510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

iNSitu Bipolar Hip System

K191297510(k) TraditionalCleared Sep 17, 2019126-day reviewClass II21 CFR 888.3390Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Theken Companies, LLC · KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented

Cleared on Sep 17, 2019 as substantially equivalent to K082705 (Biopro, Inc., 2008) and 1 other predicate after a 126-day review. Its family tree reaches back 4 generations to 1995. 1 later device descends from it. It has been recalled once.

Decided
Sep 17, 2019
Review
126 days
Class
Class II
Product code
KWY
Recalls
1
Predicates
2

Its longest line of descent runs back 4 generations to K953510 (1995). 8 ancestors in all, the oldest K931655 (1994). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2005: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2026: 1 descendant clearedK953510 · Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented or Uncemented Prosthesis · Encore Orthopedics, Inc. · 1995 · recalledK982447 · Plus Bipolar Prosthesis Cocrmo · Plus Orthopedics · 1998K082705 · Biopro Bipolar Head, Models 18130-18152 · Biopro, Inc. · 2008K100761 · Biopro Polar Head · Biopro, Inc. · 2010 · recalled1990200020102020
  1. Root · 4 generations backK953510 Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented or Uncemented…Encore Orthopedics, Inc. · 1995 · Recalled
  2. 3 generations backK982447 Plus Bipolar Prosthesis CocrmoPlus Orthopedics · 1998
  3. 2 generations backK082705 Biopro Bipolar Head, Models 18130-18152Biopro, Inc. · 2008
  4. 1 generation backK100761 Biopro Polar HeadBiopro, Inc. · 2010 · Recalled
  5. This deviceK191297 iNSitu Bipolar Hip SystemTheken Companies, LLC · 2019
  6. Descendant · cites it directly · 0 of its ownK243192 iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with…Nextstep Arthopedix, LLC · 2026

Most influential descendants

Showing 1 of 1 · See all
  • K243192iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with Plasma Sprayed CPTi…Nextstep Arthopedix, LLC · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K953510Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented or Uncemented ProsthesisEncore Orthopedics, Inc.199518
−3K982447Plus Bipolar Prosthesis CocrmoPlus Orthopedics1998—
−2K082705Biopro Bipolar Head, Models 18130-18152Biopro, Inc.2008—
−1K100761Biopro Polar HeadBiopro, Inc.20101
0 (this device)K191297iNSitu Bipolar Hip SystemTheken Companies, LLC2019—
+1K243192iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with Plasma Sprayed…Nextstep Arthopedix, LLC2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
200510
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K191297, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWY device

  1. #1 K253357 b-ONE® Bipolar Headscore 11.2 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082705Biopro Bipolar Head, Models 18130-18152
Summary page 4
…87 (Orthopedic) 1.4 Legally Marketed Predicate NextStep Arthropedix is utilizing the BioPro Bipolar Head as a predicate device (K082705/K100761). The iNSitu Bipolar Hip System features component designs, materials, indications and manufacturing methods tha…
Biopro, Inc.2008—
*K100761Biopro Polar Head
Summary page 4
…87 (Orthopedic) 1.4 Legally Marketed Predicate NextStep Arthropedix is utilizing the BioPro Bipolar Head as a predicate device (K082705/K100761). The iNSitu Bipolar Hip System features component designs, materials, indications and manufacturing methods that are…
Biopro, Inc.20101

Also cleared against K082705: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K243192iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with Plasma Sprayed CPTi…Nextstep Arthopedix, LLC20260

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 11, 2025Class IIZ-0227-2026Package design/selectionDownstream recall for RES 97605, BioPro femoral components. Product failed transit testing. Impacted product may have potential issues with sterile barrier.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 4 companies; a new generation was cleared every 8.8 years on median.
  • 4 of the 8 ancestors have a recall of any class.
  • Of its descendants, K243192 (Nextstep Arthopedix, LLC, 2026) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K082705.

How does its category compare?

All KWY devices

KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · KWY

1in 2025

2008: 32009: 12010: 42011: 12012: 12013: 22014: 12015: 12016: 12017: 12018: 32019: 52020: 12022: 22023: 22025: 12026 (partial year): 1
2007high 5 (2019) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWY

126in 2019

2008: 123 days median2010: 187 days median2018: 219 days median2019: 126 days median
2007high 219 (2018) · low 123 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWY

2in 2025

2007: 252009: 242010: 12012: 12013: 22016: 32017: 22018: 12019: 12024: 12025: 2
2007high 25 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWY

180in 2025

2007: 3702008: 9002009: 9612010: 6882011: 7582012: 9782013: 1,1482014: 1,0552015: 1,2132016: 8772017: 8262018: 1,3802019: 1,5302020: 1,2592021: 2762022: 2082023: 1782024: 2092025: 180
2007high 1,530 (2019) · low 178 (2023)2025

MAUDE reports naming product code KWY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—25370
200831230900
20091—24961
201041871688
20111—0758
20121—1978
20132—21,148
20141—01,055
20151—01,213
20161—3877
20171—2826
2018321911,380
2019512611,530
20201—01,259
20210—0276
20222—0208
20232—0178
20240—1209
20251—2180
20261—0—

What is it indicated for?

The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System Femoral Stems for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:

• Femoral neck and trochanteric fractures of the proximal femur;

• Osteonecrosis of the femoral head;

• Revisions procedures where other devices or treatments for these indications have failed

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K191297: iNSitu Bipolar Hip System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K191297

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026