VELYS™ Robotic-Assisted Solution

K202769510(k) TraditionalCleared Jan 14, 2021115-day reviewClass II21 CFR 882.4560Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Depuy Ireland UC · OLO Orthopedic Stereotaxic Instrument

Cleared on Jan 14, 2021 as substantially equivalent to K090953 (Praxim Medivision, SA, 2010) after a 115-day review. Its family tree reaches back 8 generations to 2000. 3 later devices descend from it. It has been recalled once.

Decided
Jan 14, 2021
Review
115 days
Class
Class II
Product code
OLO
Recalls
1
Predicates
1

Its longest line of descent runs back 8 generations to K002053 (2000). 26 ancestors in all, the oldest K983831 (1999). 3 later devices descend from it, 2 citing it directly.

  1. Root · 8 generations backK002053 Navitrak System-Optical Option, Model 900.004Orthosoft, Inc. · 2000
  2. 7 generations backK021760 Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc. · 2002
  3. 6 generations backK031196 Surgetics Ortho Kneelogics Navigation SystemPraxim · 2005
  4. 5 generations backK060282 Total Knee Surgetics Navigation SystemPraxim S.A · 2006
  5. 4 generations backK062146 Uni Knee Surgetics Navigation SystemPraxim S.A · 2006
  6. 3 generations backK063408 Ctlogics Navigations SystemsPraxim Medivision, SA · 2007
  7. 2 generations backK081232 Total Knee Surgetics Navigation System with PraxitelesPraxim · 2008
  8. 1 generation backK090953 Total Knee Surgetics Navigation System with IblockPraxim Medivision, SA · 2010 · Recalled
  9. This deviceK202769 VELYS™ Robotic-Assisted SolutionDepuy Ireland UC · 2021
  10. Descendant · cites it directly · 1 of its ownK233227 VELYS™ Robotic-Assisted SolutionDepuy Ireland UC · 2023
  11. Descendant · 2 generations on · 0 of its ownK260480 VELYS™ Robotic-Assisted SolutionDepuy Ireland UC · 2026
  12. Descendant · cites it directly · 0 of its ownK240721 ExcelsiusFlex™Globus Medical, Inc. · 2024

Most influential descendants

Showing 3 of 3 · See all
  • K233227VELYS™ Robotic-Assisted SolutionDepuy Ireland UC · 2023 · cites it directly1descendant
  • K260480VELYS™ Robotic-Assisted SolutionDepuy Ireland UC · 2026 · 2 generations on0descendants
  • K240721ExcelsiusFlex™Globus Medical, Inc. · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K002053Navitrak System-Optical Option, Model 900.004Orthosoft, Inc.2000—
−7K021760Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc.2002—
−6K031196Surgetics Ortho Kneelogics Navigation SystemPraxim2005—
−5K060282Total Knee Surgetics Navigation SystemPraxim S.A2006—
−4K062146Uni Knee Surgetics Navigation SystemPraxim S.A2006—
−3K063408Ctlogics Navigations SystemsPraxim Medivision, SA2007—
−2K081232Total Knee Surgetics Navigation System with PraxitelesPraxim2008—
−1K090953Total Knee Surgetics Navigation System with IblockPraxim Medivision, SA20102
0 (this device)K202769VELYS™ Robotic-Assisted SolutionDepuy Ireland UC2021—
+1K233227VELYS™ Robotic-Assisted SolutionDepuy Ireland UC2023—
+2K260480VELYS™ Robotic-Assisted SolutionDepuy Ireland UC2026—
+1K240721ExcelsiusFlex™Globus Medical, Inc.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199910
200020
200160
200250
200520
200620

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202410
202610
FIG. 1 — Lineage of K202769, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OLO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OLO device

  1. #1 K240662 Spine Guidance Software (version 5.0); Xia 3/…score 23.7 · 15 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090953Total Knee Surgetics Navigation System with Iblock
Summary page 4
…hopedics Regulation 882.4560 Product Code(s) OLO Legally marketed device(s) to TOTAL KNEE SURGETICS Navigation System with iBlock (K090953) which equivalence is claimed Reason for 510(k) submission New Device Device description The VELYS™ Robotic-Assis…
Praxim Medivision, SA20102

Also cleared against K090953: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K233227VELYS™ Robotic-Assisted SolutionDepuy Ireland UC20231
K240721ExcelsiusFlex™Globus Medical, Inc.20240

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 11, 2022Class IIZ-1029-2022Software designSystem software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a delay in treatment

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 5 companies; a new generation was cleared every 1.5 years on median.
  • 4 of the 26 ancestors have a recall of any class.
  • Of its descendants, K233227 (Depuy Ireland UC, 2023) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K090953.

How does its category compare?

All OLO devices

OLO Orthopedic Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · OLO

40in 2025

2009: 112010: 92011: 112012: 72013: 72014: 102015: 102016: 142017: 332018: 242019: 352020: 342021: 282022: 402023: 452024: 442025: 402026 (partial year): 42
2007high 45 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OLO

119in 2025

2009: 126 days median2010: 113 days median2011: 138 days median2012: 129 days median2013: 137 days median2014: 119 days median2015: 176 days median2016: 120 days median2017: 91 days median2018: 97 days median2019: 95 days median2020: 89 days median2021: 91 days median2022: 89 days median2023: 90 days median2024: 90 days median2025: 119 days median2026 (partial year): 109 days median
2007high 176 (2015) · low 89 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OLO

1in 2025

2010: 12011: 12012: 12014: 32015: 22016: 32017: 22018: 12019: 12020: 62022: 52023: 7, 1 Class I2024: 42025: 12026 (partial year): 6
2007high 7 (2023) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · OLO

3,306in 2025

2008: 12009: 42010: 42011: 102012: 602013: 362014: 2452015: 4662016: 7502017: 1,0762018: 1,2302019: 1,5852020: 2,7462021: 1,6962022: 1,5682023: 1,4932024: 2,3532025: 3,306
2007high 3,306 (2025) · low 0 (2007)2025

MAUDE reports naming product code OLO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—01
20091112604
2010911314
201111138110
20127129160
20137137036
2014101193245
2015101762466
2016141203750
2017339121,076
2018249711,230
2019359511,585
2020348962,746
2021289101,696
2022408951,568
202345907 (1)1,493
2024449042,353
20254011913,306
2026421096—

What is it indicated for?

The VELYS™ Robotic-Assisted Solution is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures, planning the position of the femoral and tibial implant components intraoperatively, and preparing the bones during total knee arthroplasty.

The VELYS™ Robotic-Assisted Solution is indicated for use with the ATTUNE Total Knee System and its cleared indications for use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K202769: VELYS™ Robotic-Assisted Solution.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K202769

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026