510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Electrosurgical Generator ESG-410, Foot Switches, PK Cutting Forceps

K203277510(k) TraditionalCleared Apr 28, 2021173-day reviewClass II21 CFR 878.4400Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Winter & Ibe GmbH · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Apr 28, 2021 as substantially equivalent to K141225 (Olympus Winter & Ibe GmbH, 2014) and 1 other predicate after a 173-day review. Its family tree reaches back 12 generations to 1994. 1 later device descends from it. It has been recalled 4 times.

Decided
Apr 28, 2021
Review
173 days
Class
Class II
Product code
GEI
Recalls
4
Predicates
2

Its longest line of descent runs back 12 generations to K936109 (1994). 88 ancestors in all, the oldest K800617 (1980). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2014: 1 other ancestor2011: 1 other ancestor2006: 1 other ancestor2003: 2 other ancestors2002: 5 other ancestors2000: 13 other ancestors1999: 4 other ancestors1994: 5 other ancestors1991: 5 other ancestors1997: 6 other ancestors1996: 9 other ancestors2001: 3 other ancestors1998: 6 other ancestors1995: 7 other ancestors1992: 3 other ancestors1980: 1 other ancestor1993: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestor1986: 1 other ancestor2023: 1 descendant clearedReference device: K151649 · Valleylab FT10 Electrosurgical Platform · Covidien · 2015Reference device: K060484 · Erbe Vio Esu, Model Vio 300 D · Erbe USA, Inc. · 2006K936109 · Sharplan Model 1120 CO2 Surgical Laser · Sharplan Lasers, Inc. · 1994K960521 · Sharplan Lasers, Inc. Silktouch Flash Scanner · Sharplan Lasers, Inc. · 1996K963339 · Coherent Ultrapulse CO2 Surgical Lasers · Coherent Medical Group · 1997K974789 · Coherent Ultrapulse S Series CO2 Surgical Lasers Model C and L · Lumenis, Inc. · 1998K990947 · Modified Versapulse Select Single Wavelength, HO: Yag and Dual Wavelength, HO: Yag/ND: Yag Surgical Lasers and Delivery · Lumenis, Inc. · 1999K011703 · Lumenis Versapulse Powersuite Holmium (HO:Yag) and Dual Wavelength (HO:Yag/ND:Yag) Surgical Lasers and Delivery Devices · Lumenis · 2001 · recalled×2: 2 generations cleared close together, K014290 (2002), K021777 (2002) K041285 · Gyrus G3 System (Generator, Assessories, Somnoplasty Electrodes and Plasmacision Electrodes) · Gyrus Medical, Ltd. · 2004 · recalledK050550 · Gyrus General Purpose Electrosurgical Generator · Gyrus Medical, Inc. · 2005K100896 · Gyrus Acmi Halo PKS Cutting Forceps · Gyrus Acmi, Inc. · 2010 · recalledK142759 · PK Cutting Forceps for ESG-400 Workstation · Gyrus Acmi, Inc. · 201519801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 4 other ancestors4 generations back: 6 other ancestors5 generations back: 8 other ancestors6 generations back: 11 other ancestors7 generations back: 16 other ancestors8 generations back: 9 other ancestors9 generations back: 11 other ancestors10 generations back: 5 other ancestors11 generations back: 3 other ancestors12 generations back: 1 other ancestor 1 generation forward: 1 descendant K142759 · PK Cutting Forceps for ESG-400 Workstation · Gyrus Acmi, Inc. · 2015K100896 · Gyrus Acmi Halo PKS Cutting Forceps · Gyrus Acmi, Inc. · 2010 · recalledK050550 · Gyrus General Purpose Electrosurgical Generator · Gyrus Medical, Inc. · 2005K041285 · Gyrus G3 System (Generator, Assessories, Somnoplasty Electrodes and Plasmacision Electrodes) · Gyrus Medical, Ltd. · 2004 · recalledK021777 · Gyrus Ent Cogent 2/G2 System (Generator,Accessories Somnoplasty Electrodes and Plasmacision Electrodes) · Gyrus Ent LLC · 2002K014290 · Entec Plasma Wands · Arthrocare Corp. · 2002K011703 · Lumenis Versapulse Powersuite Holmium (HO:Yag) and Dual Wavelength (HO:Yag/ND:Yag) Surgical Lasers and Delivery Devices · Lumenis · 2001 · recalledK990947 · Modified Versapulse Select Single Wavelength, HO: Yag and Dual Wavelength, HO: Yag/ND: Yag Surgical Lasers and Delivery · Lumenis, Inc. · 1999K974789 · Coherent Ultrapulse S Series CO2 Surgical Lasers Model C and L · Lumenis, Inc. · 1998K963339 · Coherent Ultrapulse CO2 Surgical Lasers · Coherent Medical Group · 1997K960521 · Sharplan Lasers, Inc. Silktouch Flash Scanner · Sharplan Lasers, Inc. · 1996K936109 · Sharplan Model 1120 CO2 Surgical Laser · Sharplan Lasers, Inc. · 1994−10−50
  1. Root · 12 generations backK936109 Sharplan Model 1120 CO2 Surgical LaserSharplan Lasers, Inc. · 1994
  2. 8 generations between
  3. 3 generations backK050550 Gyrus General Purpose Electrosurgical GeneratorGyrus Medical, Inc. · 2005
  4. 2 generations backK100896 Gyrus Acmi Halo PKS Cutting ForcepsGyrus Acmi, Inc. · 2010 · Recalled
  5. 1 generation backK142759 PK Cutting Forceps for ESG-400 WorkstationGyrus Acmi, Inc. · 2015
  6. This deviceK203277 Electrosurgical Generator ESG-410, Foot Switches, PK Cutting ForcepsOlympus Winter & Ibe GmbH · 2021
  7. Descendant · cites it directly · 0 of its ownK231777 Electrosurgical Generator ESG-410 and Accessories (WA91327U…Olympus Winter & Ibe GmbH · 2023 · Recalled
Show all 12 generations
  1. Root · 12 generations backK936109 Sharplan Model 1120 CO2 Surgical LaserSharplan Lasers, Inc. · 1994
  2. 11 generations backK960521 Sharplan Lasers, Inc. Silktouch Flash ScannerSharplan Lasers, Inc. · 1996
  3. 10 generations backK963339 Coherent Ultrapulse CO2 Surgical LasersCoherent Medical Group · 1997
  4. 9 generations backK974789 Coherent Ultrapulse S Series CO2 Surgical Lasers Model C and LLumenis, Inc. · 1998
  5. 8 generations backK990947 Modified Versapulse Select Single Wavelength, HO: Yag and Dual…Lumenis, Inc. · 1999
  6. 7 generations backK011703 Lumenis Versapulse Powersuite Holmium (HO:Yag) and Dual Wavelength…Lumenis · 2001 · Recalled
  7. 6 generations backK014290 Entec Plasma WandsArthrocare Corp. · 2002
  8. 5 generations backK021777 Gyrus Ent Cogent 2/G2 System (Generator,Accessories Somnoplasty…Gyrus Ent LLC · 2002
  9. 4 generations backK041285 Gyrus G3 System (Generator, Assessories, Somnoplasty Electrodes and…Gyrus Medical, Ltd. · 2004 · Recalled
  10. 3 generations backK050550 Gyrus General Purpose Electrosurgical GeneratorGyrus Medical, Inc. · 2005
  11. 2 generations backK100896 Gyrus Acmi Halo PKS Cutting ForcepsGyrus Acmi, Inc. · 2010 · Recalled
  12. 1 generation backK142759 PK Cutting Forceps for ESG-400 WorkstationGyrus Acmi, Inc. · 2015
  13. This deviceK203277 Electrosurgical Generator ESG-410, Foot Switches, PK Cutting ForcepsOlympus Winter & Ibe GmbH · 2021

Most influential descendants

Showing 1 of 1 · See all
  • K231777Electrosurgical Generator ESG-410 and Accessories (WA91327U, WA91327W), Accessories…Olympus Winter & Ibe GmbH · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K936109Sharplan Model 1120 CO2 Surgical LaserSharplan Lasers, Inc.1994—
−11K960521Sharplan Lasers, Inc. Silktouch Flash ScannerSharplan Lasers, Inc.1996—
−10K963339Coherent Ultrapulse CO2 Surgical LasersCoherent Medical Group1997—
−9K974789Coherent Ultrapulse S Series CO2 Surgical Lasers Model C and LLumenis, Inc.1998—
−8K990947Modified Versapulse Select Single Wavelength, HO: Yag and Dual Wavelength, HO…Lumenis, Inc.1999—
−7K011703Lumenis Versapulse Powersuite Holmium (HO:Yag) and Dual Wavelength…Lumenis20012
−6K014290Entec Plasma WandsArthrocare Corp.2002—
−5K021777Gyrus Ent Cogent 2/G2 System (Generator,Accessories Somnoplasty Electrodes and…Gyrus Ent LLC2002—
−4K041285Gyrus G3 System (Generator, Assessories, Somnoplasty Electrodes and…Gyrus Medical, Ltd.20042
−3K050550Gyrus General Purpose Electrosurgical GeneratorGyrus Medical, Inc.2005—
−2K100896Gyrus Acmi Halo PKS Cutting ForcepsGyrus Acmi, Inc.20101
−1K142759PK Cutting Forceps for ESG-400 WorkstationGyrus Acmi, Inc.2015—
0 (this device)K203277Electrosurgical Generator ESG-410, Foot Switches, PK Cutting ForcepsOlympus Winter & Ibe GmbH2021—
+1K231777Electrosurgical Generator ESG-410 and Accessories (WA91327U, WA91327W)…Olympus Winter & Ibe GmbH20232

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198610
198910
199010
199150
199230
199310
199450
199570
199690
199760
199860
199940
2000130
200130
200250
200320
200610
201110
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
FIG. 1 — Lineage of K203277, 1994–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141225Esg-400, Footswitch Double Pedal, Footswitch Single Pedal, P-Cord
Summary page 6
…ted devices: | Predicate Device | Manufacturer | 510(k) No | ESG-400 | Olympus Winter & Ibe GmbH | K141225 PK Cutting Forceps Olympus Surgical Technologies K142759 America T…
Olympus Winter & Ibe GmbH2014—
*K142759PK Cutting Forceps for ESG-400 Workstation
Summary page 6
…| 510(k) No | ESG-400 | Olympus Winter & Ibe GmbH | K141225 PK Cutting Forceps Olympus Surgical Technologies K142759 America Table 2.1: Identification of predicate device The following reference devices…
Gyrus Acmi, Inc.2015—
†K151649Valleylab FT10 Electrosurgical PlatformCovidien2015—
†K060484Erbe Vio Esu, Model Vio 300 DErbe USA, Inc.2006—

Also cleared against K141225: 5 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231777Electrosurgical Generator ESG-410 and Accessories (WA91327U, WA91327W), Accessories…Olympus Winter & Ibe GmbH20230

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 25, 2026Class IIZ-1640-2026Nonconforming Material/ComponentOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Jul 25, 2025Class IIZ-2339-2025Component change controlPotential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
Jun 21, 2024Class IIZ-2415-2024Component design/selectionElectrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery
Dec 15, 2023Class IIZ-1167-2024Under Investigation by firmThe jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 10 companies; a new generation was cleared every 1.9 years on median.
  • 7 of the 88 ancestors have a recall of any class.
  • Of its descendants, K231777 (Olympus Winter & Ibe GmbH, 2023) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K141225.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The electrosurgical generator, in conjunction with electrosurgical accessories and ancillary equipment, is intended for

cutting and coagulation of tissue in the following medical fields:

* Open surgery

* Laparoscopic surgery

* Endoscopic surgery

Only for use by a qualified physician in an adequate medical environment.

The PK Cutting Forceps are indicated for electrosurgical coagulation, mechanical cutting, and grasping of tissue during

the performance of laparoscopic and open general surgical procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K203277: Electrosurgical Generator ESG-410, Foot Switches, PK Cutting Forceps.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K203277

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026