510(k) · Cleared
ClearedAdvanced Intermezzo Implant System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Megagen Implant Co., Ltd. · DZE Implant, Endosseous, Root-Form
Cleared on Oct 8, 2019 as substantially equivalent to K051018 (Megagen Co., Ltd., 2005) after a 162-day review. Its family tree reaches back 1 generation to 2005.
- Decided
- Oct 8, 2019
- Review
- 162 days
- Class
- Class II
- Product code
- DZE
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 1 generation to K051018 (2005). No later device cites it as a predicate yet.
- Root · 1 generation backK051018 Intermezzo Implant SystemsMegagen Co., Ltd. · 2005
- This deviceK191127 Advanced Intermezzo Implant SystemMegagen Implant Co., Ltd. · 2019
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K051018 | Intermezzo Implant Systems | Megagen Co., Ltd. | 2005 | — |
| 0 (this device) | K191127 | Advanced Intermezzo Implant System | Megagen Implant Co., Ltd. | 2019 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DZE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DZE device
- #1 K211090 ZENEX Implant Systemscore 25.4 · 15 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K051018 | Intermezzo Implant SystemsSummary page 4…ct Code: DZE, NHA ▪ Classification regulation: Class II, 21 CFR 872.3640 4. Predicate Device: ▪ Primary Predicate: K051018 Intermezzo Implant System ▪ Reference Devices: K112540 S-MiNi Implant K123870 X… | 2005 |
| † | K112540 | S-Mini Implant System | 2012 |
| † | K123870 | Xpeed Anyridge Internal Implant System | 2013 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Its one ancestor has no recall.
- 5 other devices also cite its predicate K051018.
How does its category compare?
All DZE devicesDZE Implant, Endosseous, Root-Form: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.
FDA decisions per year · DZE
31in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DZE
184in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DZE
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DZE
465,788in 2025
MAUDE reports naming product code DZE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 43 | 107 | 11 | 225 |
| 2008 | 61 | 118 | 4 | 393 |
| 2009 | 41 | 110 | 3 | 164 |
| 2010 | 34 | 162 | 3 | 360 |
| 2011 | 26 | 194 | 0 | 362 |
| 2012 | 39 | 174 | 11 | 358 |
| 2013 | 42 | 195 | 1 | 391 |
| 2014 | 36 | 265 | 7 | 1,124 |
| 2015 | 36 | 215 | 3 | 2,894 |
| 2016 | 40 | 219 | 15 | 1,116 |
| 2017 | 48 | 223 | 4 | 9,719 |
| 2018 | 41 | 200 | 4 | 12,892 |
| 2019 | 31 | 231 | 8 | 167,016 |
| 2020 | 41 | 230 | 0 | 255,524 |
| 2021 | 34 | 182 | 6 | 479,618 |
| 2022 | 34 | 243 | 0 | 483,224 |
| 2023 | 35 | 248 | 0 | 537,585 |
| 2024 | 40 | 222 | 1 | 473,629 |
| 2025 | 31 | 184 | 1 | 465,788 |
| 2026 | 15 | 194 | 2 | — |
What is it indicated for?
Advanced Intermezzo Implant Systems is threaded one-piece implants designed for orthodontic one-stage surgical procedures in upper and lower jaw to provide a means of prosthetic attachment to restore a patient's chewing function. Advanced Intermezzo Implant System consists of single-stage, root-form dental implants. The system is designed to provide immediate provisional implant to provide temporary support for prosthetic devices during the healing phase…
Read the full statement (120 words)
Advanced Intermezzo Implant Systems is threaded one-piece implants designed for orthodontic one-stage surgical procedures in upper and lower jaw to provide a means of prosthetic attachment to restore a patient's chewing function.
Advanced Intermezzo Implant System consists of single-stage, root-form dental implants.
The system is designed to provide immediate provisional implant to provide temporary support for prosthetic devices during the healing phase of permanent root form implants. Depends on a patient's quality of bone condition, Advanced Intermezzo Fixtures are to be removed within six to ten weeks after the surgery. The system is intended for immediate placement in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cement retained, screw retained, or overdenture restorations.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K191127 · printed Nov 4, 2026
Cite this page
510(k) Report. “K191127: Advanced Intermezzo Implant System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K191127
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026