510(k) Report

510(k) · Cleared

Cleared

HVT 2.0

K203357510(k) TraditionalCleared Aug 25, 2021282-day reviewClass II21 CFR 868.5450

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vapotherm, Inc. · BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Cleared on Aug 25, 2021 as substantially equivalent to K191010 (Vapotherm, Inc., 2019) after a 282-day review. Its family tree reaches back 9 generations to 1992. 1 later device descends from it.

Decided
Aug 25, 2021
Review
282 days
Class
Class II
Product code
BTT
Recalls
None
Predicates
1

Its longest line of descent runs back 9 generations to K913368 (1992). 21 ancestors in all, the oldest K902506 (1991). 1 later device descends from it, 1 citing it directly.

  1. Root · 9 generations backK913368 Models MR700/MR720/MR730 Dual Servo Resp Humid AccFisher & Paykel Electronics, Ltd. · 1992
  2. 5 generations between
  3. 3 generations backK092846 Airvo Series HumidifiersFisher & Paykel Healthcare, Ltd. · 2010
  4. 2 generations backK103316 FlowrestVapotherm, Inc. · 2011
  5. 1 generation backK191010 Palladium High Flow Therapy SystemVapotherm, Inc. · 2019
  6. This deviceK203357 HVT 2.0Vapotherm, Inc. · 2021
  7. Descendant · cites it directly · 0 of its ownK254078 HVT 2.0Vapotherm, Inc. · 2026
Show all 9 generations
  1. Root · 9 generations backK913368 Models MR700/MR720/MR730 Dual Servo Resp Humid AccFisher & Paykel Electronics, Ltd. · 1992
  2. 8 generations backK953392 HC500 Servo-Controlled Heated Respiratory Humidifier (& Accessories)Fisher & Paykel Electronics, Ltd. · 1996
  3. 7 generations backK983112 Respiratory Humidifier, Model # MR850Fisher & Paykel Healthcare, Ltd. · 1998 · Recalled
  4. 6 generations backK020332 Respiratory Humidifier, Model MR850Fisher & Paykel Healthcare, Ltd. · 2003 · Recalled
  5. 5 generations backK033710 MR850 Respiratory Humidifier, Model MR850JHUFisher & Paykel Healthcare, Ltd. · 2004
  6. 4 generations backK073706 Respiratory Humidifier, Model MR850Fisher & Paykel Healthcare, Ltd. · 2008
  7. 3 generations backK092846 Airvo Series HumidifiersFisher & Paykel Healthcare, Ltd. · 2010
  8. 2 generations backK103316 FlowrestVapotherm, Inc. · 2011
  9. 1 generation backK191010 Palladium High Flow Therapy SystemVapotherm, Inc. · 2019
  10. This deviceK203357 HVT 2.0Vapotherm, Inc. · 2021

Most influential descendants

Showing 1 of 1 · See all
  • K254078HVT 2.0Vapotherm, Inc. · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K913368Models MR700/MR720/MR730 Dual Servo Resp Humid AccFisher & Paykel Electronics, Ltd.1992—
−8K953392HC500 Servo-Controlled Heated Respiratory Humidifier (& Accessories)Fisher & Paykel Electronics, Ltd.1996—
−7K983112Respiratory Humidifier, Model # MR850Fisher & Paykel Healthcare, Ltd.19985
−6K020332Respiratory Humidifier, Model MR850Fisher & Paykel Healthcare, Ltd.20031
−5K033710MR850 Respiratory Humidifier, Model MR850JHUFisher & Paykel Healthcare, Ltd.2004—
−4K073706Respiratory Humidifier, Model MR850Fisher & Paykel Healthcare, Ltd.2008—
−3K092846Airvo Series HumidifiersFisher & Paykel Healthcare, Ltd.2010—
−2K103316FlowrestVapotherm, Inc.2011—
−1K191010Palladium High Flow Therapy SystemVapotherm, Inc.2019—
0 (this device)K203357HVT 2.0Vapotherm, Inc.2021—
+1K254078HVT 2.0Vapotherm, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199210
199310
199420
199510
199620
199810
200010
200420

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K203357, 1992–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BTT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BTT device

  1. #1 K222197 F&P Optiflow Junior 2/2+ Nasal Cannula…score 16.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K191010Palladium High Flow Therapy System
Summary page 4
…spiratory gas humidifier Classification Name: 21 CFR 868.5450, Class II Product Code: BTT Predicate Device: K191010 – Palladium High Flow Therapy System Device Description: The HVT 2.0 system is intended to deliver warmed and humidified high-fl…
Vapotherm, Inc.2019—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K254078HVT 2.0Vapotherm, Inc.20260

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 29 years and passes through 3 companies; a new generation was cleared every 2.7 years on median.
  • 2 of the 21 ancestors have a recall of any class.
  • Of its descendants, K254078 (Vapotherm, Inc., 2026) has gone furthest, with 0 descendants of its own.

How does its category compare?

All BTT devices

BTT Humidifier, Respiratory Gas, (Direct Patient Interface): FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · BTT

4in 2025

2007: 62008: 42009: 12010: 42011: 32012: 22013: 62014: 72015: 12016: 32017: 32018: 22019: 12020: 22021: 22022: 32023: 72024: 72025: 42026 (partial year): 4
2007high 7 (2014) · low 1 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BTT

104in 2025

2007: 73 days median2008: 199 days median2010: 113 days median2011: 116 days median2013: 171 days median2014: 231 days median2016: 157 days median2017: 303 days median2022: 290 days median2023: 258 days median2024: 219 days median2025: 104 days median2026 (partial year): 184 days median
2007high 303 (2017) · low 73 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BTT

1in 2025

2007: 12008: 12010: 22011: 12014: 5, 4 Class I2025: 12026 (partial year): 1
2007high 5 (2014) · low 0 (2009)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · BTT

1,743in 2025

2007: 3652008: 4562009: 5582010: 5622011: 6652012: 5832013: 7182014: 7182015: 6702016: 1,0942017: 1,0072018: 9182019: 8942020: 1,1882021: 1,5102022: 7492023: 7162024: 5012025: 1,743
2007high 1,743 (2025) · low 365 (2007)2025

MAUDE reports naming product code BTT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076731365
200841991456
20091—0558
201041132562
201131161665
20122—0583
201361710718
201472315 (4)718
20151—0670
2016315701,094
2017330301,007
20182—0918
20191—0894
20202—01,188
20212—01,510
202232900749
202372580716
202472190501
2025410411,743
202641841—

What is it indicated for?

The HVT 2.0 system is intended to deliver warmed and humidified high-flow respiratory gases to spontaneously breathing adult patients and pediatric patients (5 kg and up). The device is intended to be used in hospital, sub-acute facility, and home-use settings.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K203357: HVT 2.0.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K203357

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026