510(k) Report

510(k) · Cleared

Cleared

Bfit Sleep, Bfit Sleep with DentiTrac, Bfit Engage, Bfit Engage with DentiTrac

K190353510(k) TraditionalCleared Apr 10, 2020420-day reviewClass II21 CFR 872.5570

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Residential Home Sleep Services · PLC Sleep Appliances With Patient Monitoring

Cleared on Apr 10, 2020 as substantially equivalent to K170692 (Respire Medical Holding, 2017) and 1 other predicate after a 420-day review. Its family tree reaches back 10 generations to 1989.

Decided
Apr 10, 2020
Review
420 days
Class
Class II
Product code
PLC
Recalls
None
Predicates
2

Its longest line of descent runs back 10 generations to K882303 (1989). 28 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 10 generations backK882303 H&M Anti-Snoring DeviceHays & Meade, Inc. · 1989
  2. 6 generations between
  3. 3 generations backK140278 Somnodent Fusion, Classic; Somnodent Fusion, FlexSomnomed, Inc. · 2014
  4. 2 generations backK150369 SomnoDent Classic with DentiTrac Micro-recorder, SomnoDent Flex with…Somnomed, Inc. · 2015
  5. 1 generation backK170692 Respire Pink Series with DentiTracRespire Medical Holding · 2017
  6. This deviceK190353 Bfit Sleep, Bfit Sleep with DentiTrac, Bfit Engage, Bfit Engage with…Residential Home Sleep Services · 2020
Show all 10 generations
  1. Root · 10 generations backK882303 H&M Anti-Snoring DeviceHays & Meade, Inc. · 1989
  2. 9 generations backK962516 Tap Anti-Snoring DeviceNellcor Puritan Bennett, Inc. · 1996
  3. 8 generations backK042161 Anti-Snoring/Sleep Apnea DeviceRj & VK Bird Pty, Ltd. · 2004
  4. 7 generations backK050592 Somnomed Mas RxaSomnomed, Ltd. · 2005
  5. 6 generations backK073004 Somnomed BflexSomnomed, Inc. · 2008
  6. 5 generations backK121340 Somnodent G2Somnomed, Inc. · 2012
  7. 4 generations backK130558 Somnodent Herbst; Classic, FlexSomnomed, Inc. · 2013
  8. 3 generations backK140278 Somnodent Fusion, Classic; Somnodent Fusion, FlexSomnomed, Inc. · 2014
  9. 2 generations backK150369 SomnoDent Classic with DentiTrac Micro-recorder, SomnoDent Flex with…Somnomed, Inc. · 2015
  10. 1 generation backK170692 Respire Pink Series with DentiTracRespire Medical Holding · 2017
  11. This deviceK190353 Bfit Sleep, Bfit Sleep with DentiTrac, Bfit Engage, Bfit Engage with…Residential Home Sleep Services · 2020
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K882303H&M Anti-Snoring DeviceHays & Meade, Inc.1989—
−9K962516Tap Anti-Snoring DeviceNellcor Puritan Bennett, Inc.1996—
−8K042161Anti-Snoring/Sleep Apnea DeviceRj & VK Bird Pty, Ltd.2004—
−7K050592Somnomed Mas RxaSomnomed, Ltd.2005—
−6K073004Somnomed BflexSomnomed, Inc.2008—
−5K121340Somnodent G2Somnomed, Inc.2012—
−4K130558Somnodent Herbst; Classic, FlexSomnomed, Inc.2013—
−3K140278Somnodent Fusion, Classic; Somnodent Fusion, FlexSomnomed, Inc.2014—
−2K150369SomnoDent Classic with DentiTrac Micro-recorder, SomnoDent Flex with DentiTrac…Somnomed, Inc.2015—
−1K170692Respire Pink Series with DentiTracRespire Medical Holding2017—
0 (this device)K190353Bfit Sleep, Bfit Sleep with DentiTrac, Bfit Engage, Bfit Engage with DentiTracResidential Home Sleep Services2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199620
199730
199810
199910
200330
200610
200710
201110
201320
201510
201710
FIG. 1 — Lineage of K190353, 1989–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PLC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PLC device

  1. #1 K252765 ProSomnus RPMO2 OSA Device (RPMO2 OSA)score 17.1 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K170692Respire Pink Series with DentiTrac
Summary page 4
…ulation Number 872.5570 Panel Dental Predicate Devices Primary Predicate Device: Respire Pink Series with DentiTrac, K170692 Reference Device: ASD Elastic, K163580 Intended Use/Indications for The BFit Intraoral Applian…
Respire Medical Holding2017—
*K163580ASD Oral Appliances
Summary page 4
…ntal Predicate Devices Primary Predicate Device: Respire Pink Series with DentiTrac, K170692 Reference Device: ASD Elastic, K163580 Intended Use/Indications for The BFit Intraoral Appliances are likely to reduce snoring and mild to Use…
American Sleep Dentistry2017—

Also cleared against K170692: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 7 companies; a new generation was cleared every 2.6 years on median.
  • 2 of the 28 ancestors have a recall of any class.
  • 1 other device also cites its predicate K170692.

How does its category compare?

All PLC devices

PLC Sleep Appliances With Patient Monitoring: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · PLC

1in 2024

2015: 12016: 22017: 22019: 12020: 12024: 12026 (partial year): 2
2007high 2 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PLC

Too few PLC 510(k)s in any year since 2007 to give a median.

Recalls per year · PLC

No recall names PLC since 2007.

Adverse event reports per year · PLC

1in 2023

2016: 12017: 12020: 12023: 1
2007high 1 (2016) · low 0 (2007)2025

MAUDE reports naming product code PLC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20151—00
20162—01
20172—01
20180—00
20191—00
20201—01
20210—00
20220—00
20230—01
20241—00
20250—00
20262—0—

What is it indicated for?

The BFit Intraoral Appliances are likely to reduce snoring and mild to moderate Obstructive Sleep Apnea (OSA) in adult patients 18 years of age or older.

Optionally, the DentiTrac Micro-Recorder may be incorporated into the Bfit device. The DentiTrac is intended to measure patient compliance to oral appliance therapy in combination with the DentiTrac System.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K190353: Bfit Sleep, Bfit Sleep with DentiTrac, Bfit Engage, Bfit Engage with DentiTrac.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K190353

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026