510(k) Report

510(k) · Cleared

Cleared

Huber Needle Infusion Set, Safety Huber Needle Infusion Set

K200463510(k) TraditionalCleared Apr 7, 2021407-day reviewClass II21 CFR 880.5570

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jiangsu Caina Medical Co.,Ltd · PTI Non-Coring (Huber) Needle

Cleared on Apr 7, 2021 as substantially equivalent to K170897 (B.Braun Medical, Inc., 2017) after a 407-day review. Its family tree reaches back 10 generations to 2000. 2 later devices descend from it.

Decided
Apr 7, 2021
Review
407 days
Class
Class II
Product code
PTI
Recalls
None
Predicates
1

Its longest line of descent runs back 10 generations to K001901 (2000). 37 ancestors in all, the oldest K863606 (1986). 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2004: 3 other ancestors2002: 3 other ancestors2006: 1 other ancestor2005: 2 other ancestors2001: 2 other ancestors2000: 1 other ancestor1996: 1 other ancestor1994: 3 other ancestors1988: 1 other ancestor1987: 2 other ancestors1992: 1 other ancestor1991: 1 other ancestor1989: 4 other ancestors1986: 1 other ancestor2021: 1 descendant cleared2026: 1 descendant clearedK001901 · Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline Catheter, Model 3134135 · C.R. Bard, Inc. · 2000 · recalledK012902 · Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath Dual Lumen Picc Catheter, Model 3246100 · C.R. Bard, Inc. · 2001K031129 · Poly Per-Q-Cath Picc Catheters; Poly Radpicc · C.R. Bard, Inc. · 2003K034019 · Poly Per-Q-Cath Picc Catheters · C.R. Bard, Inc. · 2004 · recalledK050931 · 6 FR DL Powerpicc Catheter · C.R. Bard, Inc. · 2005K051672 · Powerpicc · Bard Access Systems, Inc. · 2005 · recalledK060812 · Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex Catheter-Multiple · C.R. Bard, Inc. · 2006 · recalledK062414 · Vortex CT Port Access System and Lifeguard Safety Infusion Set · Rita Medical Systems, Inc. · 2007 · recalledK073050 · Safestep Max Power-Injectable Infusion Set · Specialized Health Products International, Inc. · 2008 · recalledK170897 · Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety II Needle with Caresite Luer Access Device and Y site · B.Braun Medical, Inc. · 20171990200020102020
  1. Root · 10 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
  2. 6 generations between
  3. 3 generations backK062414 Vortex CT Port Access System and Lifeguard Safety Infusion SetRita Medical Systems, Inc. · 2007 · Recalled
  4. 2 generations backK073050 Safestep Max Power-Injectable Infusion SetSpecialized Health Products International, Inc. · 2008 · Recalled
  5. 1 generation backK170897 Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety…B.Braun Medical, Inc. · 2017
  6. This deviceK200463 Huber Needle Infusion Set, Safety Huber Needle Infusion SetJiangsu Caina Medical Co.,Ltd · 2021
  7. Descendant · cites it directly · 0 of its ownK254095 25G Huber NeedleMartech Medical Products · 2026
  8. Descendant · cites it directly · 0 of its ownK211121 Intera Non-coring (Huber) Refill Needles, Intera Non-coring (Huber)…Intera Oncology, Inc. · 2021
Show all 10 generations
  1. Root · 10 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
  2. 9 generations backK012902 Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly…C.R. Bard, Inc. · 2001
  3. 8 generations backK031129 Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc. · 2003
  4. 7 generations backK034019 Poly Per-Q-Cath Picc CathetersC.R. Bard, Inc. · 2004 · Recalled
  5. 6 generations backK050931 6 FR DL Powerpicc CatheterC.R. Bard, Inc. · 2005
  6. 5 generations backK051672 PowerpiccBard Access Systems, Inc. · 2005 · Recalled
  7. 4 generations backK060812 Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex…C.R. Bard, Inc. · 2006 · Recalled
  8. 3 generations backK062414 Vortex CT Port Access System and Lifeguard Safety Infusion SetRita Medical Systems, Inc. · 2007 · Recalled
  9. 2 generations backK073050 Safestep Max Power-Injectable Infusion SetSpecialized Health Products International, Inc. · 2008 · Recalled
  10. 1 generation backK170897 Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety…B.Braun Medical, Inc. · 2017
  11. This deviceK200463 Huber Needle Infusion Set, Safety Huber Needle Infusion SetJiangsu Caina Medical Co.,Ltd · 2021

Most influential descendants

Showing 2 of 2 · See all
  • K25409525G Huber NeedleMartech Medical Products · 2026 · cites it directly0descendants
  • K211121Intera Non-coring (Huber) Refill Needles, Intera Non-coring (Huber) Special bolus…Intera Oncology, Inc. · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K001901Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline…C.R. Bard, Inc.20002
−9K012902Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath…C.R. Bard, Inc.2001—
−8K031129Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc.2003—
−7K034019Poly Per-Q-Cath Picc CathetersC.R. Bard, Inc.20041
−6K0509316 FR DL Powerpicc CatheterC.R. Bard, Inc.2005—
−5K051672PowerpiccBard Access Systems, Inc.20051
−4K060812Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex Catheter-MultipleC.R. Bard, Inc.20061
−3K062414Vortex CT Port Access System and Lifeguard Safety Infusion SetRita Medical Systems, Inc.20072
−2K073050Safestep Max Power-Injectable Infusion SetSpecialized Health Products International, Inc.20082
−1K170897Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety II Needle…B.Braun Medical, Inc.2017—
0 (this device)K200463Huber Needle Infusion Set, Safety Huber Needle Infusion SetJiangsu Caina Medical Co.,Ltd2021—
+1K25409525G Huber NeedleMartech Medical Products2026—
+1K211121Intera Non-coring (Huber) Refill Needles, Intera Non-coring (Huber) Special…Intera Oncology, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198720
198810
198940
199110
199210
199430
199610
200010
200120
200230
200430
200520
200610
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202610
FIG. 1 — Lineage of K200463, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PTI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PTI device

  1. #1 K254095 25G Huber Needlescore 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K170897Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety II Needle with…
Summary page 6
…Huber Needle Safety Huber Needle Remake K170897 Infusion Set Infusion Set Product Code PTI PTI…
B.Braun Medical, Inc.2017—

Also cleared against K170897: 1 other device →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K25409525G Huber NeedleMartech Medical Products20260
K211121Intera Non-coring (Huber) Refill Needles, Intera Non-coring (Huber) Special bolus…Intera Oncology, Inc.20210

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 6 companies; a new generation was cleared every 12 months on median.
  • 21 of the 37 ancestors have a recall of any class.
  • Of its descendants, K254095 (Martech Medical Products, 2026) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K170897.

How does its category compare?

All PTI devices

PTI Non-Coring (Huber) Needle: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · PTI

3in 2025

2017: 52021: 22022: 12023: 12024: 12025: 32026 (partial year): 1
2007high 5 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PTI

120in 2025

2017: 221 days median2025: 120 days median
2007high 221 (2017) · low 120 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PTI

No recall names PTI since 2007.

Adverse event reports per year · PTI

33in 2025

2017: 22018: 32019: 122020: 132021: 172022: 132023: 92024: 362025: 33
2007high 36 (2024) · low 0 (2007)2025

MAUDE reports naming product code PTI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
2017522102
20180—03
20190—012
20200—013
20212—017
20221—013
20231—09
20241—036
20253120033
20261—0—

What is it indicated for?

Huber Needle Infusion Set The Huber Needle Infusion Set is a device with a non-coring right angle needle intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids and drugs, as well as blood sampling into the port. The 19G-22G needles of device are also suitable for power injection of contrast media to a maximum pressure of 325psi. When used with ports indicated for power injection, the maximum recommended…

Read the full statement (199 words)

Huber Needle Infusion Set

The Huber Needle Infusion Set is a device with a non-coring right angle needle intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids and drugs, as well as blood sampling into the port.

The 19G-22G needles of device are also suitable for power injection of contrast media to a maximum pressure of 325psi. When used with ports indicated for power injection, the maximum recommended infusion rate is approximately 5ml/s for 19G and 20G, 2ml/s for 22G.

Safety Huber Needle Infusion Set

The Safety Huber Needle Infusion Set is a device with a non-coring right angle needle intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids and drugs, as well as blood sampling into the port. The safety feature is manually activated during needle removal, and is designed to aid in the prevention of accidental needle sticks.

The 19G-22G needles of device are also suitable for power injection of contrast media to a maximum pressure of 325psi. When used with ports indicated for power injection, the maximum recommended infusion rate is approximately 5ml/s for 19G and 20G, 2ml/s for 22G.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K200463: Huber Needle Infusion Set, Safety Huber Needle Infusion Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K200463

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026