# K211452: Encevis

- Pathway: 510(k) (Traditional)
- Applicant: Austrian Institute of Technology GmbH
- Decided: Dec 2, 2021 after 206 days
- Product code: OMB (Automatic Event Detection Software For Full-Montage Electroencephalograph)
- Device class: 2
- Regulation: 21 CFR 882.1400
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K132306 (2013). 3 ancestors in all.
3 later devices descend from it; 1 cites it directly.
0 of 7 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K132306: Persyst 12 Eeg Review and Analysis Software, Persyst Development Corp. (2013) Summary page 6.
- K171720: encevis, Ait Austrian Institute of Technology GmbH (2018) Summary page 6.

## Most influential descendants
- K240993: encevis (2.1) (2024), 2 descendants of its own
- K261712: Ceribell EEG Analysis Software (2026), 0 descendants of its own
- K260998: Ceribell Neurology Review Software (2026), 0 descendants of its own

## Indications for use
1. encevis is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EEG. This device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information. 2. The seizure detection component of encevis is intended to mark previously acquired sections of adult (greater than or equal to 18 years) EEG recordings that may correspond to electrographic seizures, in order to assist qualified clinical practitioners in the assessment of EEG traces. EEG recordings should be obtained with a full scalp montage according to the standard 10/20-system. 3. The spike detection component of encevis is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to spikes, in order to assist qualified clinical practitioners in the assessment of EEG traces. The Spike Detection component is intended to be used in adult patients greater than or equal to 18 years. encevis Spike Detection performance has not been assessed for intracranial recordings. 4. encevis includes the calculation and display of a set of quantitative measures intended to monitor and analyze the EEG waveform. These include frequency bands, rhythmic and periodic patterns and burst suppression. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms. 5. The aEEG functionality included in encevis is intended to monitor the state of the brain. 6. encevis provides notifications on an on-screen display for seizure detection, spike detection, quantitative EEG and aEEG that can be used when processing a record during acquisition. Delays of up to several minutes can occur between the beginning of a seizure, the occurrence of a spike or detection of quantitative EEG features and when the encevis notifications will be shown to a user. encevis notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a trained expert. 7. encevis PureEEG (Artifact Reduction) is intended to reduce EMG and electrode artifacts in a standard 10-20 EEG recording. PureEEG does not remove the entire artifact signal, and is not effective for other types of artifacts. PureEEG may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms. 8. This device does not provide any diagnostic conclusion about the patient's condition to the user.