm, 638 nm, and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument. The Navios EX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to ten fluorescent detection channels using three lasers (488 nm, 638 nm, and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.

Source page: https://510kreport.com/device/K183592
Methodology: https://510kreport.com/methodology