510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Steripath Gen2 Blood Collection System

K192247510(k) TraditionalCleared Feb 28, 2020193-day reviewClass II21 CFR 862.1675Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Magnolia Medical Technologies, Inc. · JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Cleared on Feb 28, 2020 as substantially equivalent to K030573 (Bd Vacutainer Systems Preanalyical Solutions, 2003) after a 193-day review. Its family tree reaches back 5 generations to 1997. 5 later devices descend from it. It has been recalled 2 times.

Decided
Feb 28, 2020
Review
193 days
Class
Class II
Product code
JKA
Recalls
2
Predicates
1

Its longest line of descent runs back 5 generations to K970259 (1997). 8 ancestors in all, the oldest K921636 (1992). 5 later devices descend from it, 3 citing it directly.

  1. Root · 5 generations backK970259 E-Z Set Infusion Set/Saf-T E-Z Set Infusion SetBecton Dickinson Vascular Access, Inc. · 1997
  2. 4 generations backK980414 Vacutainer Brand Safety-Lok Blood Collection Set Model Multiple…Bd Becton Dickinson Vacutainer Systems Preanalytic · 1998 · Recalled
  3. 3 generations backK011984 BD Vacutainer Push Button Blood Colection SetBecton, Dickinson & CO · 2001 · Recalled
  4. 2 generations backK022875 Modification to BD Vacutainer Push Button Blood Collection SetBecton, Dickinson & CO · 2002
  5. 1 generation backK030573 BD Vacutainer Push Button Blood Collection Set, Models 367338…Bd Vacutainer Systems Preanalyical Solutions · 2003 · Recalled
  6. This deviceK192247 Steripath Gen2 Blood Collection SystemMagnolia Medical Technologies, Inc. · 2020
  7. Descendant · cites it directly · 2 of its ownK200661 Steripath Micro Blood Collection SystemMagnolia Medical Technologies, Inc. · 2020
  8. Descendant · cites it directly · 0 of its ownK231282 LifeOutcomes C-Quest™ Blood Culture Sampling DeviceLifeoutcomes, LLC · 2024
  9. Descendant · 2 generations on · 0 of its ownK220677 Kurin Blood Culture Collection Set with Kurin Lock TechnologyKurin, Inc. · 2023
  10. Descendant · 2 generations on · 0 of its ownK222299 Steripath Micro Blood Collection SystemMagnolia Medical Technologies · 2022
  11. Descendant · cites it directly · 0 of its ownK201237 Velano Vascular Blood Collection AdapterVelano Vascular · 2022

Most influential descendants

Showing 5 of 5 · See all
  • K200661Steripath Micro Blood Collection SystemMagnolia Medical Technologies, Inc. · 2020 · cites it directly2descendants
  • K231282LifeOutcomes C-Quest™ Blood Culture Sampling DeviceLifeoutcomes, LLC · 2024 · cites it directly0descendants
  • K220677Kurin Blood Culture Collection Set with Kurin Lock TechnologyKurin, Inc. · 2023 · 2 generations on0descendants
  • K222299Steripath Micro Blood Collection SystemMagnolia Medical Technologies · 2022 · 2 generations on0descendants
  • K201237Velano Vascular Blood Collection AdapterVelano Vascular · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K970259E-Z Set Infusion Set/Saf-T E-Z Set Infusion SetBecton Dickinson Vascular Access, Inc.1997—
−4K980414Vacutainer Brand Safety-Lok Blood Collection Set Model Multiple, Vacutainer…Bd Becton Dickinson Vacutainer Systems Preanalytic19981
−3K011984BD Vacutainer Push Button Blood Colection SetBecton, Dickinson & CO20017
−2K022875Modification to BD Vacutainer Push Button Blood Collection SetBecton, Dickinson & CO2002—
−1K030573BD Vacutainer Push Button Blood Collection Set, Models 367338, 367344, 367336…Bd Vacutainer Systems Preanalyical Solutions20035
0 (this device)K192247Steripath Gen2 Blood Collection SystemMagnolia Medical Technologies, Inc.2020—
+1K200661Steripath Micro Blood Collection SystemMagnolia Medical Technologies, Inc.2020—
+1K231282LifeOutcomes C-Quest™ Blood Culture Sampling DeviceLifeoutcomes, LLC2024—
+2K220677Kurin Blood Culture Collection Set with Kurin Lock TechnologyKurin, Inc.2023—
+2K222299Steripath Micro Blood Collection SystemMagnolia Medical Technologies2022—
+1K201237Velano Vascular Blood Collection AdapterVelano Vascular2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199310
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202220
202310
202410
FIG. 1 — Lineage of K192247, 1997–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JKA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JKA device

  1. #1 K200027 Blood Collection Needle with/without Holder…score 15.6 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030573BD Vacutainer Push Button Blood Collection Set, Models 367338, 367344, 367336, 367342…
Summary page 4
…Number: 862.1675 Regulatory Class: Class II Product Code: JKA and FPA Predicate Device: BD Vacutainer® Push Button Blood Collection Set (K030573) Device Description: The Steripath® Gen2 Blood Collection System diverts and sequesters the initia…
Bd Vacutainer Systems Preanalyical Solutions20035

Also cleared against K030573: 4 other devices →

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K200661Steripath Micro Blood Collection SystemMagnolia Medical Technologies, Inc.20202
K231282LifeOutcomes C-Quest™ Blood Culture Sampling DeviceLifeoutcomes, LLC20240
K201237Velano Vascular Blood Collection AdapterVelano Vascular20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 2, 2021Class IIZ-0318-2022Process controlDue to an uptick of complaints associated with premature needle retraction of blood collection system.
Aug 2, 2021Class IIZ-2393-2021Labeling Change ControlProduct was shipped with the incorrect Transfer Adapter Instructions For Use (IFU).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 1.0 years on median.
  • 3 of the 8 ancestors have a recall of any class.
  • Of its descendants, K200661 (Magnolia Medical Technologies, Inc., 2020) has gone furthest, with 2 descendants of its own.
  • 4 other devices also cite its predicate K030573.

How does its category compare?

All JKA devices

JKA Tubes, Vials, Systems, Serum Separators, Blood Collection: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · JKA

7in 2025

2007: 42008: 22009: 22010: 42012: 12013: 32014: 12015: 12016: 42017: 22018: 12019: 62020: 62021: 42022: 92023: 172024: 62025: 72026 (partial year): 8
2007high 17 (2023) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JKA

115in 2025

2007: 281 days median2010: 28 days median2013: 244 days median2016: 116 days median2019: 181 days median2020: 181 days median2021: 217 days median2022: 164 days median2023: 189 days median2024: 240 days median2025: 115 days median2026 (partial year): 250 days median
2007high 281 (2007) · low 28 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JKA

13in 2025

2007: 32008: 32009: 32010: 42011: 142012: 22013: 12014: 12015: 62016: 12017: 22018: 22019: 92021: 32022: 22023: 22025: 132026 (partial year): 1
2007high 14 (2011) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JKA

6,089in 2025

2007: 732008: 782009: 422010: 372011: 672012: 542013: 332014: 682015: 412016: 652017: 2,0392018: 9,7312019: 6,4692020: 3,7742021: 3,6102022: 3,3752023: 4,4002024: 4,6092025: 6,089
2007high 9,731 (2018) · low 33 (2013)2025

MAUDE reports naming product code JKA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074281373
20082—378
20092—342
2010428437
20110—1467
20121—254
20133244133
20141—168
20151—641
20164116165
20172—22,039
20181—29,731
2019618196,469
2020618103,774
2021421733,610
2022916423,375
20231718924,400
2024624004,609
20257115136,089
202682501—

What is it indicated for?

The Steripath® Gen2 Blood Collection System is a system to draw blood for in vitro diagnostic testing. The Steripath® Gen2 Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard…

Read the full statement (109 words)

The Steripath® Gen2 Blood Collection System is a system to draw blood for in vitro diagnostic testing.

The Steripath® Gen2 Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion.

Additionally, components of the system may be used for infusion following sample collection after disconnection of the Initial Specimen Diversion Device® (ISDD®). Venipuncture needles are indicated for short term infusion (less than 2 hours).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K192247: Steripath Gen2 Blood Collection System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K192247

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026