510(k) Report

510(k) · Cleared

Cleared

CUBE Air Purifier

K211139510(k) TraditionalCleared Nov 26, 2021224-day reviewClass II21 CFR 880.6500

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Samsung Electronics Co., Ltd. · FRA Purifier, Air, Ultraviolet, Medical

Cleared on Nov 26, 2021 as substantially equivalent to K161468 (Transformair, LLC, 2017) after a 224-day review. Its family tree reaches back 4 generations to 2003.

Decided
Nov 26, 2021
Review
224 days
Class
Class II
Product code
FRA
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K023830 (2003). No later device cites it as a predicate yet.

  1. Root · 4 generations backK023830 Airocide TI02Kes Science & Technology, Inc. · 2003
  2. 3 generations backK033448 Electro-Optical Air Sterilizer with IonizerJohn Mfg., Ltd. · 2004
  3. 2 generations backK052732 Abracair Air Cleaner, Model QTZ300-60; Abracair Air Cleaner, Model…Abracair, Inc. · 2006
  4. 1 generation backK161468 Transformair Indoor Air PurifierTransformair, LLC · 2017
  5. This deviceK211139 CUBE Air PurifierSamsung Electronics Co., Ltd. · 2021
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K023830Airocide TI02Kes Science & Technology, Inc.2003—
−3K033448Electro-Optical Air Sterilizer with IonizerJohn Mfg., Ltd.2004—
−2K052732Abracair Air Cleaner, Model QTZ300-60; Abracair Air Cleaner, Model TI 100-30PAbracair, Inc.2006—
−1K161468Transformair Indoor Air PurifierTransformair, LLC2017—
0 (this device)K211139CUBE Air PurifierSamsung Electronics Co., Ltd.2021—
FIG. 1 — Lineage of K211139, 2003–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRA device

  1. #1 K212644 Aura Stormscore 13.9 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K161468Transformair Indoor Air Purifier
Summary page 4
…roduct Code: FRA • Panel: General Hospital • Regulation Number: 21 CFR §880.6500 • Device Class: Class II • Date prepared: 10/14/2021 Predicate Device Primary Predicate K161468, Transformair Indoor Air Purifier by Transformair, LLC. Reference Device K200500, Molekule Air Pro RX by Molekule, Inc. Indicat…
Transformair, LLC2017—
†K200500Molekule Air Pro RXMolekule, Inc.2020—

Also cleared against K161468: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 5 companies; a new generation was cleared every 4.8 years on median.
  • None of the 4 ancestors has a recall.
  • 2 other devices also cite its predicate K161468.

How does its category compare?

All FRA devices

FRA Purifier, Air, Ultraviolet, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · FRA

1in 2025

2014: 12017: 12020: 22021: 52022: 42023: 22024: 42025: 1
2007high 5 (2021) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRA

205in 2024

2021: 190 days median2022: 175 days median2024: 205 days median
2007high 205 (2024) · low 175 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRA

No recall names FRA since 2007.

Adverse event reports per year · FRA

2in 2024

2015: 12021: 12022: 12024: 2
2007high 2 (2024) · low 0 (2007)2025

MAUDE reports naming product code FRA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20141—00
20150—01
20160—00
20171—00
20180—00
20190—00
20202—00
2021519001
2022417501
20232—00
2024420502
20251—00
20260—0—

What is it indicated for?

The CUBE Air Purifier is a device intended for medical purposes that is used to destroy bacteria and viruses in the air by exposure to ultraviolet radiation. The CUBE Air Purifier has been demonstrated to destroy the following MS2 bacteriophage, Phi-X174 bacteriophage, Staphylococcus epidermidis, Escherichia coli entrained on the filter of the subject device under the following exposure conditions: Organisms — Name — Average Maximum log…

Read the full statement (104 words)

The CUBE Air Purifier is a device intended for medical purposes that is used to destroy bacteria and viruses in the air by exposure to ultraviolet radiation.

The CUBE Air Purifier has been demonstrated to destroy the following MS2 bacteriophage, Phi-X174 bacteriophage, Staphylococcus epidermidis, Escherichia coli entrained on the filter of the subject device under the following exposure conditions:

Organisms — Name — Average Maximum log reduction/exposure time (hours), Room temperature test

Virus — MS2 bacteriophage — 5.33±0.23 /60 mins

Virus — Phi-X174 bacteriophage — 5.34±0.11 /60 mins

Bacteria — Staphylococcus epidermidis — 5.36±0.28 /60 mins

Bacteria — Escherichia coli — 5.17±0.05 /60 mins

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K211139: CUBE Air Purifier.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K211139

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026