510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

EsophaCap Swallowable Cellular Retrieval Device (changed from Cell-Mata)

K203450510(k) TraditionalCleared May 4, 2021162-day reviewClass II21 CFR 874.4710Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Capnostics, LLC · EOX Esophagoscope (Flexible Or Rigid)

Cleared on May 4, 2021 as substantially equivalent to K934193 (Biosearch Medical Products, Inc., 1993) and 3 other predicates after a 162-day review. Its family tree reaches back 4 generations to 1993. It has been recalled 2 times.

Decided
May 4, 2021
Review
162 days
Class
Class II
Product code
EOX
Recalls
2
Predicates
4

Its longest line of descent runs back 4 generations to K934193 (1993). 5 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 4 generations backK934193 Cell-Mate (TM) Mass Cytology-Cellular Retrieval SystBiosearch Medical Products, Inc. · 1993
  2. 3 generations backK142695 Cytosponge Cell Collection DeviceCovidien, LLC · 2014
  3. 2 generations backK152794 Cytosponge Cell Collection DeviceCovidien, LLC · 2016 · Recalled
  4. 1 generation backK181020 Cytosponge Cell Collection DeviceCovidien, LLC · 2018 · Recalled
  5. This deviceK203450 EsophaCap Swallowable Cellular Retrieval Device (changed from…Capnostics, LLC · 2021
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K934193Cell-Mate (TM) Mass Cytology-Cellular Retrieval SystBiosearch Medical Products, Inc.1993—
−3K142695Cytosponge Cell Collection DeviceCovidien, LLC2014—
−2K152794Cytosponge Cell Collection DeviceCovidien, LLC20162
−1K181020Cytosponge Cell Collection DeviceCovidien, LLC20181
0 (this device)K203450EsophaCap Swallowable Cellular Retrieval Device (changed from Cell-Mata)Capnostics, LLC2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
201010
FIG. 1 — Lineage of K203450, 1993–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in EOX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new EOX device

  1. #1 K222366 EsoCheck Cell Collection Devicescore 7.8 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K934193Cell-Mate (TM) Mass Cytology-Cellular Retrieval Syst
Summary page 4
…OX Device Panel: Ear, Nose and Throat Devices Device Classification: Class II Legally Marketed Predicate Devices for Claimed Equivalency: a) K934193: Cell-MateTM Mass Cytology -Cellular Retrieval System b) K142695, K152794 and K181020: Cytosponge TM…
Biosearch Medical Products, Inc.1993—
*K142695Cytosponge Cell Collection Device
Summary page 4
…ification: Class II Legally Marketed Predicate Devices for Claimed Equivalency: a) K934193: Cell-MateTM Mass Cytology -Cellular Retrieval System b) K142695, K152794 and K181020: Cytosponge TM Cell Collection Device…
Covidien, LLC2014—
*K152794Cytosponge Cell Collection Device
Summary page 4
…: Class II Legally Marketed Predicate Devices for Claimed Equivalency: a) K934193: Cell-MateTM Mass Cytology -Cellular Retrieval System b) K142695, K152794 and K181020: Cytosponge TM Cell Collection Device Not…
Covidien, LLC20162
*K181020Cytosponge Cell Collection Device
Summary page 4
…Legally Marketed Predicate Devices for Claimed Equivalency: a) K934193: Cell-MateTM Mass Cytology -Cellular Retrieval System b) K142695, K152794 and K181020: Cytosponge TM Cell Collection Device Note: The CytoS…
Covidien, LLC20181

Also cleared against K934193: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 24, 2024Class IIZ-1900-2024Under Investigation by firmThe EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter…
Apr 24, 2024Class IIZ-1899-2024Under Investigation by firmThe EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 3 companies; a new generation was cleared every 2.7 years on median.
  • 3 of the 5 ancestors have a recall of any class.
  • 1 other device also cites its predicate K934193.

How does its category compare?

All EOX devices

EOX Esophagoscope (Flexible Or Rigid): FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · EOX

1in 2024

2007: 22011: 12012: 12013: 12014: 12016: 12018: 22019: 22020: 12021: 22022: 22023: 12024: 1
2007high 2 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · EOX

Too few EOX 510(k)s in any year since 2007 to give a median.

Recalls per year · EOX

2in 2011

2011: 2
2007high 2 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · EOX

2in 2024

2009: 12010: 12012: 12015: 12016: 32017: 12020: 22022: 12023: 122024: 2
2007high 12 (2023) · low 0 (2007)2025

MAUDE reports naming product code EOX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20080—00
20090—01
20100—01
20111—20
20121—01
20131—00
20141—00
20150—01
20161—03
20170—01
20182—00
20192—00
20201—02
20212—00
20222—01
20231—012
20241—02
20250—00
20260—0—

What is it indicated for?

The gathering and recovery of cells and cellular material from the mucosa in the esophagus for cytological and histological analyses.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K203450: EsophaCap Swallowable Cellular Retrieval Device (changed from Cell-Mata).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K203450

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026