510(k) Report

510(k) · Cleared

Cleared

Shamrock Powder Free Blue Nitrile Examination Gloves (Tested for use with Chemotherapy Drugs and Fentanyl)

K211810510(k) TraditionalCleared May 23, 2022346-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pt. Shamrock Manufacturing Corpora · LZA Polymer Patient Examination Glove

Cleared on May 23, 2022 as substantially equivalent to K111248 (Yty Industry (Manjung) Sdn Bhd, 2011) after a 346-day review. Its family tree reaches back 3 generations to 2009. 1 later device descends from it.

Decided
May 23, 2022
Review
346 days
Class
Class I
Product code
LZA
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K082957 (2009). 1 later device descends from it, 1 citing it directly.

  1. Root · 3 generations backK082957 Non-Sterile, Powder Free Nitrile Examination Glove, Blue Color…Siam Sempermed Corp., Ltd. · 2009
  2. 2 generations backK101822 Non-Sterile Powder-Free Blue Nitrile Examination Gloves Tested for…Sgmp Co., Ltd. · 2010
  3. 1 generation backK111248 Non-Sterile, Powder-Free Nitrile Examination Gloves (Original Blue…Yty Industry (Manjung) Sdn Bhd · 2011
  4. This deviceK211810 Shamrock Powder Free Blue Nitrile Examination Gloves (Tested for use…Pt. Shamrock Manufacturing Corpora · 2022
  5. Descendant · cites it directly · 0 of its ownK232064 Powder-Free Frost Nitrile Examination Gloves (Blue) (Tested for use…Isikel Manufacturing, LLC · 2023

Most influential descendants

Showing 1 of 1 · See all
  • K232064Powder-Free Frost Nitrile Examination Gloves (Blue) (Tested for use with Chemotherapy…Isikel Manufacturing, LLC · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K082957Non-Sterile, Powder Free Nitrile Examination Glove, Blue Color, Polymer…Siam Sempermed Corp., Ltd.2009—
−2K101822Non-Sterile Powder-Free Blue Nitrile Examination Gloves Tested for Use with…Sgmp Co., Ltd.2010—
−1K111248Non-Sterile, Powder-Free Nitrile Examination Gloves (Original Blue, Cobalt Blue)Yty Industry (Manjung) Sdn Bhd2011—
0 (this device)K211810Shamrock Powder Free Blue Nitrile Examination Gloves (Tested for use with…Pt. Shamrock Manufacturing Corpora2022—
+1K232064Powder-Free Frost Nitrile Examination Gloves (Blue) (Tested for use with…Isikel Manufacturing, LLC2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
FIG. 1 — Lineage of K211810, 2009–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZA device

  1. #1 K192333 Blue Nitrile Examination Gloves Powder Freescore 71.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K111248Non-Sterile, Powder-Free Nitrile Examination Gloves (Original Blue, Cobalt Blue)
Summary page 5
…Class: Class I Predicate Device: YTY Industry (Manjung): Non-sterile, Powder-free Nitrile Examination Gloves (Cobalt Blue) tested for use with Chemotherapy Drugs (K111248) Device Description: Shamrock Powder Free Blue Nitrile Examination Gloves (Tested for use with Chemotherapy Drugs and Fentanyl)…
Yty Industry (Manjung) Sdn Bhd2011—

Also cleared against K111248: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K232064Powder-Free Frost Nitrile Examination Gloves (Blue) (Tested for use with Chemotherapy…Isikel Manufacturing, LLC20230

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 1.5 years on median.
  • None of the 3 ancestors has a recall.
  • Of its descendants, K232064 (Isikel Manufacturing, LLC, 2023) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K111248.

How does its category compare?

All LZA devices

LZA Polymer Patient Examination Glove: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · LZA

24in 2025

2007: 212008: 182009: 282010: 312011: 262012: 332013: 202014: 272015: 202016: 302017: 442018: 222019: 252020: 302021: 1392022: 1462023: 622024: 302025: 242026 (partial year): 15
2007high 146 (2022) · low 18 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZA

110in 2025

2007: 95 days median2008: 97 days median2009: 112 days median2010: 97 days median2011: 122 days median2012: 110 days median2013: 130 days median2014: 197 days median2015: 112 days median2016: 171 days median2017: 124 days median2018: 178 days median2019: 212 days median2020: 101 days median2021: 111 days median2022: 111 days median2023: 92 days median2024: 89 days median2025: 110 days median2026 (partial year): 103 days median
2007high 212 (2019) · low 89 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZA

3so far in 2026, the most since 2007

2015: 12016: 12026 (partial year): 3
2007high 3 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZA

73in 2025

2007: 72008: 22009: 122010: 122011: 92012: 112013: 102014: 82015: 122016: 252017: 42018: 32019: 212020: 42021: 202022: 192023: 242024: 762025: 73
2007high 76 (2024) · low 2 (2008)2025

MAUDE reports naming product code LZA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007219507
2008189702
200928112012
20103197012
20112612209
201233110011
201320130010
20142719708
201520112112
201630171125
20174412404
20182217803
201925212021
20203010104
2021139111020
2022146111019
20236292024
20243089076
202524110073
2026151033—

What is it indicated for?

A patient examination glove is a disposable device intended for medical purpose that is worn on examiner’s hand to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05(2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs and Fentanyl. Chemotherapy Drug and Permeation The following chemicals have been tested with these gloves: Chemotherapy…

Read the full statement (235 words)

A patient examination glove is a disposable device intended for medical purpose that is worn on examiner’s hand to

prevent contamination between patient and examiner.

These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05(2019) Standard Practice for

Assessment of Medical Gloves to Permeation by Chemotherapy Drugs and Fentanyl.

Chemotherapy Drug and Permeation

The following chemicals have been tested with these gloves:

Chemotherapy Drug and Concentration Breakthrough Detection Time in Minutes

Carboplatin (Paraplatin), 10 mg/ml (10,000ppm) > 240 min.

Carmustine (BCNU), 3.3 mg/ml (3,300ppm) 46.6 min.

Chloroquine 50mg/ml (50,000ppm) > 240 min

Cisplatin, 1.0 mg/ml (1,000ppm) > 240 min

Cyclophosphamide (Cytoxan), 20.0mg/ml (20,000 ppm) > 240 min

Dacarbazine, 10.0mg/ml (10,000ppm) > 240 min

Docetaxel, 10mg/ml (10,000ppm) > 240 min

Doxorubicin HCI, 2.0mg/ml (2,000ppm) > 240 min

Etoposide, 20.0mg/ml (20,000ppm) > 240 min

Fluorouracil, 50.0mg/ml (50,000ppm) > 240 min

Ifosfamide, 50 mg/ml (50,000 ppm) > 240 min

Methotrexate, 25mg/ml (25,000 ppm) > 240 min

Mitomycin C, 0.5 mg/ml (500 ppm) > 240 min

Paclitaxel, 6.0 mg/ml (6,000 ppm) > 240 min

Thiotepa, 10.0mg/ml (10,000ppm) 64.8 min

Vincristine Sulfate, 1 mg/ml (1,000 ppm) > 240 min

Fentanyl Citrate Injection 100mcg/2ml (50mcg/1ml) > 240 min

Please note that the following drugs that have low permeation time are:

Carmustine (BCNU), 3.3 mg/ml 46.6 min

Thiotepa, 10.0mg/ml 64.8 min

CAUTION: Testing showed an average of breakthrough time of 46.6 min for Carmustine and 64.8 min for Thiotepa.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K211810: Shamrock Powder Free Blue Nitrile Examination Gloves (Tested for use with Chemotherapy Drugs and Fentanyl).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K211810

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026