# K210810: Steiking Packaging for Medical Devices

- Pathway: 510(k) (Traditional)
- Applicant: Wipak OY
- Decided: Jan 23, 2022 after 312 days
- Product code: FRG (Wrap, Sterilization)
- Device class: 2
- Regulation: 21 CFR 880.6850
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K953776 (1995). 1 ancestor in all.
2 later devices descend from it; 1 cites it directly.
0 of 4 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K953776: Steriking Sterilization Pouches & Rolls, Wihuri OY (1995) Summary page 5.

## Most influential descendants
- K221016: Steriking Packaging for Medical Devices (2022), 1 descendant of its own
- K232625: Steriking® Pouch for Robotic Instruments (2024), 0 descendants of its own

## Indications for use
Steriking Packaging for Medical Devices serve as an enclosure for medical devices during steam sterilization that maintains the sterility of the enclosed medical devices until use of the medical devices with a combined weight of metal and plastics of 2.6 pounds or less. The recommended sterilization cycles are as follows: Pre-vacuum steam at 1320C for 4 minutes; Drying time of 20 minutes Pre-vacuum steam at 1350C for 3 minutes; Drying time of 16 minutes The Steriking Packaging for medical devices consists of a paper backing (Bleached wood pulp, grammage 100 g/m2) with transparent plastic film laminate front (2 sheets of laminated plastic with a total grammage of 55 g/m?2, 1 sheet of oriented polyester 12 microns thick, 1 sheet of coextruded polypropylene 40 microns thick. The plastic laminate is triple heat sealed to the backing paper. The open end of the pouch is to be heat sealed once a device is inserted. Heat sealing parameters to provide a sterile barrier are 1650C-2000C (3290F-3920F). Steriking Packaging for Medical Devices maintains the sterility of the enclosed devices for up to 12 months post Steam sterilization and before sterilization has a maximum shelf life of 5 years from the date of manufacture. Steriking Packaging for Medical Devices Dimensional configurations (2 sizes 200mm x §00mm, 250mm x 900mm)