510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Invacare Perfecto2V Oxygen Concentrator

K200890510(k) TraditionalCleared Aug 6, 2020125-day reviewClass II21 CFR 868.5440Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Invacare Corporation · CAW Generator, Oxygen, Portable

Cleared on Aug 6, 2020 as substantially equivalent to K020386 (Invacare Corp., 2002) after a 125-day review. Its family tree reaches back 3 generations to 1995. 1 later device descends from it. It has been recalled once.

Decided
Aug 6, 2020
Review
125 days
Class
Class II
Product code
CAW
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to K953815 (1995). 1 later device descends from it, 1 citing it directly.

  1. Root · 3 generations backK953815 5 LPM Oxygen ConcentratorDevilbiss Health Care, Inc. · 1995
  2. 2 generations backK991722 Devilbiss 5 Liter Oxygen ConcentratorSunrise Medical Hhg, Inc. · 1999 · Recalled
  3. 1 generation backK020386 Invacare Platinum, Models Irc 5LX and Irc 5LX02Invacare Corp. · 2002 · Recalled
  4. This deviceK200890 Invacare Perfecto2V Oxygen ConcentratorInvacare Corporation · 2020
  5. Descendant · cites it directly · 0 of its ownK203210 Invacare Platinum 5NXG Oxygen ConcentratorInvacare Corporation · 2021 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K203210Invacare Platinum 5NXG Oxygen ConcentratorInvacare Corporation · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K9538155 LPM Oxygen ConcentratorDevilbiss Health Care, Inc.1995—
−2K991722Devilbiss 5 Liter Oxygen ConcentratorSunrise Medical Hhg, Inc.19991
−1K020386Invacare Platinum, Models Irc 5LX and Irc 5LX02Invacare Corp.20022
0 (this device)K200890Invacare Perfecto2V Oxygen ConcentratorInvacare Corporation2020—
+1K203210Invacare Platinum 5NXG Oxygen ConcentratorInvacare Corporation20211

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
FIG. 1 — Lineage of K200890, 1995–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CAW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CAW device

  1. #1 K230052 Inogen Rove 6score 29.9 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020386Invacare Platinum, Models Irc 5LX and Irc 5LX02
Summary page 4
…21 CFR 868.5440 Regulatory Class: 2 Product Code: CAW PREDICATE DEVICE: Model Platinum 5 Oxygen Concentrator (K020386) Patient Population: Patients requiring supplemental oxygen DEVICE DESCRIPTION per 21 CFR 808.92(a)(4) The purpo…
Invacare Corp.20022

Also cleared against K020386: 4 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K203210Invacare Platinum 5NXG Oxygen ConcentratorInvacare Corporation20210

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 30, 2020Class IIZ-0514-2021Device DesignBreakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 4.0 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K203210 (Invacare Corporation, 2021) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K020386.

How does its category compare?

All CAW devices

CAW Generator, Oxygen, Portable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · CAW

12in 2025

2007: 62008: 112009: 82010: 52011: 62012: 52013: 22014: 62015: 12016: 12017: 12018: 32019: 22020: 52021: 22022: 42023: 82024: 22025: 122026 (partial year): 6
2007high 12 (2025) · low 1 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CAW

178in 2025

2007: 143 days median2008: 90 days median2009: 193 days median2010: 133 days median2011: 174 days median2012: 148 days median2014: 277 days median2018: 203 days median2020: 224 days median2022: 458 days median2023: 261 days median2025: 178 days median2026 (partial year): 260 days median
2007high 458 (2022) · low 90 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CAW

1in 2025

2007: 12008: 32009: 22011: 12013: 22014: 22016: 12017: 12019: 22020: 12022: 22024: 5, 1 Class I2025: 1
2007high 5 (2024) · low 0 (2010)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · CAW

611in 2025

2007: 232008: 252009: 432010: 482011: 1672012: 3952013: 2,3092014: 24,5382015: 18,9732016: 3,1672017: 1072018: 1162019: 1202020: 1252021: 1942022: 3052023: 3572024: 4502025: 611
2007high 24,538 (2014) · low 23 (2007)2025

MAUDE reports naming product code CAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076143123
20081190325
20098193243
20105133048
201161741167
201251480395
20132—22,309
20146277224,538
20151—018,973
20161—13,167
20171—1107
201832030116
20192—2120
202052241125
20212—0194
202244582305
202382610357
20242—5 (1)450
2025121781611
202662600—

What is it indicated for?

The Invacare® Perfecto.™V Oxygen Concentrator is indicated for patients with respiratory disorders requiring supplemental oxygen at flow rates of 1 to 5 liters per minute. For flow rates below 1 liter per minute use with Invacare Pediatric Flowmeter accessory. It is not intended to sustain or support life.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K200890: Invacare Perfecto2V Oxygen Concentrator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K200890

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026