# K203790: GEM Premier 5000

- Pathway: 510(k) (Special)
- Applicant: Instrumentation Laboratory CO
- Decided: Aug 20, 2021 after 235 days
- Product code: CHL (Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph)
- Device class: 2
- Regulation: 21 CFR 862.1120
- Recalls of this device: 25 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 10 generations back to K801848 (1980). 40 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 42 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K173403: GEM Premier 5000, Instrumentation Laboratory CO (2017) Summary page 5.
- K160415: GEM Premier 5000 (Measured parameters: Hematocrit, Total Hemoglobin, Carboxyhemoglobin, Methemoglobin, Deoxyhemoglobin, Oxyhemoglobin, Oxygen Saturation), Instrumentation Laboratory CO (2016) Summary page 5.
- K160402: GEM Premier 5000 (Measured Parameters:Glucose, Lactate and Total Bilirubin), Instrumentation Laboratory CO (2016) Summary page 5.
- K160225: GEM Premier 5000 (Measured Parameters:Sodium, Potassium, Chloride, Ionized Calcium), GEM CVP 5 tBili, GEM System Evaluator, GEM Hematocrit Evaluator, Instrumentation Laboratory CO (2016) Summary page 5.

## Most influential descendants
- K223608: GEM Premier 7000 with IQM3 (2023), 0 descendants of its own

## Recalls of this device
- May 21, 2026 · Class II · Z-2900-2026: The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognit
- May 8, 2026 · Class II · Z-2821-2026: BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a rac
- May 8, 2026 · Class II · Z-2824-2026: BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a rac
- May 8, 2026 · Class II · Z-2823-2026: BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a rac
- May 8, 2026 · Class II · Z-2825-2026: BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a rac
- May 8, 2026 · Class II · Z-2826-2026: BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a rac
- Jan 20, 2026 · Class II · Z-1541-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1526-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1542-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1530-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1525-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1533-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1536-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1524-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1531-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1538-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1540-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1529-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1539-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1534-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1532-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1528-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1527-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1537-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u
- Jan 20, 2026 · Class II · Z-1543-2026: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-u

## Indications for use
The GEM Premier 5000 is a portable critical care system for use by health care professionals to rapidly analyze heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2*) parameters from arterial, venous or capillary heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient’s acid/base status, electrolyte and metabolite balance and oxygen delivery capacity. *s02 = ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus deoxyhemoglobin. * pH, pCO2, and pO2 measurements in whole blood are used in the diagnosis and treatment of life-threatening acid-base disturbances. * Electrolytes in the human body have multiple roles. Nearly all metabolic processes depend on or vary with electrolytes: * Sodium (Na+) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison’s disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance. * Potassium (K+) measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. * lonized calcium (Ca++) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. * Chloride (Cl-) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders, such as cystic fibrosis and diabetic acidosis. » Hematocrit (Hct) measurements in whole blood of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red cells). * Glucose (Glu) measurement is used in the diagnosis, monitoring and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. * Lactate (Lac) measurement is used: » to evaluate the acid-base status of patients suspected of having lactic acidosis; * to monitor tissue hypoxia and strenuous physical exertion; » in the diagnosis of hyperlactatemia. » Total Bilirubin (tBili) measurement is used to aid in assessing the risk of kernicterus and hyperbilirubinemia in neonates. » CO-Oximetry (tHb, COHb, MetHb, O2Hb, HHb, and sO2) evaluates the ability of the blood to carry oxygen by measuring total hemoglobin and determining the percentage of functional and dysfunctional hemoglobin species. » Total Hemoglobin (tHb): Total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia. » COHb: Carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning. » MetHb: Methemoglobin measurements are used to determine different conditions of methemoglobinemia. » HHb: Deoxyhemoglobin, as a fraction of total hemoglobin, is used in combination with oxyhemoglobin to measure oxygen status. * O2Hb: Oxyhemoglobin, as a fraction of total hemoglobin, is used in combination with deoxyhemoglobin to measure oxygen status. * sO2: Oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus deoxyhemoglobin, is used to measure oxygen status.

Source page: https://510kreport.com/device/K203790