510(k) Report

510(k) · Cleared

Cleared

Arthrex DynaNite Compression Plate

K193345510(k) TraditionalCleared Jul 8, 2020218-day reviewClass II21 CFR 888.3030

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Arthrex, Inc. · HRS Plate, Fixation, Bone

Cleared on Jul 8, 2020 as substantially equivalent to K161303 (Mx Orthopedics, Corp., 2017) and 1 other predicate after a 218-day review. Its family tree reaches back 7 generations to 1997.

Decided
Jul 8, 2020
Review
218 days
Class
Class II
Product code
HRS
Recalls
None
Predicates
2

Its longest line of descent runs back 7 generations to K964226 (1997). 20 ancestors in all, the oldest K895117 (1990). No later device cites it as a predicate yet.

  1. Root · 7 generations backK964226 Memory Staple (20 Sizes Available)Landos, Inc. · 1997
  2. 6 generations backK991482 Uni-Clip StapleNew Deal, S.A. · 1999
  3. 5 generations backK011716 The Uni-Clip StapleNew Deal, S.A. · 2001 · Recalled
  4. 4 generations backK061594 Large Uni-Clip StapleNew Deal, S.A. · 2006
  5. 3 generations backK070447 Newdeal Compression PlateNewdeal Sas · 2007 · Recalled
  6. 2 generations backK091609 Newdeal Compression PlateNewdeal Sas · 2009 · Recalled
  7. 1 generation backK161303 dynaMX™ Compression PlateMx Orthopedics, Corp. · 2017
  8. This deviceK193345 Arthrex DynaNite Compression PlateArthrex, Inc. · 2020
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K964226Memory Staple (20 Sizes Available)Landos, Inc.1997—
−6K991482Uni-Clip StapleNew Deal, S.A.1999—
−5K011716The Uni-Clip StapleNew Deal, S.A.20011
−4K061594Large Uni-Clip StapleNew Deal, S.A.2006—
−3K070447Newdeal Compression PlateNewdeal Sas20072
−2K091609Newdeal Compression PlateNewdeal Sas20091
−1K161303dynaMX™ Compression PlateMx Orthopedics, Corp.2017—
0 (this device)K193345Arthrex DynaNite Compression PlateArthrex, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199020
199820
200020
200520
200610
200810
200910
201010
201510
FIG. 1 — Lineage of K193345, 1997–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HRS device

  1. #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K161303dynaMX™ Compression Plate
Summary page 4
…Name 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class II Predicate Device K161303: MXO dynaMX Compression Plate (Primary Predicate) K090047: Synthes (USA) 1.5mm Mini Fragment LCP System Ref…
Mx Orthopedics, Corp.2017—
*K090047Synthes (Usa) 1.5MM Mini Fragment LCP System
Summary page 4
…nces and accessories Regulatory Class II Predicate Device K161303: MXO dynaMX Compression Plate (Primary Predicate) K090047: Synthes (USA) 1.5mm Mini Fragment LCP System Reference Device K172052: Arthrex DynaNite Nitinol Staple Purpose of…
Synthes (Usa)2009—
†K172052Arthrex DynaNite Nitinol StapleArthrex, Inc.2017—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 2.3 years on median.
  • 4 of the 20 ancestors have a recall of any class.

How does its category compare?

All HRS devices

HRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · HRS

54in 2025

2007: 352008: 352009: 472010: 492011: 432012: 592013: 562014: 762015: 562016: 442017: 572018: 402019: 462020: 482021: 402022: 472023: 402024: 542025: 542026 (partial year): 51
2007high 76 (2014) · low 35 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HRS

90in 2025

2007: 63 days median2008: 87 days median2009: 85 days median2010: 100 days median2011: 88 days median2012: 106 days median2013: 89 days median2014: 57 days median2015: 97 days median2016: 109 days median2017: 122 days median2018: 116 days median2019: 89 days median2020: 88 days median2021: 87 days median2022: 144 days median2023: 120 days median2024: 96 days median2025: 90 days median2026 (partial year): 82 days median
2007high 144 (2022) · low 57 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HRS

1in 2025

2007: 22008: 112009: 32010: 162011: 52012: 42013: 112015: 72016: 52017: 32018: 22019: 252020: 12021: 92022: 32023: 12025: 12026 (partial year): 3
2007high 25 (2019) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HRS

2,306in 2025

2007: 3392008: 3112009: 4352010: 4912011: 7732012: 1,6632013: 2,6432014: 2,4132015: 3,0952016: 2,2762017: 2,2202018: 2,6632019: 3,3592020: 2,7902021: 2,4422022: 1,7002023: 2,0492024: 1,8492025: 2,306
2007high 3,359 (2019) · low 311 (2008)2025

MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735632339
2008358711311
200947853435
20104910016491
201143885773
20125910641,663
20135689112,643
2014765702,413
2015569773,095
20164410952,276
20175712232,220
20184011622,663
20194689253,359
2020488812,790
2021408792,442
20224714431,700
20234012012,049
2024549601,849
2025549012,306
202651823—

What is it indicated for?

The Arthrex DynaNite Compression Plate is intended to be used for fixation such as: LisFranc arthrodesis, mono or bi-cortical osteotomies in the forefoot, first metatarsophalangeal arthrodesis, Akin osteotomy, midfoot and hindfoot arthrodeses or osteotomies, fixation of osteotomies for hallux valgus treatment (Scarf and Chevron), and arthrodesis of the metatarsocuneiform joint to reposition and stabilize metatarsus primus varus.

The Arthrex DynaNite Compression Plate is intended to be used in conjunction with the Arthrex Compression FT Screws.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K193345: Arthrex DynaNite Compression Plate.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K193345

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026