# K193215: Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems

- Pathway: 510(k) (Abbreviated)
- Applicant: Philips Medical Systems Nederland B.V.
- Decided: Apr 10, 2020 after 141 days
- Product code: LNH (System, Nuclear Magnetic Resonance Imaging)
- Device class: 2
- Regulation: 21 CFR 892.1000
- Recalls of this device: 152 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 8 generations back to K031815 (2003). 13 ancestors in all.
11 later devices descend from it; 6 cite it directly.
4 of 25 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K183063: Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems, Philips Medical Systems Nederland B.V. (2019) Summary page 5.
- K162940: MultiBand SENSE, Philips Medical Systems Nederland B.V. (2016) Summary page 5.
- K173079: Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4, Philips Medical Systems Nederland B.V. (2018) Summary page 6.

## Most influential descendants
- K212673: MR 5300 (2021), 6 descendants of its own
- K213583: Achieva, Ingenia, Ingenia CX, Ingenia Elition and Ingenia Ambition MR Systems (2022), 5 descendants of its own
- K213516: Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei (2022), 5 descendants of its own
- K223442: MR 5300 and MR 7700 R11 MR Systems (2022), 4 descendants of its own
- K223458: Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems (2023), 2 descendants of its own

## Recalls of this device
- Aug 28, 2026 · Class II · Z-3311-2026: When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is u
- Aug 28, 2026 · Class II · Z-3310-2026: When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is u
- Aug 28, 2026 · Class II · Z-3308-2026: When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is u
- Aug 21, 2026 · Class II · Z-3239-2026: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system me
- Aug 21, 2026 · Class II · Z-3237-2026: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system me
- Aug 21, 2026 · Class II · Z-3232-2026: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system me
- Aug 21, 2026 · Class II · Z-3243-2026: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system me
- Aug 21, 2026 · Class II · Z-3233-2026: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system me
- Aug 21, 2026 · Class II · Z-3236-2026: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system me
- Aug 21, 2026 · Class II · Z-3238-2026: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system me
- Aug 21, 2026 · Class II · Z-3241-2026: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system me
- Apr 14, 2026 · Class II · Z-1954-2026: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings 
- Apr 14, 2026 · Class II · Z-1968-2026: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings 
- Apr 14, 2026 · Class II · Z-1956-2026: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings 
- Apr 14, 2026 · Class II · Z-1965-2026: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings 
- Apr 14, 2026 · Class II · Z-1960-2026: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings 
- Apr 14, 2026 · Class II · Z-1953-2026: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings 
- Apr 14, 2026 · Class II · Z-1951-2026: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings 
- Apr 14, 2026 · Class II · Z-1967-2026: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings 
- Apr 14, 2026 · Class II · Z-1961-2026: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings 
- Dec 3, 2025 · Class II · Z-1208-2026: The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Dec 3, 2025 · Class II · Z-1212-2026: The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Dec 3, 2025 · Class II · Z-1210-2026: The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Dec 3, 2025 · Class II · Z-1207-2026: The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Dec 3, 2025 · Class II · Z-1211-2026: The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Dec 3, 2025 · Class II · Z-1206-2026: The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Dec 3, 2025 · Class II · Z-1209-2026: The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Jun 30, 2025 · Class II · Z-2207-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2208-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2206-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2201-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2210-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2205-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2211-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2212-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2209-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2200-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Jun 30, 2025 · Class II · Z-2204-2025: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- May 29, 2025 · Class II · Z-1929-2025: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a 
- May 29, 2025 · Class II · Z-1932-2025: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a 
- May 29, 2025 · Class II · Z-1935-2025: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a 
- May 29, 2025 · Class II · Z-1933-2025: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a 
- May 29, 2025 · Class II · Z-1931-2025: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a 
- May 29, 2025 · Class II · Z-1934-2025: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a 
- May 29, 2025 · Class II · Z-1928-2025: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a 
- May 29, 2025 · Class II · Z-1930-2025: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a 
- May 13, 2025 · Class II · Z-1795-2025: Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
- May 13, 2025 · Class II · Z-1797-2025: Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
- May 13, 2025 · Class II · Z-1803-2025: Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
- May 13, 2025 · Class II · Z-1798-2025: Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.

## Indications for use
Philips Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. This MR system enables trained physicians to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin. Image appearance is determined by many different physical properties of the tissue and the anatomy, the MR scan technique applied, and presence of contrast agents. The use of contrast agents for diagnostic imaging applications should be performed consistent with the approved labeling for the contrast agent. The trained clinical user can adjust the MR scan parameters to customize image appearance, accelerate image acquisition, and synchronize with the patient’s breathing or cardiac cycle. The systems can use combinations of images to produce physical parameters, and related derived images. Images, spectra, and measurements of physical parameters, when interpreted by a trained physician, provide information that may assist diagnosis and therapy planning. The accuracy of determined physical parameters depends on system and scan parameters, and must be controlled and validated by the clinical user. In addition the Philips MR systems provide imaging capabilities, such as MR fluoroscopy, to guide and evaluate interventional and minimally invasive procedures in the head, body and extremities. MR Interventional procedures, performed inside or adjacent to the Philips MR system, must be performed with MR Conditional or MR Safe instrumentation as selected and evaluated by the clinical user for use with the specific MR system configuration in the hospital. The appropriateness and use of information from a Philips MR system for a specific interventional procedure and specific MR system configuration must be validated by the clinical user.