# K192872: ulrichINJECT CT Motion

- Pathway: 510(k) (Traditional)
- Applicant: Ulrich GmbH & Co. KG
- Decided: Jul 2, 2020 after 268 days
- Product code: IZQ (Injector, Contrast Medium, Automatic)
- Device class: 2
- Regulation: 21 CFR 870.1650
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 5 generations back to K934086 (1995). 5 ancestors in all.
5 later devices descend from it; 2 cite it directly.
0 of 11 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K171392: ulrichINJECT CT motion, Ulrich GmbH & Co. KG (2017) Summary page 4.

## Most influential descendants
- K233737: ulricheasyINJECT Max 2M (XD 10140); ulricheasyINJECT Max 3 (XD 10150); ulricheasyINJECT Max 3 (XD 10180) (2024), 2 descendants of its own
- K241850: ulricheasyINJECT Max 2M (XD 10140); ulricheasyINJECT Max 3 (XD 10150); ulricheasyINJECT Max 3 (XD 10180) (2024), 1 descendant of its own
- K210541: ulrichINJECT CT motion (2021), 1 descendant of its own
- K252281: ulricheasyINJECT Max 2M (XD 10140); ulricheasyINJECT Max 3 (XD 10150); ulricheasyINJECT Max 3 (XD 10180) (2025), 0 descendants of its own
- K251295: ulrichINJECT CT Motion (XD 8000) (2025), 0 descendants of its own

## Indications for use
ulrichINJECT CT motion is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCl), to human subjects undergoing diagnostic examinations in computed tomography (CT) applications. ulrichINJECT CT motion is specifically indicated for use in CT procedures for the delivery of Omnipaque™ (Iohexol) Injection, solution - GE Healthcare Inc. contrast media as supplied in Imaging Bulk Packages (IBP), Omnipaque™ (Iohexol) Injection, solution – GE Healthcare Inc., and Visipaque™ (iodixanol) Injection - GE Healthcare Inc. contrast media as supplied in single dose bottles. Pump tubing-flex is used for a maximum time of twenty four (24) hours. When used with Omnipaque™ IBP, Omnipaque™ single dose bottles, or Visipaque™ single dose bottles, a maximum of 19 bottles of contrast media can be used or maximum time of twenty four (24) hours of Pump tubing-flex, or whichever comes first. Time per contrast media or saline container depends on each contrast media's or saline's use time expiration with a maximum of eight (8) hours per contrast media or saline container. Spike for CT disposable is for single-bottle use only and must be discarded with the media container. The Patient tubing must be discarded after each patient procedure. ulrichINJECT CT motion is to be used only by and under quasi-continuous supervision of trained healthcare professionals in an appropriate licensed healthcare facility, in a room designated for radiological procedures that involve intravascular administration of contrast agent.

Source page: https://510kreport.com/device/K192872
Methodology: https://510kreport.com/methodology
