# K210500: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD

- Pathway: 510(k) (Traditional)
- Applicant: Elekta Limited
- Decided: Jun 11, 2021 after 109 days
- Product code: IYE (Accelerator, Linear, Medical)
- Device class: 2
- Regulation: 21 CFR 892.5050
- Recalls of this device: 10 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 10 generations back to K926449 (1995). 33 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 35 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K192242: Synergy Platform, Synergy, Infinity, VersaHD, Elekta Limited (2019) Summary page 4.

## Most influential descendants
- K252188: EMLA (Elekta Evo); EMLA (VersaHD); EMLA (Elekta Harmony Pro); EMLA (Elekta Infintiy); EMLA (Elekta Harmony); EMLA (Elekta Synergy) (2026), 0 descendants of its own

## Recalls of this device
- Dec 14, 2025 · Class II · Z-1164-2026: Due to a manufacturing issue and electrical grounding of systems.
- Oct 25, 2023 · Class II · Z-0443-2024: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fu
- Oct 25, 2023 · Class II · Z-0444-2024: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fu
- Oct 25, 2023 · Class II · Z-0447-2024: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fu
- Oct 25, 2023 · Class II · Z-0442-2024: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fu
- Oct 25, 2023 · Class II · Z-0445-2024: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fu
- Oct 25, 2023 · Class II · Z-0446-2024: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fu
- May 5, 2023 · Class II · Z-1665-2023: A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601
- Sep 28, 2021 · Class II · Z-0115-2022: There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected ma
- Sep 17, 2021 · Class II · Z-0116-2022: There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected ma

## Indications for use
The Elekta Medical Linear Accelerator (EMLA) is intended to be used for external beam radiation therapy (EBRT) treatments as determined by a licensed medical practitioner. It is intended to assist a licensed medical practitioner in the delivery of EBRT to defined target volumes, while sparing surrounding normal tissue and critical organs from excess radiation. −Elekta Synergy and Elekta Harmony are the default entry-level configurations. They are intended to be used for single or multiple fractions using standard dose fractionation, hyperfractionation, and hypofractionation in all areas of the body where such treatment is indicated. −Elekta Infinity is the default mid-level configuration. It is intended to be used for single or multiple fractions using standard dose fractionation, hyperfractionation, hypofractionation and stereotactic delivery (stereotactic body radiation therapy – SBRT; stereotactic ablative radiotherapy – SABR) in all areas of the body where such treatment is indicated. −Versa HD is the default high-level configuration. It is intended to be used for single or multiple fractions using standard fractionation, hyperfractionation, hypofractionation and stereotactic delivery (stereotactic body radiation therapy – SBRT; stereotactic ablative radiotherapy – SABR; stereotactic radio surgery - SRS) in all areas of the body where such treatment is indicated and for the treatment of functional disorders, such as trigeminal neuralgia. The EMLA is indicated for the delivery of curative and palliative intent EBRT to Adult and Pediatric patients with primary benign and malignant tumor and metastasis (or secondaries) anywhere in the body.