510(k) · Cleared
ClearedTyber Medical Anatomical Plating System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Tyber Medical, LLC · HRS Plate, Fixation, Bone
Cleared on Jan 20, 2020 as substantially equivalent to K162674 (Biopro, Inc., 2017) after a 59-day review. Its family tree reaches back 13 generations to 1996. 11 later devices descend from it.
- Decided
- Jan 20, 2020
- Review
- 59 days
- Class
- Class II
- Product code
- HRS
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 13 generations to K962616 (1996). 207 ancestors in all, the oldest K792022 (1979). 11 later devices descend from it, 4 citing it directly.
- Root · 13 generations backK962616 Synthes Dorsal Distal Radius Plate (DDRP) SystemSynthes (Usa) · 1996
- 9 generations between
- 3 generations backK140376 Variax 2 SystemStryker Trauma AG · 2014
- 2 generations backK141992 Variax 2 SystemStryker Trauma AG · 2014
- 1 generation backK162674 BioPro Foot Plating SystemsBiopro, Inc. · 2017
- This deviceK193222 Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2020
- Descendant · cites it directly · 10 of its ownK203817 Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2021
- Descendant · cites it directly · 9 of its ownK222465 Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2023
- Descendant · cites it directly · 3 of its ownK232652 Anatomical Plating System (various)Tyber Medical, LLC · 2023 · Recalled
- Descendant · 2 generations on · 2 of its ownK241218 Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2024
- Descendant · 2 generations on · 2 of its ownK233423 Tyber Medical Anatomical Plating System – Line extension to the…Tyber Medical, LLC · 2023
Show all 13 generations
- Root · 13 generations backK962616 Synthes Dorsal Distal Radius Plate (DDRP) SystemSynthes (Usa) · 1996
- 12 generations backK982732 Synthes (Usa) Distal Radius Plate System (DRPS)Synthes · 1998
- 11 generations backK023007 Distal Radius Fracture Repair SystemHand Innovations, Inc. · 2002
- 10 generations backK030198 Hand Innovations, Inc. Distal Radius Fracture Repair SystemHand Innovations, Inc. · 2003
- 9 generations backK032705 Distal Radius Fracture Repair SystemHand Innovations, Inc. · 2003
- 8 generations backK050932 Distal Volar Radius Anatomical Plate SystemHand Innovations, Inc. · 2005 · Recalled
- 7 generations backK051735 Peri-Loc Locking Bone Plates and Locking Bone Screws for the Upper…Smith & Nephew, Inc. · 2005
- 6 generations backK071563 Peri-Loc Periarticular Locked Plating SystemSmith & Nephew, Inc. · 2007
- 5 generations backK090675 Smith & Nephew, Inc. VLP Foot Plating, Screw System and AccessoriesSmith & Nephew, Inc. · 2009
- 4 generations backK100776 Synthes 2.4 MM/2.7 MM Variable Angle LCP Forefoot/Midfoot SystemSynthes (Usa) · 2010
- 3 generations backK140376 Variax 2 SystemStryker Trauma AG · 2014
- 2 generations backK141992 Variax 2 SystemStryker Trauma AG · 2014
- 1 generation backK162674 BioPro Foot Plating SystemsBiopro, Inc. · 2017
- This deviceK193222 Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2020
Showing 5 of 11 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 11 · See all- K203817Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2021 · cites it directly10descendants
- K222465Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2023 · cites it directly9descendants
- K232652Anatomical Plating System (various)Tyber Medical, LLC · 2023 · cites it directly3descendants
- K241218Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2024 · 2 generations on2descendants
- K233423Tyber Medical Anatomical Plating System – Line extension to the Mini-Frag SystemTyber Medical, LLC · 2023 · 2 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −13 | K962616 | Synthes Dorsal Distal Radius Plate (DDRP) System | Synthes (Usa) | 1996 | — |
| −12 | K982732 | Synthes (Usa) Distal Radius Plate System (DRPS) | Synthes | 1998 | — |
| −11 | K023007 | Distal Radius Fracture Repair System | Hand Innovations, Inc. | 2002 | — |
| −10 | K030198 | Hand Innovations, Inc. Distal Radius Fracture Repair System | Hand Innovations, Inc. | 2003 | — |
| −9 | K032705 | Distal Radius Fracture Repair System | Hand Innovations, Inc. | 2003 | — |
| −8 | K050932 | Distal Volar Radius Anatomical Plate System | Hand Innovations, Inc. | 2005 | 1 |
| −7 | K051735 | Peri-Loc Locking Bone Plates and Locking Bone Screws for the Upper Extremity | Smith & Nephew, Inc. | 2005 | — |
| −6 | K071563 | Peri-Loc Periarticular Locked Plating System | Smith & Nephew, Inc. | 2007 | — |
| −5 | K090675 | Smith & Nephew, Inc. VLP Foot Plating, Screw System and Accessories | Smith & Nephew, Inc. | 2009 | — |
| −4 | K100776 | Synthes 2.4 MM/2.7 MM Variable Angle LCP Forefoot/Midfoot System | Synthes (Usa) | 2010 | — |
| −3 | K140376 | Variax 2 System | Stryker Trauma AG | 2014 | — |
| −2 | K141992 | Variax 2 System | Stryker Trauma AG | 2014 | — |
| −1 | K162674 | BioPro Foot Plating Systems | Biopro, Inc. | 2017 | — |
| 0 (this device) | K193222 | Tyber Medical Anatomical Plating System | Tyber Medical, LLC | 2020 | — |
| +1 | K203817 | Tyber Medical Anatomical Plating System | Tyber Medical, LLC | 2021 | — |
| +1 | K222465 | Tyber Medical Anatomical Plating System | Tyber Medical, LLC | 2023 | — |
| +1 | K232652 | Anatomical Plating System (various) | Tyber Medical, LLC | 2023 | 20 |
| +2 | K241218 | Tyber Medical Anatomical Plating System | Tyber Medical, LLC | 2024 | — |
| +2 | K233423 | Tyber Medical Anatomical Plating System – Line extension to the Mini-Frag System | Tyber Medical, LLC | 2023 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1979 | 1 | 0 |
| 1980 | 1 | 0 |
| 1981 | 1 | 0 |
| 1988 | 1 | 0 |
| 1989 | 3 | 0 |
| 1990 | 3 | 0 |
| 1991 | 2 | 0 |
| 1992 | 5 | 0 |
| 1993 | 3 | 0 |
| 1994 | 6 | 0 |
| 1995 | 5 | 0 |
| 1996 | 13 | 0 |
| 1997 | 2 | 0 |
| 1998 | 7 | 0 |
| 1999 | 10 | 0 |
| 2000 | 13 | 0 |
| 2001 | 13 | 0 |
| 2002 | 13 | 0 |
| 2003 | 6 | 0 |
| 2004 | 9 | 0 |
| 2005 | 13 | 0 |
| 2006 | 12 | 0 |
| 2007 | 9 | 0 |
| 2008 | 12 | 0 |
| 2009 | 10 | 0 |
| 2010 | 11 | 0 |
| 2011 | 3 | 0 |
| 2012 | 5 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2021 | 1 | 0 |
| 2023 | 5 | 0 |
| 2024 | 1 | 0 |
| 2025 | 2 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HRS device
- #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K162674 | BioPro Foot Plating SystemsSummary page 5…aded metallic bone fixation fastener (21 CFR 888.3040) Classification Panel: Orthopedic Product Code HRS/HWC Predicate Devices BioPro Foot Plating Systems – K162674 Device Description The Tyber Medical Anatomical Plating System consists of the following categories:… | 2017 |
What descends from it? (11)
See all 114 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 24 years and passes through 7 companies; a new generation was cleared every 1.8 years on median.
- 44 of the 207 ancestors have a recall of any class.
- Of its descendants, K203817 (Tyber Medical, LLC, 2021) has gone furthest, with 10 descendants of its own.
- 2 other devices also cite its predicate K162674.
How does its category compare?
All HRS devicesHRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.
FDA decisions per year · HRS
54in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HRS
90in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HRS
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HRS
2,306in 2025
MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 35 | 63 | 2 | 339 |
| 2008 | 35 | 87 | 11 | 311 |
| 2009 | 47 | 85 | 3 | 435 |
| 2010 | 49 | 100 | 16 | 491 |
| 2011 | 43 | 88 | 5 | 773 |
| 2012 | 59 | 106 | 4 | 1,663 |
| 2013 | 56 | 89 | 11 | 2,643 |
| 2014 | 76 | 57 | 0 | 2,413 |
| 2015 | 56 | 97 | 7 | 3,095 |
| 2016 | 44 | 109 | 5 | 2,276 |
| 2017 | 57 | 122 | 3 | 2,220 |
| 2018 | 40 | 116 | 2 | 2,663 |
| 2019 | 46 | 89 | 25 | 3,359 |
| 2020 | 48 | 88 | 1 | 2,790 |
| 2021 | 40 | 87 | 9 | 2,442 |
| 2022 | 47 | 144 | 3 | 1,700 |
| 2023 | 40 | 120 | 1 | 2,049 |
| 2024 | 54 | 96 | 0 | 1,849 |
| 2025 | 54 | 90 | 1 | 2,306 |
| 2026 | 51 | 82 | 3 | — |
What is it indicated for?
The intended use of the Tyber Medical Anatomical Plating System is to draw two or more aligned bone fragments together to facilitate healing in an adult patient and is composed of the following bone plate categories: I. Forefoot System: The Tyber Medical Forefoot Plating System is Indicated for Use in fixation of small bones and small bone fragments in the foot (Phalanges and Metatarsals) for stabilization of fractures, joint fusions, osteotomies…
Read the full statement (268 words)
The intended use of the Tyber Medical Anatomical Plating System is to draw two or more aligned bone fragments together to facilitate healing in an adult patient and is composed of the following bone plate categories:
I. Forefoot System:
The Tyber Medical Forefoot Plating System is Indicated for Use in fixation of small bones and small bone fragments in the foot (Phalanges and Metatarsals) for stabilization of fractures, joint fusions, osteotomies, nonunions, malunions, reconstruction of small bones, revision surgeries and replantations in an adult patient.
The Tyber Medical Forefoot System is not for Spinal Use.
II. Mid & Hindfoot System:
The Tyber Medical Mid & Hindfoot Plating System is Indicated for Use in fixation of medium/large bones and medium/large bone multi-fragments in the foot (Cuneiform, Cuboid, Navicular, Talus and Calcaneus) for stabilization of fractures, joint fusions, osteotomies, nonunions, malunions, reconstruction of medium/large bones, revision surgeries and replantations in an adult patient.
The Tyber Medical Mid & Hindfoot System is not for Spinal Use.
III. Ankle Fracture System:
The Tyber Medical Ankle Fracture System is Indicated for Use in:
1). Fixation of fractures of the distal tibia included, but not limited to, ankle fractures, perarticular fractures, corrective osteotomies, non-unions, intra- and extra- articular and distal tibia fractures with a shaft extension, and malleolar fractures;
2). In intra- and extra articular fractures, osteotomies, medial malleolar fractures and non-unions of the metaphyseal and diaphyseal region of the distal fibula;
3). In distal tibia/fibula long bones which include the metaphyseal and diaphyseal regions of the tibia and fibula in the ankle.
The Tyber Medical Ankle Fracture System is not for Spinal Use.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K193222 · printed Nov 4, 2026
Cite this page
510(k) Report. “K193222: Tyber Medical Anatomical Plating System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K193222
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026