510(k) Report

510(k) · Cleared

Cleared

Idylla MSI Test

K211181510(k) TraditionalCleared Feb 27, 2023678-day reviewClass II21 CFR 864.1866

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biocartis NV · PZJ Lynch Syndrome Test System

Cleared on Feb 27, 2023 as substantially equivalent to K200129 (Promega Corporation, 2021) after a 678-day review. Its family tree reaches back 2 generations to 2017.

Decided
Feb 27, 2023
Review
678 days
Class
Class II
Product code
PZJ
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to DEN170030 (2017). No later device cites it as a predicate yet.

  1. Root · 2 generations backDEN170030 VENTANA anti-MLH-1(M1) Mouse Monoclonal Primary Antibody, VENTANA…Ventana Medical Systems · 2017
  2. 1 generation backK200129 OncoMate MSI Dx Analysis SystemPromega Corporation · 2021
  3. This deviceK211181 Idylla MSI TestBiocartis NV · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2DEN170030VENTANA anti-MLH-1(M1) Mouse Monoclonal Primary Antibody, VENTANA anti-PMS2…Ventana Medical Systems2017—
−1K200129OncoMate MSI Dx Analysis SystemPromega Corporation2021—
0 (this device)K211181Idylla MSI TestBiocartis NV2023—
FIG. 1 — Lineage of K211181, 2017–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PZJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PZJ device

  1. #1 K213348 BOND MMR Antibody Panelscore 4.9 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K200129OncoMate MSI Dx Analysis System
Cleared while this submission was under review
Summary page 5
…NV Proposed Predicate Device OncoMate™ MSI Dx Analysis System K200129 2. Device Description The Biocartis Idylla™ System covers the entire process from sample to result with fully integrated sample…
Promega Corporation2021—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 3 companies; a new generation was cleared every 3.7 years on median.
  • None of the 2 ancestors has a recall.

How does its category compare?

All PZJ devices

PZJ Lynch Syndrome Test System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · PZJ

2in 2023

2017: 12021: 12023: 22026 (partial year): 1
2007high 2 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PZJ

Too few PZJ 510(k)s in any year since 2007 to give a median.

Recalls per year · PZJ

No recall names PZJ since 2007.

Adverse event reports per year · PZJ

1in 2019

2019: 1
2007high 1 (2019) · low 0 (2007)2025

MAUDE reports naming product code PZJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20171—00
20180—00
20190—01
20200—00
20211—00
20220—00
20232—00
20240—00
20250—00
20261—0—

What is it indicated for?

For in vitro diagnostic use. For use on the Biocartis Idylla™ System only. The Idylla™ MSI Test, for use on the Idylla™ System, uses formalin-fixed, paraffin-embedded (FFPE) tissue sections of human CRC tumor, from which nucleic acids are liberated, then analyzed using PCR amplification of seven monomorphic biomarkers (ACVR2A, BTBD7, DIDO1, MRE11, RYR3, SEC31A and SULF2) and subsequent melt-curve analysis. The Idylla™ MSI Test reports results as either…

Read the full statement (186 words)

For in vitro diagnostic use.

For use on the Biocartis Idylla™ System only.

The Idylla™ MSI Test, for use on the Idylla™ System, uses formalin-fixed, paraffin-embedded (FFPE) tissue sections of human CRC tumor, from which nucleic acids are liberated, then analyzed using PCR amplification of seven monomorphic biomarkers (ACVR2A, BTBD7, DIDO1, MRE11, RYR3, SEC31A and SULF2) and subsequent melt-curve analysis. The Idylla™ MSI Test reports results as either microsatellite stable (MSS), or microsatellite instability high (MSI-H) or invalid.

Idylla™ MSI Test is indicated for use by healthcare professionals for the qualitative identification of microsatellite instability (MSI) in colorectal cancer (CRC) tumors, indicative of mismatch repair deficiency, as an aid in the identification of potential Lynch syndrome to help identify patients that would benefit from additional genetic testing to diagnose Lynch syndrome.

The results from the Idylla™ MSI Test should be interpreted by healthcare professionals in conjunction with other clinical findings, family history, and other laboratory data. The Idylla™ MSI Test should not be used for diagnosis of CRC.

The clinical performance of this device to guide treatment decision for MSI high patients has not been established.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K211181: Idylla MSI Test.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K211181

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026