510(k) Report

510(k) · Cleared

Cleared

Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour)

K211758510(k) TraditionalCleared Aug 3, 202157-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hartalega NGC Sdn. Bhd. · OPC Powder-Free Polychloroprene Patient Examination Glove

Cleared on Aug 3, 2021 after a 57-day review. FDA’s file for it is a 510(k) statement, which names no predicate.

Decided
Aug 3, 2021
Review
57 days
Class
Class I
Product code
OPC
Recalls
None
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K211758Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OPC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All OPC devices

OPC Powder-Free Polychloroprene Patient Examination Glove: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · OPC

1in 2021

2021: 1
2007high 1 (2021) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OPC

Too few OPC 510(k)s in any year since 2007 to give a median.

Recalls per year · OPC

No recall names OPC since 2007.

Adverse event reports per year · OPC

1in 2023

2021: 12023: 1
2007high 1 (2021) · low 0 (2007)2025

MAUDE reports naming product code OPC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20170—00
20180—00
20190—00
20200—00
20211—01
20220—00
20230—01
20240—00
20250—00
20260—0—

What is it indicated for?

Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical…

Read the full statement (163 words)

Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and Concentration Minimum Breakthrough Detection Time in Minutes

Carmustine (3.3 mg/ml) 14.9

Cisplatin (1.0 mg/ml) >240

Cyclophosphamide (20.0 mg/ml) >240

Dacarbazine (10.0 mg/ml) >240

Doxorubicin Hydrochloride (2.0 mg/ml) >240

Etoposide (20.0 mg/ml) >240

Fluorouracil (50.0 mg/ml) >240

Methotrexate (25.0 mg/ml) >240

Mitomycin C (0.5 mg/ml) >240

Paclitaxel (6.0 mg/ml) >240

Thiotepa (10.0 mg/ml) 22.9

Vincristine Sulfate (1.0 mg/ml) >240

Please note that Carmustine and Thiotepa have extremely low permeation times of 14.9 minutes and 22.9 minutes.

Warning: Do not use with Carmustine

Warning: Do not use with Thiotepa

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K211758: Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K211758

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026