510(k) · Cleared
ClearedHuber Needle Infusion Set, Safety Huber Needle Infusion Set
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Jiangsu Caina Medical Co.,Ltd · PTI Non-Coring (Huber) Needle
Cleared on Apr 7, 2021 as substantially equivalent to K170897 (B.Braun Medical, Inc., 2017) after a 407-day review. Its family tree reaches back 10 generations to 2000. 2 later devices descend from it.
- Decided
- Apr 7, 2021
- Review
- 407 days
- Class
- Class II
- Product code
- PTI
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 10 generations to K001901 (2000). 37 ancestors in all, the oldest K863606 (1986). 2 later devices descend from it, 2 citing it directly.
- Root · 10 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
- 6 generations between
- 3 generations backK062414 Vortex CT Port Access System and Lifeguard Safety Infusion SetRita Medical Systems, Inc. · 2007 · Recalled
- 2 generations backK073050 Safestep Max Power-Injectable Infusion SetSpecialized Health Products International, Inc. · 2008 · Recalled
- 1 generation backK170897 Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety…B.Braun Medical, Inc. · 2017
- This deviceK200463 Huber Needle Infusion Set, Safety Huber Needle Infusion SetJiangsu Caina Medical Co.,Ltd · 2021
- Descendant · cites it directly · 0 of its ownK254095 25G Huber NeedleMartech Medical Products · 2026
- Descendant · cites it directly · 0 of its ownK211121 Intera Non-coring (Huber) Refill Needles, Intera Non-coring (Huber)…Intera Oncology, Inc. · 2021
Show all 10 generations
- Root · 10 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
- 9 generations backK012902 Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly…C.R. Bard, Inc. · 2001
- 8 generations backK031129 Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc. · 2003
- 7 generations backK034019 Poly Per-Q-Cath Picc CathetersC.R. Bard, Inc. · 2004 · Recalled
- 6 generations backK050931 6 FR DL Powerpicc CatheterC.R. Bard, Inc. · 2005
- 5 generations backK051672 PowerpiccBard Access Systems, Inc. · 2005 · Recalled
- 4 generations backK060812 Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex…C.R. Bard, Inc. · 2006 · Recalled
- 3 generations backK062414 Vortex CT Port Access System and Lifeguard Safety Infusion SetRita Medical Systems, Inc. · 2007 · Recalled
- 2 generations backK073050 Safestep Max Power-Injectable Infusion SetSpecialized Health Products International, Inc. · 2008 · Recalled
- 1 generation backK170897 Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety…B.Braun Medical, Inc. · 2017
- This deviceK200463 Huber Needle Infusion Set, Safety Huber Needle Infusion SetJiangsu Caina Medical Co.,Ltd · 2021
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −10 | K001901 | Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline… | C.R. Bard, Inc. | 2000 | 2 |
| −9 | K012902 | Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath… | C.R. Bard, Inc. | 2001 | — |
| −8 | K031129 | Poly Per-Q-Cath Picc Catheters; Poly Radpicc | C.R. Bard, Inc. | 2003 | — |
| −7 | K034019 | Poly Per-Q-Cath Picc Catheters | C.R. Bard, Inc. | 2004 | 1 |
| −6 | K050931 | 6 FR DL Powerpicc Catheter | C.R. Bard, Inc. | 2005 | — |
| −5 | K051672 | Powerpicc | Bard Access Systems, Inc. | 2005 | 1 |
| −4 | K060812 | Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex Catheter-Multiple | C.R. Bard, Inc. | 2006 | 1 |
| −3 | K062414 | Vortex CT Port Access System and Lifeguard Safety Infusion Set | Rita Medical Systems, Inc. | 2007 | 2 |
| −2 | K073050 | Safestep Max Power-Injectable Infusion Set | Specialized Health Products International, Inc. | 2008 | 2 |
| −1 | K170897 | Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety II Needle… | B.Braun Medical, Inc. | 2017 | — |
| 0 (this device) | K200463 | Huber Needle Infusion Set, Safety Huber Needle Infusion Set | Jiangsu Caina Medical Co.,Ltd | 2021 | — |
| +1 | K254095 | 25G Huber Needle | Martech Medical Products | 2026 | — |
| +1 | K211121 | Intera Non-coring (Huber) Refill Needles, Intera Non-coring (Huber) Special… | Intera Oncology, Inc. | 2021 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1986 | 1 | 0 |
| 1987 | 2 | 0 |
| 1988 | 1 | 0 |
| 1989 | 4 | 0 |
| 1991 | 1 | 0 |
| 1992 | 1 | 0 |
| 1994 | 3 | 0 |
| 1996 | 1 | 0 |
| 2000 | 1 | 0 |
| 2001 | 2 | 0 |
| 2002 | 3 | 0 |
| 2004 | 3 | 0 |
| 2005 | 2 | 0 |
| 2006 | 1 | 0 |
| 2007 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2021 | 1 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in PTI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new PTI device
- #1 K254095 25G Huber Needlescore 15.0 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K170897 | Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety II Needle with…Summary page 6…Huber Needle Safety Huber Needle Remake K170897 Infusion Set Infusion Set Product Code PTI PTI… | 2017 |
What descends from it? (2)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 21 years and passes through 6 companies; a new generation was cleared every 12 months on median.
- 21 of the 37 ancestors have a recall of any class.
- Of its descendants, K254095 (Martech Medical Products, 2026) has gone furthest, with 0 descendants of its own.
- 1 other device also cites its predicate K170897.
How does its category compare?
All PTI devicesPTI Non-Coring (Huber) Needle: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.
FDA decisions per year · PTI
3in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · PTI
120in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · PTI
No recall names PTI since 2007.
Adverse event reports per year · PTI
33in 2025
MAUDE reports naming product code PTI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 5 | 221 | 0 | 2 |
| 2018 | 0 | — | 0 | 3 |
| 2019 | 0 | — | 0 | 12 |
| 2020 | 0 | — | 0 | 13 |
| 2021 | 2 | — | 0 | 17 |
| 2022 | 1 | — | 0 | 13 |
| 2023 | 1 | — | 0 | 9 |
| 2024 | 1 | — | 0 | 36 |
| 2025 | 3 | 120 | 0 | 33 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
Huber Needle Infusion Set The Huber Needle Infusion Set is a device with a non-coring right angle needle intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids and drugs, as well as blood sampling into the port. The 19G-22G needles of device are also suitable for power injection of contrast media to a maximum pressure of 325psi. When used with ports indicated for power injection, the maximum recommended…
Read the full statement (199 words)
Huber Needle Infusion Set
The Huber Needle Infusion Set is a device with a non-coring right angle needle intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids and drugs, as well as blood sampling into the port.
The 19G-22G needles of device are also suitable for power injection of contrast media to a maximum pressure of 325psi. When used with ports indicated for power injection, the maximum recommended infusion rate is approximately 5ml/s for 19G and 20G, 2ml/s for 22G.
Safety Huber Needle Infusion Set
The Safety Huber Needle Infusion Set is a device with a non-coring right angle needle intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids and drugs, as well as blood sampling into the port. The safety feature is manually activated during needle removal, and is designed to aid in the prevention of accidental needle sticks.
The 19G-22G needles of device are also suitable for power injection of contrast media to a maximum pressure of 325psi. When used with ports indicated for power injection, the maximum recommended infusion rate is approximately 5ml/s for 19G and 20G, 2ml/s for 22G.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K200463 · printed Nov 6, 2026
Cite this page
510(k) Report. “K200463: Huber Needle Infusion Set, Safety Huber Needle Infusion Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K200463
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026