510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Discovery XR656 HD with VolumeRad

K191699510(k) TraditionalCleared Jul 24, 201929-day reviewClass II21 CFR 892.1680Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Hualun Medical Systems Co., Ltd. · KPR System, X-Ray, Stationary

Cleared on Jul 24, 2019 as substantially equivalent to K132261 (Ge Medical Systems, LLC, 2013) and 1 other predicate after a 29-day review. Its family tree reaches back 8 generations to 1986. 1 later device descends from it. It has been recalled 3 times.

Decided
Jul 24, 2019
Review
29 days
Class
Class II
Product code
KPR
Recalls
3
Predicates
2

Its longest line of descent runs back 8 generations to K862120 (1986). 12 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 8 generations backK862120 SCX SystemGeneral Electric Co. · 1986 · Recalled
  2. 7 generations backK982196 Digital Radiographic Imaging SystemGE Medical Systems · 1998
  3. 6 generations backK992066 Revolution TX/I Digital Radiographic Table SystemGE Medical Systems · 1999
  4. 5 generations backK012389 Revolution XR/D Digital Radiographic Imaging SystemGe Medical Systems, Inc. · 2001 · Recalled
  5. 4 generations backK042602 Revolution XR/D Digital Radiographic Imaging System with Manual…Ge Medical Systems, LLC · 2005
  6. 3 generations backK050704 Modification to: Revolution XR/D Digital Radiographic Imaging System…General Electric Co. · 2005 · Recalled
  7. 2 generations backK143270 Optima XR646Ge Hualun Medical Systems Co., Ltd. · 2015
  8. 1 generation backK172869 Discovery XR656 HDGe Hualun Medical Systems Co., Ltd. · 2017 · Recalled
  9. This deviceK191699 Discovery XR656 HD with VolumeRadGe Hualun Medical Systems Co., Ltd. · 2019
  10. Descendant · cites it directly · 0 of its ownK250788 Definium Tempo SelectGe Hualun Medical Systems Co., Ltd. · 2025

Most influential descendants

Showing 1 of 1 · See all
  • K250788Definium Tempo SelectGe Hualun Medical Systems Co., Ltd. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K862120SCX SystemGeneral Electric Co.19863
−7K982196Digital Radiographic Imaging SystemGE Medical Systems1998—
−6K992066Revolution TX/I Digital Radiographic Table SystemGE Medical Systems1999—
−5K012389Revolution XR/D Digital Radiographic Imaging SystemGe Medical Systems, Inc.20014
−4K042602Revolution XR/D Digital Radiographic Imaging System with Manual Image PastingGe Medical Systems, LLC2005—
−3K050704Modification to: Revolution XR/D Digital Radiographic Imaging System with…General Electric Co.20052
−2K143270Optima XR646Ge Hualun Medical Systems Co., Ltd.2015—
−1K172869Discovery XR656 HDGe Hualun Medical Systems Co., Ltd.20171
0 (this device)K191699Discovery XR656 HD with VolumeRadGe Hualun Medical Systems Co., Ltd.2019—
+1K250788Definium Tempo SelectGe Hualun Medical Systems Co., Ltd.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199410
200510
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K191699, 1986–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K132261Discovery XR656 with Volumerad (Digital Tomosynthesis)
Summary page 4
…0 Product Code: Classification: Class II Product Codes: KPR, MQB, IZF Predicate Device Discovery XR656 With VolumeRad (Digital Tomosynthesis) (K132261) #1: Regulation Name: Stationary X-Ray System Regulation: 21 CFR 892.1680 Cl…
Ge Medical Systems, LLC20131
*K172869Discovery XR656 HD
Summary page 4
…Regulation: 21 CFR 892.1680 Classification: Class II Product Codes: KPR, MQB, IZF Predicate Device Discovery XR656 HD (K172869) #2: Regulation Name: Stationary X-Ray System Regulation: 21 CFR 892.1680 Cl…
Ge Hualun Medical Systems Co., Ltd.20171
†K173602Optima XR240amx, AutoGridGe Medical Systems, LLC20182
†K181526XRpad2 4343 HWC-M Flat Panel DetectorVarex Imaging Corporation2018—

Also cleared against K132261: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K250788Definium Tempo SelectGe Hualun Medical Systems Co., Ltd.20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 28, 2025Class IIZ-1044-2025Radiation Control for Health and Safety ActThe user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the…
Nov 7, 2023Class IIZ-0552-2024Process designGE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension…
Nov 7, 2023Class IIZ-0553-2024Process designGE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 5 companies; a new generation was cleared every 2.7 years on median.
  • 7 of the 12 ancestors have a recall of any class.
  • Of its descendants, K250788 (Ge Hualun Medical Systems Co., Ltd., 2025) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K132261.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

The Discovery XR656 HD is intended to generate digital radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position and the system is intended for use in all routine radiography exams. Optional image pasting function enables the operator to stitch sequentially acquired radiographs into a…

Read the full statement (217 words)

The Discovery XR656 HD is intended to generate digital radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position and the system is intended for use in all routine radiography exams. Optional image pasting function enables the operator to stitch sequentially acquired radiographs into a single image.

The Discovery XR656 HD incorporates AutoGrid, which is an optional image processing software installed as a part of the systems Helix image processing software. AutoGrid can be used in lieu of an anti-scatter grid to improve image contrast in general radiographic images by reducing the effects of scatter radiation.

When the VolumeRAD option is included on the system, the system can generate tomographic images of human anatomy including the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages.

When the VolumeRAD option is used for patients undergoing thoracic imaging, it is indicated for the detection of lung nodules. VolumeRad generates diagnostic images of the chest that aid the radiologist in achieving superior detectability of lung nodules versus posterior-anterior and left lateral views of the chest, at a comparable radiation level.

The device is not intended for mammographic applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K191699: Discovery XR656 HD with VolumeRad.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K191699

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026