510(k) Report

510(k) · Cleared

Cleared

Pounce Thrombus Retrieval System

K192814510(k) TraditionalCleared Sep 21, 2020356-day reviewClass II21 CFR 870.5150

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Surmodics,Inc. · QEW Peripheral Mechanical Thrombectomy With Aspiration

Cleared on Sep 21, 2020 as substantially equivalent to K110315 (Nexgen Medical Systems, Inc., 2011) after a 356-day review. Its family tree reaches back 3 generations to 1997. 5 later devices descend from it.

Decided
Sep 21, 2020
Review
356 days
Class
Class II
Product code
QEW
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K970657 (1997). 7 ancestors in all, the oldest K892410 (1989). 5 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor1989: 1 other ancestor2005: 1 other ancestor2000: 1 other ancestor2021: 1 descendant cleared2022: 1 descendant cleared2023: 1 descendant cleared2024: 1 descendant cleared2026: 1 descendant clearedReference device: K102925 · Nexgen Expandable Catheter · Nexgen Medical Systems, Inc. · 2011K970657 · Lmi Thrombectomy Catheter · Lucas Medical, Inc. · 1997K070403 · Pronto .035 Extraction Catheter · Vascular Solutions, Inc. · 2007K110315 · Nexgen Peripheral Mechanical Retrieval Device · Nexgen Medical Systems, Inc. · 20111990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 2 descendants4 generations forward: 1 descendant K110315 · Nexgen Peripheral Mechanical Retrieval Device · Nexgen Medical Systems, Inc. · 2011K070403 · Pronto .035 Extraction Catheter · Vascular Solutions, Inc. · 2007K970657 · Lmi Thrombectomy Catheter · Lucas Medical, Inc. · 19970
  1. Root · 3 generations backK970657 Lmi Thrombectomy CatheterLucas Medical, Inc. · 1997
  2. 2 generations backK070403 Pronto .035 Extraction CatheterVascular Solutions, Inc. · 2007
  3. 1 generation backK110315 Nexgen Peripheral Mechanical Retrieval DeviceNexgen Medical Systems, Inc. · 2011
  4. This deviceK192814 Pounce Thrombus Retrieval SystemSurmodics,Inc. · 2020
  5. Descendant · cites it directly · 4 of its ownK211492 Pounce Thrombectomy SystemSurmodics,Inc. · 2021
  6. Descendant · 2 generations on · 3 of its ownK220501 Pounce Thrombectomy SystemSurmodics,Inc. · 2022
  7. Descendant · 3 generations on · 2 of its ownK231022 Pounce™ Thrombectomy SystemSurmodics,Inc. · 2023
  8. Descendant · 4 generations on · 0 of its ownK260599 INDIGO® Aspiration System – INDIGO LinkPenumbra, Inc. · 2026
  9. Descendant · 3 generations on · 0 of its ownK241362 Pounce XL Thrombectomy System (PTS-1011-7F135)Surmodics,Inc. · 2024

Most influential descendants

Showing 5 of 5 · See all
  • K211492Pounce Thrombectomy SystemSurmodics,Inc. · 2021 · cites it directly4descendants
  • K220501Pounce Thrombectomy SystemSurmodics,Inc. · 2022 · 2 generations on3descendants
  • K231022Pounce™ Thrombectomy SystemSurmodics,Inc. · 2023 · 3 generations on2descendants
  • K260599INDIGO® Aspiration System – INDIGO LinkPenumbra, Inc. · 2026 · 4 generations on0descendants
  • K241362Pounce XL Thrombectomy System (PTS-1011-7F135)Surmodics,Inc. · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K970657Lmi Thrombectomy CatheterLucas Medical, Inc.1997—
−2K070403Pronto .035 Extraction CatheterVascular Solutions, Inc.2007—
−1K110315Nexgen Peripheral Mechanical Retrieval DeviceNexgen Medical Systems, Inc.2011—
0 (this device)K192814Pounce Thrombus Retrieval SystemSurmodics,Inc.2020—
+1K211492Pounce Thrombectomy SystemSurmodics,Inc.2021—
+2K220501Pounce Thrombectomy SystemSurmodics,Inc.2022—
+3K231022Pounce™ Thrombectomy SystemSurmodics,Inc.2023—
+4K260599INDIGO® Aspiration System – INDIGO LinkPenumbra, Inc.2026—
+3K241362Pounce XL Thrombectomy System (PTS-1011-7F135)Surmodics,Inc.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
200010
200510
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202210
202310
202410
202610
FIG. 1 — Lineage of K192814, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in QEW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QEW device

  1. #1 K213402 Triever24, Triever20score 28.6 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K110315Nexgen Peripheral Mechanical Retrieval Device
Summary page 4
…iew Panel: Cardiovascular Predicate Devices: 1. Predicate Device: NexGen Peripheral Mechanical Retrieval Device (K110315) 2. Reference device: NexGen Peripheral Expandable Catheter (K102925…
Nexgen Medical Systems, Inc.2011—
†K102925Nexgen Expandable CatheterNexgen Medical Systems, Inc.2011—

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211492Pounce Thrombectomy SystemSurmodics,Inc.20214

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 8.9 years on median.
  • 2 of the 7 ancestors have a recall of any class.
  • Of its descendants, K211492 (Surmodics,Inc., 2021) has gone furthest, with 4 descendants of its own.

How does its category compare?

All QEW devices

QEW Peripheral Mechanical Thrombectomy With Aspiration: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · QEW

14in 2025

2008: 22009: 22012: 12013: 22014: 22015: 42016: 42017: 22018: 142019: 82020: 162021: 112022: 122023: 232024: 142025: 142026 (partial year): 10
2007high 23 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QEW

58in 2025

2015: 88 days median2016: 63 days median2018: 61 days median2019: 51 days median2020: 105 days median2021: 60 days median2022: 140 days median2023: 102 days median2024: 65 days median2025: 58 days median2026 (partial year): 74 days median
2007high 140 (2022) · low 51 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · QEW

7in 2022

2022: 7, 7 Class I
2007high 7 (2022) · low 0 (2007)2026

Class I (7 since 2007) Class II/III not classified

Adverse event reports per year · QEW

207in 2025

2019: 332020: 1972021: 3222022: 1252023: 1652024: 1622025: 207
2007high 322 (2021) · low 0 (2007)2025

MAUDE reports naming product code QEW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—00
20092—00
20100—00
20110—00
20121—00
20132—00
20142—00
201548800
201646300
20172—00
2018146100
2019851033
2020161050197
202111600322
2022121407 (7)125
2023231020165
202414650162
202514580207
202610740—

What is it indicated for?

The Thrombus Retrieval System is intended for the non-surgical removal of thrombi and emboli from the arterial peripheral vasculature.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K192814: Pounce Thrombus Retrieval System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K192814

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026