510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

VITROS XT Chemistry Products TBIL-ALKP Slides

K190807510(k) TraditionalCleared Apr 26, 201928-day reviewClass II21 CFR 862.1110Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ortho-Clinical Diagnostics, Inc. · CIG Diazo Colorimetry, Bilirubin

Cleared on Apr 26, 2019 after a 28-day review. The FDA summary names no predicate by number. It has been recalled 3 times.

Decided
Apr 26, 2019
Review
28 days
Class
Class II
Product code
CIG
Recalls
3
Predicates
0

The FDA summary names no predicate by number.

FIG. 1 — Lineage of K190807Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CIG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CIG device

  1. #1 K251998 Atellica CH Diazo Total Bilirubin (D_TBil)score 11.8 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 24, 2025Class IIZ-1835-2025Storagetransit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test…
Aug 5, 2023Class IIZ-2623-2023Under Investigation by firmSamples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System…
Jul 13, 2020Class IIZ-2875-2020OtherEtrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides

No ancestor or descendant has a Class I recall.

How does its category compare?

All CIG devices

CIG Diazo Colorimetry, Bilirubin: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · CIG

2in 2023

2007: 12008: 32009: 22010: 12011: 22013: 52015: 12017: 12019: 12022: 12023: 22026 (partial year): 1
2007high 5 (2013) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CIG

64in 2013

2008: 199 days median2013: 64 days median
2007high 199 (2008) · low 64 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CIG

1in 2019

2009: 32013: 32019: 1
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CIG

52in 2025

2007: 22008: 82009: 32010: 62011: 382012: 142013: 82014: 32015: 72016: 42017: 112018: 232019: 252020: 312021: 222022: 262023: 342024: 392025: 52
2007high 52 (2025) · low 2 (2007)2025

MAUDE reports naming product code CIG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—02
2008319908
20092—33
20101—06
20112—038
20120—014
201356438
20140—03
20151—07
20160—04
20171—011
20180—023
20191—125
20200—031
20210—022
20221—026
20232—034
20240—039
20250—052
20261—0—

What is it indicated for?

The TBIL test within the VITROS XT Chemistry Products TBIL-ALKP Slides quantitatively measure total bilirubin (TBIL) concentration in serum and plasma using VITROS XT 7600 Integrated Systems. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis and treatment of liver, hemolytic hematological and metabolic disorders, including hepatitis and gall bladder…

Read the full statement (115 words)

The TBIL test within the VITROS XT Chemistry Products TBIL-ALKP Slides quantitatively measure total bilirubin (TBIL) concentration in serum and plasma using VITROS XT 7600 Integrated Systems. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis and treatment of liver, hemolytic hematological and metabolic disorders, including hepatitis and gall bladder block.

The ALKP test within the VITROS XT Chemistry Products TBIL-ALKP Slides quantitatively measure alkaline phosphatase (ALKP) activity in serum and plasma using VITROS XT 7600 Integrated Systems. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K190807: VITROS XT Chemistry Products TBIL-ALKP Slides.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K190807

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026