# K210377: Stryker iBur hubs and cutting accessories

- Pathway: 510(k) (Traditional)
- Applicant: Stryker Corporation
- Decided: Jun 30, 2021 after 141 days
- Product code: HBE (Drills, Burrs, Trephines & Accessories (Simple, Powered))
- Device class: 2
- Regulation: 21 CFR 882.4310
- Recalls of this device: 10 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K112593 (2012). 2 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 4 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K143540: Stryker MIS Attachments and Cutting Accessories, Stryker Corporation (2015) Summary page 4.

## Most influential descendants
- K250374: iBur 3.0mm Diamond Match Head, Distal Bend (8431-107-030D); iBur 3.0mm Precision Round, Distal Bend (8431-009-030); iBur 4.0mm Precision Round, Distal Bend (8431-009-040); iBur 2.0mm Diamond Round, Distal Bend (8431-012-020D); iBur 3.0mm Diamond Round, Distal Bend (8431-012-030D); iBur 4.0mm Diamond Round, Distal Bend (8431-012-040D); iBur 3.0mm Coarse Diamond Round, Distal Bend (8431-013-030DC); iBur 4.0mm Coarse Diamond Round, Distal Bend (8431-013-040DC); iBur 5.0mm Coarse Diamond R (2025), 0 descendants of its own

## Recalls of this device
- Jul 24, 2024 · Class II · Z-2727-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis
- Jul 24, 2024 · Class II · Z-2724-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis
- Jul 24, 2024 · Class II · Z-2729-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis
- Jul 24, 2024 · Class II · Z-2725-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis
- Jul 24, 2024 · Class II · Z-2726-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis
- Jul 24, 2024 · Class II · Z-2731-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis
- Jul 24, 2024 · Class II · Z-2730-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis
- Jul 24, 2024 · Class II · Z-2722-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis
- Jul 24, 2024 · Class II · Z-2723-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis
- Jul 24, 2024 · Class II · Z-2728-2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tis

## Indications for use
The iBur hubs and cutting accessories are intended to be used with the Stryker Core Consolidated Operating Room (CORE) Console and electric and pneumatic motors. When used with these motors, the iBur hubs and cutting accessories are intended to cut bone in the following manner; drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/Otorhinolaryngology; and Endoscopic applications. Specific applications include Craniotomy/Craniectomy, Laminotomy/Laminectomy, Minimally Invasive Surgery (MIS) Spine, Expanded Endonasal Approach (EEA)/Anterior Skull Base/Endoscopic/Transnasal/Transsphenoidal, and Orthopedic Spine. These devices are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices.

Source page: https://510kreport.com/device/K210377