510(k) Report

510(k) · Cleared

Cleared

Galapogos

K200565510(k) TraditionalCleared Jan 17, 2021319-day reviewClass II21 CFR 868.5905

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Resmed Corp · BZD Ventilator, Non-Continuous (Respirator)

Cleared on Jan 17, 2021 as substantially equivalent to K160836 (Resmed, Ltd., 2016) after a 319-day review. Its family tree reaches back 10 generations to 1992. 3 later devices descend from it.

Decided
Jan 17, 2021
Review
319 days
Class
Class II
Product code
BZD
Recalls
None
Predicates
1

Its longest line of descent runs back 10 generations to K913396 (1992). 51 ancestors in all, the oldest K905404 (1991). 3 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1991: 2 other ancestors2009: 2 other ancestors2008: 1 other ancestor2004: 6 other ancestors2007: 1 other ancestor2006: 1 other ancestor2003: 4 other ancestors2002: 3 other ancestors1998: 4 other ancestors1996: 5 other ancestors2001: 2 other ancestors1999: 2 other ancestors1992: 3 other ancestors1997: 1 other ancestor1994: 1 other ancestor1993: 1 other ancestor1995: 2 other ancestors2024: 1 descendant cleared2025: 2 descendants clearedReference device: K151901 · AirView · Resmed, Ltd. · 2015K913396 · Cannula Style Thermocouples · Pro-Tech, Inc. · 1992K982293 · Pressure Transducer Airflow Sensor · Pro-Tech Services, Inc. · 1998K020607 · Sleepflo, Model 101501 · Biomec, Inc. · 2002K030379 · Aplab, Model 801D0030 · Sector Medical Corp. · 2003K040576 · Micromesam Basic-Set · Map Medizintechnik Fur Arzt Und Patient GmbH · 2004K061405 · Apnealink, Model 22302 · Resmed Germany, Inc. · 2006K070263 · Apnealink · Resmed Germany, Inc. · 2007K083575 · Apnealink Plus, Model: 22328 · Resmed Germany, Inc. · 2009K140124 · S9 Elouera · Resmed Corp · 2014K160836 · Menai System · Resmed, Ltd. · 20161990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 1 other ancestor4 generations back: 3 other ancestors5 generations back: 8 other ancestors6 generations back: 11 other ancestors7 generations back: 8 other ancestors8 generations back: 4 other ancestors9 generations back: 5 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K160836 · Menai System · Resmed, Ltd. · 2016K140124 · S9 Elouera · Resmed Corp · 2014K083575 · Apnealink Plus, Model: 22328 · Resmed Germany, Inc. · 2009K070263 · Apnealink · Resmed Germany, Inc. · 2007K061405 · Apnealink, Model 22302 · Resmed Germany, Inc. · 2006K040576 · Micromesam Basic-Set · Map Medizintechnik Fur Arzt Und Patient GmbH · 2004K030379 · Aplab, Model 801D0030 · Sector Medical Corp. · 2003K020607 · Sleepflo, Model 101501 · Biomec, Inc. · 2002K982293 · Pressure Transducer Airflow Sensor · Pro-Tech Services, Inc. · 1998K913396 · Cannula Style Thermocouples · Pro-Tech, Inc. · 1992−10−50
  1. Root · 10 generations backK913396 Cannula Style ThermocouplesPro-Tech, Inc. · 1992
  2. 6 generations between
  3. 3 generations backK083575 Apnealink Plus, Model: 22328Resmed Germany, Inc. · 2009
  4. 2 generations backK140124 S9 EloueraResmed Corp · 2014
  5. 1 generation backK160836 Menai SystemResmed, Ltd. · 2016
  6. This deviceK200565 GalapogosResmed Corp · 2021
  7. Descendant · cites it directly · 2 of its ownK241216 myAirResmed Corp · 2024
  8. Descendant · 2 generations on · 1 of its ownK250624 myAirResmed Corp · 2025
  9. Descendant · 3 generations on · 0 of its ownK251657 Personalized Therapy Comfort Settings (PTCS)Resmed Corp · 2025
Show all 10 generations
  1. Root · 10 generations backK913396 Cannula Style ThermocouplesPro-Tech, Inc. · 1992
  2. 9 generations backK982293 Pressure Transducer Airflow SensorPro-Tech Services, Inc. · 1998
  3. 8 generations backK020607 Sleepflo, Model 101501Biomec, Inc. · 2002
  4. 7 generations backK030379 Aplab, Model 801D0030Sector Medical Corp. · 2003
  5. 6 generations backK040576 Micromesam Basic-SetMap Medizintechnik Fur Arzt Und Patient GmbH · 2004
  6. 5 generations backK061405 Apnealink, Model 22302Resmed Germany, Inc. · 2006
  7. 4 generations backK070263 ApnealinkResmed Germany, Inc. · 2007
  8. 3 generations backK083575 Apnealink Plus, Model: 22328Resmed Germany, Inc. · 2009
  9. 2 generations backK140124 S9 EloueraResmed Corp · 2014
  10. 1 generation backK160836 Menai SystemResmed, Ltd. · 2016
  11. This deviceK200565 GalapogosResmed Corp · 2021

Most influential descendants

Showing 3 of 3 · See all
  • K241216myAirResmed Corp · 2024 · cites it directly2descendants
  • K250624myAirResmed Corp · 2025 · 2 generations on1descendant
  • K251657Personalized Therapy Comfort Settings (PTCS)Resmed Corp · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K913396Cannula Style ThermocouplesPro-Tech, Inc.1992—
−9K982293Pressure Transducer Airflow SensorPro-Tech Services, Inc.1998—
−8K020607Sleepflo, Model 101501Biomec, Inc.2002—
−7K030379Aplab, Model 801D0030Sector Medical Corp.2003—
−6K040576Micromesam Basic-SetMap Medizintechnik Fur Arzt Und Patient GmbH2004—
−5K061405Apnealink, Model 22302Resmed Germany, Inc.2006—
−4K070263ApnealinkResmed Germany, Inc.2007—
−3K083575Apnealink Plus, Model: 22328Resmed Germany, Inc.2009—
−2K140124S9 EloueraResmed Corp2014—
−1K160836Menai SystemResmed, Ltd.2016—
0 (this device)K200565GalapogosResmed Corp2021—
+1K241216myAirResmed Corp2024—
+2K250624myAirResmed Corp2025—
+3K251657Personalized Therapy Comfort Settings (PTCS)Resmed Corp2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199120
199230
199310
199410
199520
199650
199710
199840
199920
200120
200230
200340
200460
200610
200710
200810
200920

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
202520
FIG. 1 — Lineage of K200565, 1992–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BZD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZD device

  1. #1 K250624 myAirscore 15.7 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K160836Menai System
Summary page 4
…ator (IPPB) (21 CFR 868.5905) Product Code: 73 BZD Predicate Device: Monte Carlo (mobile application in the Menai System) K160836 Device Description: Galapagos is a mobile interface where the patient can control the device thr…
Resmed, Ltd.2016—
†K151901AirViewResmed, Ltd.2015—

Also cleared against K160836: 2 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K241216myAirResmed Corp20242

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 29 years and passes through 8 companies; a new generation was cleared every 2.6 years on median.
  • 3 of the 51 ancestors have a recall of any class.
  • Of its descendants, K241216 (Resmed Corp, 2024) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K160836.

How does its category compare?

All BZD devices

BZD Ventilator, Non-Continuous (Respirator): FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · BZD

7in 2025

2007: 212008: 172009: 232010: 122011: 232012: 212013: 232014: 142015: 112016: 122017: 52018: 52019: 62020: 62021: 62022: 42023: 52024: 102025: 72026 (partial year): 10
2007high 23 (2009) · low 4 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZD

189in 2025

2007: 83 days median2008: 108 days median2009: 90 days median2010: 122 days median2011: 125 days median2012: 133 days median2013: 127 days median2014: 172 days median2015: 224 days median2016: 217 days median2017: 221 days median2018: 280 days median2019: 231 days median2020: 137 days median2021: 192 days median2022: 253 days median2023: 204 days median2024: 132 days median2025: 189 days median2026 (partial year): 214 days median
2007high 280 (2018) · low 83 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZD

8in 2023

2007: 292008: 22009: 12011: 12019: 4, 4 Class I2023: 8, 7 Class I2026 (partial year): 2, 2 Class I
2007high 29 (2007) · low 0 (2010)2026

Class I (13 since 2007) Class II/III not classified

Adverse event reports per year · BZD

60,279in 2025

2007: 652008: 672009: 1472010: 1432011: 1892012: 1392013: 822014: 862015: 1142016: 1182017: 1482018: 1012019: 1882020: 1262021: 6,0662022: 90,7682023: 27,9872024: 68,2622025: 60,279
2007high 90,768 (2022) · low 65 (2007)2025

MAUDE reports naming product code BZD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200721832965
200817108267
200923901147
2010121220143
2011231251189
2012211330139
201323127082
201414172086
2015112240114
2016122170118
201752210148
201852800101
201962314 (4)188
202061370126
2021619206,066
20224253090,768
202352048 (7)27,987
202410132068,262
20257189060,279
2026102142 (2)—

What is it indicated for?

The Galapagos app is intended for patients who are prescribed a compatible ResMed S10 platform device to simulate therapy prior to using their device with their prescribed settings. It is an optional software accessory to allow patients to acclimate to their therapy device.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K200565: Galapogos.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K200565

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026