510(k) · Cleared
ClearedHarrier-SA Lumbar Interbody System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Choice Spine, LLC · MAX Intervertebral Fusion Device With Bone Graft, Lumbar
Cleared on Jul 5, 2019 after a 44-day review. The summary cites 2 numbers, none of which could be confirmed as its predicate (see the table below).
- Decided
- Jul 5, 2019
- Review
- 44 days
- Class
- Class II
- Product code
- MAX
- Recalls
- None
- Predicates
- 2
The summary cites 2 numbers, none of which could be confirmed as its predicate (see the table below).
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MAX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MAX device
- #1 K203714 NuVasive Thoracolumbar Interbody Systems…score 102.5 · 24 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.
| Type | Number | Device | Decided |
|---|---|---|---|
| Kept off the family tree | K180519 | FDA’s K180519 is HARRIER-SA™ Lumbar Interbody System Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree | 2018 |
| Kept off the family tree | K163269 | FDA’s K163269 is Endoskeleton (R) TAS Interbody Fusion Device / Endoskeleton (R) TAS… Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree | 2017 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
How does its category compare?
All MAX devicesMAX Intervertebral Fusion Device With Bone Graft, Lumbar: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.
FDA decisions per year · MAX
29in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MAX
85in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MAX
5in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · MAX
811in 2025
MAUDE reports naming product code MAX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 86 | 0 | 12 |
| 2008 | 35 | 88 | 2 | 19 |
| 2009 | 31 | 102 | 9 | 90 |
| 2010 | 29 | 137 | 0 | 154 |
| 2011 | 49 | 98 | 3 | 201 |
| 2012 | 48 | 69 | 11 (1) | 327 |
| 2013 | 58 | 102 | 0 | 380 |
| 2014 | 58 | 115 | 3 | 366 |
| 2015 | 53 | 113 | 1 | 524 |
| 2016 | 64 | 124 | 6 | 647 |
| 2017 | 76 | 116 | 1 | 596 |
| 2018 | 64 | 114 | 4 | 616 |
| 2019 | 60 | 131 | 1 | 699 |
| 2020 | 56 | 106 | 0 | 640 |
| 2021 | 53 | 109 | 0 | 622 |
| 2022 | 41 | 95 | 1 | 618 |
| 2023 | 30 | 87 | 0 | 576 |
| 2024 | 38 | 91 | 1 | 729 |
| 2025 | 29 | 85 | 5 | 811 |
| 2026 | 18 | 95 | 0 | — |
What is it indicated for?
The ChoiceSpine Harrier-SA™ Lumbar Interbody System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. This device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the…
Read the full statement (152 words)
The ChoiceSpine Harrier-SA™ Lumbar Interbody System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. This device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. This device is designed to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
The ChoiceSpine Harrier-SA™ Lumbar Interbody System is a stand-alone device intended to be used with four bone screws. Supplemental fixation, cleared by the FDA for use in the lumbosacral spine, must be used with implants >20°. Supplemental fixation must also be used whenever fewer than four bone screws are used.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K191367 · printed Nov 6, 2026
Cite this page
510(k) Report. “K191367: Harrier-SA Lumbar Interbody System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K191367
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026