# K200524: SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All, go. Sim, go. Open Pro, Scan&GO

- Pathway: 510(k) (Special)
- Applicant: Siemens Medical Solutions USA, Inc.
- Decided: Apr 1, 2020 after 30 days
- Product code: JAK (System, X-Ray, Tomography, Computed)
- Device class: 2
- Regulation: 21 CFR 892.1750
- Recalls of this device: 7 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 13 generations back to K013561 (2001). 45 ancestors in all.
14 later devices descend from it; 2 cite it directly.
0 of 60 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K191891: SOMATOM X.cite, Siemens Medical Solutions USA, Inc. (2019) Summary page 6.
- K192061: SOMATOM Go Platform – SOMATOM go.Now, go.Up, go.All, go.Top, go.Sim, go.Open Pro, Scan & GO, Siemens Medical Solutions USA, Inc. (2019) Summary page 6.

## Most influential descendants
- K211373: SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X. Platform Scanners - SOMATOM X.cite and SOMATOM X.ceed, Scan&GO Software (2021), 12 descendants of its own
- K233650: SOMATOM go.Up; SOMATOM go.Now; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM X.cite; SOMATOM X.ceed (2024), 5 descendants of its own
- K211591: NAEOTOM Alpha, Scan&GO (2021), 4 descendants of its own
- K220814: NAEOTOM Alpha, Scan&GO Software (2022), 3 descendants of its own
- K233657: NAEOTOM Alpha (2024), 2 descendants of its own

## Recalls of this device
- Dec 22, 2022 · Class II · Z-1003-2023: In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped
- Mar 7, 2022 · Class II · Z-0997-2022: Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Markin
- Mar 7, 2022 · Class II · Z-0962-2022: Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional
- Mar 7, 2022 · Class II · Z-0996-2022: Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Markin
- Aug 18, 2021 · Class II · Z-2500-2021: Software versions may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Sporadic software errors may also occur during interventional workflows, re
- Aug 16, 2021 · Class II · Z-2446-2021: Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the trea
- Mar 3, 2021 · Class II · Z-1253-2021: SOMATOM systems-issues in with software syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a, may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Delay in d

## Indications for use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. This in-room scan application is a planning and information system designed to perform the necessary functions required for planning and controlling scans of supported SIEMENS CT scanners. It allows users to work in close proximity to the scanner. The in-room scan application runs on standard information technology hardware and software, utilizing the standard information technology operating systems and user interface. Communication and data exchange are done using special protocols.

Source page: https://510kreport.com/device/K200524