# K200252: Getinge Assured MI Steam Migrating Integrator

- Pathway: 510(k) (Traditional)
- Applicant: Steritec Products Mfg Co, Inc.
- Decided: May 21, 2020 after 108 days
- Product code: JOJ (Indicator, Physical/Chemical Sterilization Process)
- Device class: 2
- Regulation: 21 CFR 880.2800
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K012195 (2002). 2 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 4 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K152630: VERIFY STEAM Integrating Indicator, STERIS Corporation (2016) Summary page 5.

## Most influential descendants
- K242795: Getinge Assured MI Steam Migrating Integrator (2024), 0 descendants of its own

## Indications for use
The Getinge Assured MI Steam Migrating Integrator strip is an internal pack integrating indicator designed to react to critical process parameters of a steam sterilization cycle within a stated tolerance. The integrating indicator is intended to be placed in each pack, pouch, tray or container to function as an independent monitor of critical parameters for the following sterilization cycles: Steam Sterilization Cycles: 250°F/121°C, 30 minutes Gravity 270°F/132°C, 4 minutes Dynamic Air Removal 270°F/132°C, 15 minutes Gravity 275°F/135°C, 3 minutes Dynamic Air Removal 275°F/135°C, 10 minutes Gravity Steam Sterilization Cycles (IUSS): 270°F/132°C, 4 minutes Dynamic Air Removal 270°F/132°C, 3 minutes Gravity 270°F/132°C, 10 minutes Gravity 275°F/135°C, 3 minutes Dynamic Air Removal 275°F/135°C, 3 minutes Gravity 275°F/135°C, 10 minutes Gravity Stated values (as determined in a steam resistometer): 30 minutes at 121°C 9.1 minutes at 128°C 3.3 minutes at 132°C 1.5 minutes at 135°C

Source page: https://510kreport.com/device/K200252
Methodology: https://510kreport.com/methodology