# K210164: Vantage Elan 1.5T, MRT-2020, V6.0

- Pathway: 510(k) (Traditional)
- Applicant: Canon Medical Systems Corporation
- Decided: Mar 10, 2021 after 48 days
- Product code: LNH (System, Nuclear Magnetic Resonance Imaging)
- Device class: 2
- Regulation: 21 CFR 892.1000
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K171597 (2017). 1 ancestor in all.
No later device cites it as a predicate yet.
0 of 2 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K171597: Vantage Elan 1.5T, V4.0, Toshibamedical Systems Corporation (2017) Summary page 5.

## Recalls of this device
- May 10, 2024 · Class II · Z-2041-2024: The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followe
- Jan 29, 2024 · Class II · Z-1249-2024: For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values descri

## Indications for use
Vantage Elan systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: •Proton density (PD) (also called hydrogen density) •Spin-lattice relaxation time (T1) •Spin-spin relaxation time (T2) •Flow dynamics •Chemical Shift Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.